Charlotte Rogers1, Ronald Mooiweer1,2, Grzegorz Kowalik1, Radhouene Neji1,2, Reza Razavi1, Rene Botnar1,3,4, and Sebastien Roujol1
1School of Biomedical Engineering and Imaging Sciences, Faculty of Life Sciences and Medicine, King's College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare Limited, Camberley, United Kingdom, 3School of Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile, 4Institute for Biological and Medical Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile
Synopsis
Keywords: Myocardium, Thermometry
A novel fast simultaneous multi-slice black
blood cardiac thermometry sequence is presented which allows multiple slices to be acquired in the same cardiac phase.
A double inversion recovery pre-pulse is introduced to allow blood suppression
and to optimise the pre-pulse strategy and therefore reduce acquisition time. Simultaneous
multi-slice (SMS) imaging is integrated to further reduce the acquisition time
of the multiple slices. Thermometry results obtained with this technique show
an average temperature stability of 1.0±0.4°C in healthy subjects.
Introduction
Catheter based ablation therapy has
become a well-established treatment option for patients suffering from cardiac arrhythmias
and is used to block abnormal electrical signals in the diseased tissue. MRI-guidance
of such procedures represents a promising approach to improve procedural
outcome, which currently has a high recurrence rate of up to 50% 1. MRI
enables real-time temperature mapping in the heart and prediction of the extent
of the ablation lesions 2,3. Current cardiac MR-thermometry methods cover the
ablation region by acquiring multiple contiguous slices (usually Nslices=3-4)
per heartbeat. These acquired slices, however, require pre-pulses such as fat
suppression and saturation bands for in-flow blood saturation as well as reduced
field of view imaging, in combination with a fast EPI readout. This leads to an
acquisition time of around 300-400ms 4 per stack of slices, meaning that they
are all acquired at different cardiac phases. Lesion assessment in the
through-plane dimension is therefore severely limited 3, which is an
important restriction of the method. This could lead to inadequate lesion
formation which is known to cause a higher risk of developing recurrent
arrhythmias following the procedure5.
In this study, we sought to
address these challenges by developing a fast single-phase volumetric cardiac MR-thermometry
sequence. This was done by using simultaneous multi-slice (SMS) sequence combined
with a more time efficient pre-pulse strategy for blood signal suppression.
Methods
The proposed cardiac MR-thermometry
sequence is shown in Figure 1. Four slices were acquired using an SMS
single-shot EPI sequence with a multiband factor of 2. A double inversion
recovery (DIR) pre-pulse was implemented for blood signal suppression instead
of the typical in-flow saturation slabs that are commonly played before each
slice. This DIR pre-pulse (thickness of the slice-selective re-inversion pulse=40mm, delay (Td)
between DIR pulses and first SMS shot=225ms) therefore reduced the pre-pulse
time taken before each slice whilst also delivering good dark blood signal
suppression. Fat saturation pulses were used in both sequences prior to each
SMS-EPI shot and saturation slabs were added to allow for a reduced field of
view (FOV).
This new prototype sequence was tested in five
healthy subjects (4m, 1f) scanned on a 1.5T scanner (MAGNETOM Aera, Siemens
Healthcare, Erlangen Germany) under free breathing. The original thermometry sequence was also
tested in two of the healthy subjects to be able to compare the magnitude
images (shown in Figure 2). The common parameters for all scans were: TE=24ms, TR=63ms, Flip
Angle=60o, Bandwidth=1420 Hz/px, FOV=250x250mm2,
in-plane voxel size=3.1x3.1mm2, slice thickness=5mm, Nslices=4,
multiband factor=2, number of dynamics=150, no slice gap. A multi-baseline approach
using a look-up table of 40 dynamics, serving as a reference phase, was then used to reconstruct the motion
corrected temperature maps offline. This was done for
correction of respiratory induced phase variation together with non-rigid image
registration, and temporal temperature filtering, as previously described 6. The stability of thermometry was assessed by calculating
the standard deviation of the temperature in the myocardium over time. The
stability in all 4 slices was combined and reported for each subject.
