Georgiy Alekseevich Solomakha1, Felix Glang1, Theodor Steffen2, Klaus Scheffler1,3, and Nikolai Ivanovich Avdievich 1
1High-field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Electronical Workshop, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 3Department for Biomedical Magnetic Resonance, University of Tübingen, Tuebingen, Germany
Synopsis
Keywords: RF Arrays & Systems, RF Arrays & Systems
Recently developed dynamic parallel
imaging, i.e. rapid modulation of element's sensitivities during acquisition, greatly
improved the performance of the method. In this work, sensitivity profiles of
dipole elements were electronically reconfigured by varying impedances of
lumped element
circuits connected in series with the dipoles . This required a large number of
DC wires directly connected to the dipoles. In the present work, we developed,
constructed, and evaluated a dynamically reconfigurable 8-element coaxial
dipole array for brain imaging at 9.4T. Our design eliminates the DC wires
directly connected to the dipoles and, hence, substantially simplifies further
increasing the number of receive channels.
Introduction
Parallel imaging is an important
MRI method that increases the acquisition speed(1). A standard benchmark for evaluation
of the parallel imaging performance is a so-called g-factor(1). To minimize
the g-factor value, a large number of receive coils with spatially diverse
sensitivity is required. Recently, a novel method for improvement of
parallel imaging without additional receive elements, i.e. dynamic switching of
the individual coil's sensitivities during acquisition, has been developed(2,3).
Firstly, a loop array with a variable in-plane sensitivity was presented(2).
Following this line of research, an array of eight dipoles with
modulated sensitivity profiles along the dipole length was developed (3). This
design enabled an acceleration in the head-foot direction(3). Sensitivity
profile of each dipole was electronically reconfigured by varying impedances
of circuits placed in each arm of the dipole. This required 32 DC wires for an 8-element array prototype. Since dipoles are very
sensitive to parallel conductors, further increasing the number of receive channels
is very difficult using this design. E.g., current state-of-the-art receive
arrays have 30 and above elements. In the present work, we developed, constructed, and
evaluated a dynamically reconfigurable 8-channel coaxial dipole array for brain imaging at 9.4T. Our design eliminates the DC wires directly connected to
the dipoles. It also reduces the inductance value required for current
distribution manipulation. Methods
Use of coaxial dipoles in MRI was
previously suggested in(4). Following this work, the reconfigurable coaxial
dipole consisted of a 13-cm coaxial line with two gaps in the shield located at
2-cm distance from the ends (Fig 1A). In contrast to(4), in our design the shield and
inner conductor were shorted at the driving port location. At both ends the
core and shield were connected by lumped element circuits, impedance of which
can be switched between inductive and capacitive. By connecting an inductor and
capacitor to the opposite ends of the dipole (Fig.1A), two current distributions with its maximum shifted toward one or the other ends of the
dipole are formed (Fig.1). Dynamic switching between these two states produces
two virtual rows of elements. In the final array design, we placed dipoles in
the so-called "chess order" as shown in Fig.1B, which allows an
acceleration in the head-feet direction even for static configurations(3).
