Feng Du1,2, Nan Li1,2, Xiaoliang Zhang3, Xin Liu1,2, Hairong Zheng1,2, and Ye Li1,2
1Paul C. Lauterbur Imaging Research Center, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China, Shenzhen, China, 2Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province, Shenzhen 518055, Guang Dong, China, Shenzhen, China, 3Department of Biomedical Engineering, State University of New York, Buffalo, NY, United States, Buffalo, NY, United States
Synopsis
Keywords: RF Arrays & Systems, New Devices
Multinuclear MRI has been demonstrated as a crucial tool for comprehensively
characterizing tumor pathology and monitoring tumor treatment response as it
can provide biochemical, physical, and functional as well as structural
information. The objective of this work is to develop a
quadruple-nuclear transceiver coil array for
1H/
19F/
23Na/
31P
MRI at 3T to simultaneously detect multinucleal signals. The phantom studies were
performed on self-developed 3T MRI system to verify the performance. These results proved
that uniform excitation and highly sensitive acquisition in the region of
interest were achieved by utilizing the proposed RF coil and indicated the fesibility for multinuclear MRI applications.
Introduction
Multinuclear MRI
technique has been shown to quantify some physiological or pathological
indicators such as 19F, 31P and 23Na in tissues in addition to provide
high-quality anatomical images, which is widely used in clinical medicine and preclinical
medical research owing to noninvasion in obtaining biological information such
as tissue anatomy, physiology and metabolism[1-4]. The inherent physical
properties of weak nuclides make the imaging challenging at 3T[5]. The ability
to detect extremely weak signal depends on the corresponding RF transceiver. As the critical component of MRI systems used to transmit and receive MR
signals, RF transceivers were required for higher performance. However, the
design of multi-nuclear RF coils is complicated as interactions and
electromagnetic interference between each individually tuned coil element must
be overcome. The aim of this work is to design and construct a quadruple-nuclear transceiver coil array for 1H/19F/23Na/31P
MRI at 3T to meet the requirement of detecting many different sensitive nuclei simultaneously.
The bench tests were performed by utilizing the proposed coil array to validate
the performance and the feasibility for multi-nuclear MRI.Method
A quadruple-nuclear transceiver coil array was developed for 1H/19F/23Na/31P MRI at
3T and two-layered concentric RF coil arrangement was built as shown in Fig. 1.
The outer 180 mm-diameter tube was wrapped by four-loop transceiver coils that
resonate at 51.9MHz for 31P imaging, and the loop dimension was 190 mm × 160
mm with a width 5 mm. The inner 160mm-diameter tube wrapped by four triple-tuned
loop coils that resonate at 128.2MHz, 120.6MHz and 33.9MHz to allow
simultaneous 1H, 19F and 23Na imaging, and the
loop dimension was 165mm ×160mm with a width 10mm. The schematic diagram of
the triple-tuned circuit was shown in Fig. 2. C1, L1 and C4, L4 is used to tune
the frequency of 23Na, C2 and C3 participated in the tuning of 1H
and 19F, cooperating with the 1H/19F/23Na matching
circuit connected in series at the supply point to make the loop resonated at
three frequencies simultaneously. The acquired 1H/19F/23Na
signals acquired would be separated by the high-pass and low-pass filters added
on each receiving path.
To generate circularly polarized B1+ fields, all transceiver arrays were driven through
four output signals generated by the corresponding interface circuits with
equal magnitude but 90 degree phase increments. To realize a stable power
distribution of four channels per nuclide, the amplitude and phase
of excitation source in each channel were modulated by quadrature-hybrid
(microstrip structure for 1H/19F, LC circuit structure
for 23Na and 31P ) and pi phase shifter. The interface circuit of 1H and 19F
shared the same transceiver chain using a wide-band design, and was optimized
at 124.4 MHz, which can operate properly at 128.2MHz (1H) and 120.6MHz (19F). Two independent transceiver chains were built for 23Na
and 31P, which were optimized at 33.9MHz and 51.9MHz,
respectively. The decoupling between adjacent channels and next-adjacent
channels in the coil array was achieved by adjusting the overlap area. The
working state of all the nuclear was controlled by the detune circuits diode
connected in series in the loop, which reducing the interference between the
inner and outer coils.
