Sangwoo Kim1 and Sukhoon Oh2
1Daewon University College, Jecheon-si, Korea, Republic of, 2Korea Basic Science Institution, Ochang-eub, Korea, Republic of
Synopsis
Keywords: Safety, Thermometry
We showed a new approach of MR temperature mapping without the oil-based
phase reference of the conventional proton resonance frequency shift method
(cPRFS), named as delta field-corrected temperature mapping (DFCTM). The
performance of the DFCTM was compared with measurements using fiber optic
thermal sensor and the cPRFS at the human-tissue mimicking agar-gel phantom. The
sensor–measured temperature range was 6.05 ℃, while the DFCTM exhibited about
5.93 ℃. It exhibited a difference of -0.12 ℃ and the RMSE of the DFCTM was 0.16
℃ for the entire experiments, which means that it could realize a temperature
monitoring without oil phantoms.
Introduction
Proton resonance frequency shift (PRFS) is established as a simple and
rapid method with high spatial and temporal resolution than other methods such
as T1 relaxation time- or diffusion-based temperature mapping techniques1.
However, the MRI system phase variations caused by the eddy-current effect
affect the temperature reliability and fidelity1,2. It needs oil
phantoms or magnetic field monitoring probes to correct the phenomenon, whereas
the use of these supplies is quite cumbersome in the PRFS1-4. We
developed a new approach for temperature mapping without the oil phantom for
phase references on the PRFS process, named as delta field-corrected
temperature mapping (DFCTM).Methods
An agar-gel phantom to mimic human tissue was constructed with the agar
(8 g/L), NaCl (10 g/L), and CuSO4 (1 g/L). At 3T MRI (Archiva,
Philips Medical System, Best, Netherland), 3D dual-gradient echo (GRE) images
were acquired before and after the insertion of the fiber-optic thermal sensor without
phantom heating for the estimation of the susceptibility changes around the fiber
optic sensor. Subsequently, the agar-gel phantom was heated using a microwave
oven. Then, temperature images (PRFS) based on echo-planar imaging (EPI)
with dual echoes were acquired. All procedures for the DFCTM were shown at
Figure 1. Four oil phantoms were placed around the agar-gel phantom to estimate
the background phase drift for the conventional PRFS (cPRFS) in comparison of the
DFCTM. In addition, thermal sensor-measured temperature changes were monitored
as a reference. The susceptibility changes of the cPRFS and DFCTM around the fiber
optic sensor were lastly corrected. The scan parameters of 3D-GRE (for the
susceptibility measurements) and EPI (for PRFS) were summarized in Table 1.Results
We analyzed the root-mean-square error (RMSE) of DFCTM, oPRFS, and the
fiber optic sensor readings to determine the performance of the DFCTM. The
apparent patterns of temperature changes of DFCTM and oPRFS look different
(Fig. 2). However, DFCTM is much closer to the fiber optic sensor readings as
shown at Fig. 3. The sensor measured heating was up to 6.05 ℃ whereas the DFCTM
exhibited 5.93 ℃, which showed a difference of - 0.12 ℃ (cPRFS = + 0.31 ℃). The
RMSE of the DFCTM was 2.7 times lower than the cPRFS (0.16 ℃ vs. 0.43 ℃,
respectively).Discussion
This study showed the performance of temperature mapping of the DFCTM
(RMSE = 0.16 ℃), which is the reference phantoms-free PRFS method, in
comparison to the oPRFS and fiber optic thermal sensor. The sensitivity of the DFCTM
was within ± 1 ℃. This result indicates that the phase drift caused by the eddy
current and the phase retardation by large-volume heating could be corrected using
the magnetic field maps through dual-echo phase images. The PRFS needs oil
phantoms, which are fewer phase shifts, to correct phase drifts caused by eddy
currents that are varying according to any temperature1,2,4. Magnetic
field maps were generated through Hermitian products using dual echoes, which
can remove the phase retardation1. In the DFCTM process, each phase
difference map was calculated by subtracting different field maps, which were
acquired during temperature rises, from the first field map, and then the
difference maps were added to the field maps with temperature changes (Fig. 1).
These steps imply that phase drift caused by eddy currents in all temperature
ranges becomes similar to the first field map considerably. In addition, the
phase retardation remained only in the phase offset because it subtracted the
field maps from the original phase images, and the phase offset was subsequently
removed from the original data. This indicates that a new-found DFCTM enables
simultaneous correction of the phase drift and the phase retardation, leading
to the temperature reading from only phase information of the MRI. Therefore,
we believe that the DFCTM method can be applied to most MR thermometry studies.Acknowledgements
This work was supported by the Institute for
Information and Communications Technology Promotion (IITP) Grant by the Korean
Government through the Ministry of Science, ICT and Future Planning (MSIP)
(Development of precision analysis and imaging technology for biological radio
waves) under Grant 2021-0-00490.References
1. Winter L, Oberacker E, Paul K, Ji Y,
Oezerdem C, Ghadjar P, et al. Magnetic resonance thermometry: Methodology,
pitfalls and practical solutions. Int J Hyperthermia. 2016;32(1):63-75.
2. Rieke V. MR thermometry. J. Magn.
Reson. Imaging. 2008;27:376-390.
3. Hernandez D, Kim KS, Michel E, Lee
SY. Correction of B0 drift effects in magnetic resonance thermometry using
magnetic field monitoring technique. Concepts Magn Reson Part B (Magn Reson Engineering).
2016;46B(2):81-89.
4. Ferrer CJ, Bartels LW, van
der Velden TA, Grüll H, Heijman E, Moonen CT, et al. Field drift correction of
proton resonance frequency shift temperature mapping with multichannel fast
alternating nonselective free induction decay readouts. Magn Reson Med.
2020;83(3):962-973.