Fig.2 shows the final schematic of the reconfigurable coaxial dipole element
(Fig.2A), photos of the array (Fig.2B), and the entire set up (Figs.2C,D). For
switching between the two states, PIN diodes were connected in series to inductors
in opposite directions (Fig.2A). Thus, applying positive voltage, we produced a
50-nH inductor at one end and small capacitor (~1pF) at the other end. For the
negative bias, we had the opposite configuration. With no voltage applied, the
dipole was detuned. DC voltage was delivered through the RF cable (Fig.2B). The
final set up consisted of the developed receive-only 8-element coaxial dipole
array and 8-element transmit-only loop array (Figs.2C,D). Numerically simulated dipole sensitivity profiles (CST studio) were later combined in MatLab to
calculate g-factors. Parallel receive performance was investigated both on a phantom
and in-vivo. Sensitivity profiles were switched every 10µs for a strongly
oversampled readout with a dwell time of 1µs. Retrospective SENSE
reconstructions have been performed using coil sensitivity maps via ESPIRIT(5)
from the central 48x48 k-space lines. All data were pre-whitened according to
noise covariance obtained from a separate noise-only scan. Data was
acquired on a Siemens 9.4T human MR-scanner with an RF and gradient spoiled GRE sequence (TR/TE=20ms/10ms, FA=20°, 200x200x60 matrix, FOV=220mmx220mmx120mm, 20% slice oversampling). Results and Discussion
Figs.3 and 4 show numerically
simulated (Fig.3) and experimentally measured (Fig.4) g-factor improvement due
to switching of asymmetric dipole sensitivities as compared to static cases. Fig.5
shows in-vivo data. For example, in the case of RyxRz=2x2 acceleration,
switching provides an 8% lower mean and 33% lower maximum g-factor in-vivo, and
~20% lower mean and ~2.5 times lower maximum g-factor for the phantom
measurements. In all settings, i.e., simulation, phantom and in vivo
measurements, rapid switching between the sensitivity profiles strongly reduces
both mean and maximum g-factors compared to the conventional static cases. The
absolute values of g-factors for a specific acceleration factor and the
observed switching improvement varied between all of these settings due to
various experimental conditions like FOV and object masking.
In general, the idea behind the
developed coaxial dipole design is similar to the previously reported common reconfigurable
dipoles(3). However, the new design has the following benefits. Firstly, it
eliminates the DC wires connected directly to the dipoles and provides
an easy way of further increasing the number of receive elements in the array.
In addition, it uses lower tuning inductances, i.e. 50nH vs 120nH(3). The
proposed reconfigurable coaxial dipole elements can be also combined with loops
to further minimize g-factors and improve central SNR(6).Conclusion
We developed, constructed, and
evaluated a 8-element reconfigurable receive-only coaxial dipole array for
dynamic parallel imaging of the brain at 9.4T. This novel design eliminates
DC wires connected directly to the dipoles. It also substantially reduces the inductances in comparison to the previous design. Finally, dynamic switching provides 8% lower mean and 33% lower maximum g-factors (for the case
of RyxRz=2x2 acceleration) in-vivo compared to the static case.Acknowledgements
Financial
support of the ERC Advanced Grant “SpreadMRI”, No 834940 is gratefully
acknowledged.References
1.
Pruessmann, K. P., Weiger, M., Scheidegger, M. B., & Boesiger, P. (1999). SENSE: sensitivity encoding for
fast MRI. Magnetic Resonance in Medicine, 42(5), 952-962.
2.
Glang F, Nikulin AV, Bause J, et al. Accelerated MRI at 9.4 T with electronically modulated
time-varying receive sensitivities. Magnetic Resonance in Medicine
2022;88:742–756
3.
Anton V. Nikulin, Felix Glang, Nikolai Avdievich, Dario Bosch, Theodor Steffen
and Klaus Scheffler (2022). Dynamic Radiofrequency Coils to Accelerate Parallel Imaging at
Ultra-High Magnetic Field IEEE. Transactions on Medical Imaging (in revision)
4. van Leeuwen, C. C., Steensma, B. R., Klomp, D. W., van
den Berg, C. A., & Raaijmakers, A. J. (2022). The Coax Dipole: A fully
flexible coaxial cable dipole antenna with flattened current distribution for
body imaging at 7 Tesla. Magnetic Resonance in Medicine, 87(1), 528-540.
5. Uecker M, Lai P, Murphy MJ, et al. ESPIRiT—an
eigenvalue approach to autocalibrating parallel MRI: Where SENSE meets GRAPPA.
Magnetic Resonance in Medicine 2014;71:990–1001
6. Avdievich
NI, Nikulin AV, Ruhm L, Magill AW, Henning A, Scheffler K. Double-Row Dipole/Loop
Combined Array for Human Whole Brain Imaging at 7 T. NMR in BioMed 2022, 35
(10), e4773.