S-parameters were measured on the bench using the network analyzer to
evaluate the match and isolation of the quadruple-nuclear transceiver coil
array system.RESULT
The 1H and 19F signals
were separated efficiently by 1H and 19F signals filter respectively, as shown
in Fig. 3 (a) and (b). As shown in Fig.3 (a), S21 was -0.85 dB at 128.2 MHz and
S21 was -23.26 dB at 120.6 MHz respectively, which indicate that only 1H
signal is allowed to pass through. As shown in Fig.3 (b), S21 was -32.65 dB at
128.2 MHz and S21 was -2.6 dB at 120.6 MHz, respectively, which indicate that only
19F signal is allowed to pass through.
The measured
s-parameters of the 1H/19F/23Na/31P coil were shown as Fig. 3 (c) and Fig. 3
(d), sufficient matching was achieved with the arithmetic maximum of reflection
coefficient no larger than -10 dB at 33.9 MHz, -20 dB at 120.6 MHz, 22 dB at
128.2 MHz and -14 dB at 51.9MHz MHz respectively. The designed triple-tuned circuit made the loop
resonated at three frequencies simultaneously and the reuse of coil used for 1H/19F/23Na
imaging was achieved, resulting in the elimination of the electromagnetic
interference between 1H/19F/23Na.
The relevant amplitude
and phase offset of power divider for 1H/19F frequencies were well balanced
with the absolute amplitude and phase offsets standard deviations were smaller
than 0.2 dB and 0.5 degree respectively (Table I).
Table II summarizes the isolation of T/R
switches for all channels. The sufficient isolation between transmission and
reception for all operating frequencies was achieved with the arithmetic
maximum of reflection coefficient no larger than -50 dB. Conclusion
In this
study, a quadruple-nuclear transceiver coil array was designed and constructed
for 1H/19F/23Na/31P imaging at 3T. Our
performance evaluations showed that the proposed coil array system was
sufficiently matched, detuned and decoupled when working simultaneously, thus
can be expected to generate high SNR images for simultaneous 1H/19F/23Na/31P. In future work, phantom and animal study
on a self-made 3T multinuclear MR system will be performed.Acknowledgements
1. Dong-Ho
Ha, Sunseob choi, Jong Young Oh, et al. Application of 31P MR Spectroscopy to the Brain Tumors. Korean J Radiol. 2013 May-Jun; 14(3): 477–486.
2. Madelin,
G, Regatte, R. R, et al. Biomedical applications of sodium MRI in vivo . J.
Magn. Reson. Imaging 2013;38:511–529
3. Straathof,
M., et al. "Deuterium Metabolic Imaging of the Healthy and Diseased Brain." Neuroscience 9(2021).
4. Feng Du, Shengping Liu,et al.
"Numerical Simulation and Evaluation of a Four-Channel-by-Four-Channel
Double-Nuclear RF Coil for 1HMRI and 31P MRSI at 7 T" IEEE Trans. Magn.,
vol. 54, no. 11, Nov. 2018, Art. No.
5101105
5. de,
Graaf Robin A., et al. " On the magnetic field dependence of deuterium
metabolic imaging." NMR Biomed 33.3(2020): e4235.
References
This work is supported by National Key Research and Development Program of China, 2021YFE0204400; NSFC grant 81627901; the Strategic Priority Research Program of Chinese Academy of Sciences, XDB25000000; National Natural Science Foundation of China, U22A20344; Youth Innovation Promotion Association of CAS No. Y2021098; Key Laboratory Project of Guangdong Province, 2020B1212060051; Shenzhen city grant, RCYX20200714114735123.