Results
Successful blood signal suppression was achieved
in all slices using the DIR pre-pulses, as can be seen in Figure 2 which
shows typical images obtained for two subjects. Figure 3 displays the
temperature map of the same two subjects over the 4 slices. The data from the
subject shown in Figure 3b showed slightly better results with constant
temperature stability values across the whole myocardium, averaging at under 1°C. The temperature map shown in Figure
3a represents the worst stability values obtained across the volunteers,
averaging at around 1.4°C. The distribution of thermometry stability
over the 4 slices of each volunteer is shown in the box plot in Figure 4.
Over all the subjects, the thermometry stability was calculated to be 1.0±0.4°C.Discussion
The proposed approach demonstrates the feasibility of
single-phase cardiac MR-thermometry with a precision of ~1°C. The blood
signal suppression obtained in this study allows better visualisation of the
morphologic features of the heart, which is important to improve the
performance of the applied motion correction algorithm as well as the
temperature stability 7. Only a limited number of subjects
were included in this study and assessment of this technique on a larger cohort
is warranted to confirm these findings. Since all the slices are acquired in the same cardiac phase, this
technique has the potential to improve 3D assessment of ablation lesions, but
further evaluation will be required during in-vivo ablation. Future work will
also involve increasing in-plane spatial resolution and correcting for
through-plane motion between heartbeats. Conclusion
The method proposed in this study allows
multiple (Nslices=4) contiguous slices per heartbeat to be acquired
in the same cardiac phase. This technique enables good blood signal suppression, and good thermometry
stability of 1.0±0.4°C. The
benefit of this technique for 3D lesion assessment will be the subject of
future studies.
Acknowledgements
This
work was supported by the Engineering and Physical Sciences Research Council
(EPSRC) grants (EP/R010935/1), the Centre for
Doctoral Training in Surgical and Interventional Engineering funded by King’s College London’s Centre for Doctoral Studies, the British Heart Foundation (BHF) grants (PG/19/11/34243
and PG/21/10539), the Wellcome EPSRC Centre for Medical Engineering at King’s
College London (WT 203148/Z/16/Z), the National Institute for Health Research
(NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ National Health
Service (NHS) Foundation Trust and King’s College London. The views expressed
are those of the authors and not necessarily those of the NHS, the NIHR or the
Department of Health.References
1. Kuck K. H., Schaumann A,
Eckardt L, Willems S, Ventura R, Delacretaz E, et al. Catheter ablation of
stable ventricular tachycardia before defibrillator implantation in patients
with coronary heart disease (VTACH): a multicentre randomised controlled trial
Lancet (2010) 375:9708
2. Denis de Senneville B, Roujol S, Jais P, Moonen C, Herigault G and
Quesson B. Feasibility of fast MR-thermometry during cardiac radiofrequency
ablation NMR Biomed (2012) 25:556-62
3. Toupin S, Bour P,
Lepetit-Coiffe M, Ozenne V, Denis de Senneville B, Schneider R et al. Feasibility of real-time MR thermal dose mapping for
predicting radiofrequency ablation outcome in the myocardium in vivo. J
Cardiovas Magn Reson (2017) 19:14
4. Ozenne V, Toupin S, Bour P,
D. de Senneville B, Lepetit-Coiffe M, Boissenin M, et al.
Improved cardiac magnetic resonance thermometry and dosimetry for
monitoring lesion formation during catheter ablation Magn Reson Med (2017) 77:673-683 DOI: 10.1002/mrm.26158.
5. Sacher F, Tedrow U. B., Field E, Raymond J. M., Koplan B.
A., Epstein L. M., et al. Ventricular tachycardia ablation: evolution of
patients and procedures over 8 years. Circ Arrhythm Electrophysiol (2008)
1:3
6. Roujol S, Ries M, Quesson B, Moonen C, Denis de Senneville B. Real-time
MR-thermometry and dosimetry for interventional guidance on abdominal organs.
Magn Reson Med. (2010) 63:1080–7.
7. Hey S, Cernicanu A, de
Senneville B. D., Roujol S, Ries M, Jais P, Moonen C.T.W and Quesson B. Towards
optimized MR thermometry of the human heart at 3T. NMR Biomed (2011) 25:1