Suzanne M Wong1,2, Craig A Macsemchuk1,2, Andrew Headrick1, Phoebe Luo1, Arthur Akbulatov1, James M Drake1,2, and Adam C Waspe1,3
1Posluns Centre for Image Guided Innovation & Therapeutic Interventions, Hospital for Sick Children, Toronto, ON, Canada, 2Department of Biomedical Engineering, University of Toronto, Toronto, ON, Canada, 3Department of Medical Imaging, University of Toronto, Toronto, ON, Canada
Synopsis
Keywords: Motion Correction, Thermometry
Magnetic resonance-guided
high-intensity focused ultrasound (MRgHIFU) can noninvasively administer
controlled hyperthermia as an adjuvant cancer therapy. For clinical
translation, one of the main challenges is the sensitivity of MR thermometry to
motion artifacts. This work aims to validate a real-time hybrid principal
component analysis and projection onto dipole fields (PCA-PDF) motion compensation
algorithm on a clinical MRgHIFU system during reproducible motion profiles. The
real-time PCA-PDF algorithm maintained a temperature standard deviation of <
1°C in a phantom while the periodic motion was induced on a phantom using an
MR-compatible robot.
Introduction
Standard-of-care
cancer therapies, such as radiation and chemotherapy, can have improved
efficacy when combined with mild hyperthermia in more resistant tumour types1-3.
One localized, non-invasive method of administering mild hyperthermia is
magnetic resonance-guided high-intensity focused ultrasound (MRgHIFU). MRI is
used for treatment planning and monitors the treatment with magnetic resonance
thermometry, which maps temperature changes overlaid on the anatomical images.
However, one of the main challenges of MR thermometry is that it is highly sensitive
to motion artifacts, which can skew temperature measurements4.
Real-time
motion compensation algorithms can reduce temperature uncertainty during
treatment by removing any artifacts that appear in the thermometry when motion
occurs. The motion artifact removal algorithm used in this work is a hybrid of
principal component analysis and projection onto dipole fields (PCA-PDF)5.
The objective of this study was to validate the capabilities of the PCA-PDF
motion compensation algorithm under a controlled periodic motion profile for
hyperthermia on a clinical MRgHIFU system.Methods
The
motion compensation validation was done on a clinical MRgHIFU system consisting
of a 3T Achieva MRI (Philips Healthcare, Best, NL) and Sonalleve V1 HIFU
(Profound Medical, Toronto, Canada) system. For the MR thermometry, an
echo-planar fast-field echo imaging (EPI-FFE) sequence was used. A custom
closed-loop hyperthermia software called Proteus6 was used for
implementing the PCA-PDF algorithm.
An MR-compatible robot was
constructed to allow movement across the right-left (RL) and superior-inferior
(SI) directions while not impeding ultrasound propagation. To recreate periodic
motion, a sinusoidal movement with a period of 4s and an amplitude of 10mm was
used, which was found to be the general movement of the diaphragm in healthy
patients7. This robot was designed to induce motion on custom
phantoms fabricated using 10% gelatin and 2% silica (Figure 1). This
formulation was ideal to simulate a tissue-like stiffness which shows a speed
of sound to 1540m/s, the average speed of sound through human tissue and bulk
attenuation of 1 dB/cm/MHz8.
In initial tests, no sonications
were performed, to remove all temperature-dependent phase changes. In doing so,
we can conclude the thermal signal present during the motion to be solely due
to motion artifacts. During the atlas acquisition of 20 images, the robot moved
at the same periodic specifications as during the MR thermometry sequence. No
motion was induced for the first 2 minutes once the thermometry started, to act
as a baseline. For the following 2 minutes, the periodic motion was induced on
the phantom with the robot within the MRgHIFU system. After, an additional 2
minutes of the thermometry sequence was acquired without movement. This was
performed 5 times without motion compensation where the PCA-PDF algorithm would
be run retrospectively and 5 times with the real-time PCA-PDF algorithm. These
protocols were repeated with hyperthermia administered by the MRgHIFU system.
To analyze the thermometry data, the standard deviations of temperature
measurements were calculated for each voxel over the thermometry images when
motion was induced on the phantom. A paired t-test was run to compare the
average temperature standard deviation between the original subtraction
thermometry and both the retrospective and real-time PCA-PDF.Results
The spatial average temporal
standard deviation in the thermometry data without motion compensation was
20.9°C ± 1.4°C. After the retrospective application of the PCA-PDF algorithm to
these data sets, a substantial amount of motion artifacts were removed (Figure
2), similar to the trials where the PCA-PDF algorithm was run in real-time. The
average temperature standard deviation with the retrospective PCA-PDF was
reduced to 0.5°C ± 0.03°C. This is comparable to the
real-time PCA-PDF test, which had an average temperature standard deviation of
0.4°C ± 0.05°C (Figure 3). The PCA-PDF algorithm improved MR
thermometry by reducing temperature standard deviation with statistical
significance (p<0.001).Discussion
We can create reproducible periodic motion by using a robot positioner for the phantom. Preliminary results
demonstrated that the PCA-PDF algorithm compensated for periodic motion,
significantly reducing motion artifacts. When hyperthermia was administered,
the PCA-PDF algorithm was able to distinguish temperature-dependent phase
changes from motion-induced phase changes. Even when substantial motion
artifacts were present in the magnitude MRIs, the PCA-PDF algorithm was able to
compensate successfully (Figure 2C). There was a decrease in temperature
standard deviation from 20.9°C with the original subtraction thermometry method
to <1°C with the retrospective and real-time PCA-PDF methods. These results
verified that the PCA-PDF algorithm is comparable in both retrospective and
real-time modes, confirming that we can rely on the results during treatment.
Conclusion
The results demonstrate that the
PCA-PDF algorithm's real-time capabilities were comparable to retrospective
analysis, where the algorithm reduced temperature uncertainty to <1°C.
Results also verified that the algorithm successfully distinguished heating
from motion artifacts. Future work includes expanding the types of motion
profiles to recreate more sporadic or arbitrary movements and further
validation of the PCA-PDF algorithm. Acknowledgements
We thank Samuel
Pichardo for providing access to the Proteus software and technical assistance.
Funding is provided by the Natural Sciences and Engineering Research Council
(NSERC).References
1. Datta NR, Puric E, Klingbiel D,
Gomez S, Bodis S. Hyperthermia and Radiation Therapy in Locoregional
Recurrent Breast Cancers: A Systematic Review and Meta-analysis. Int J Radiat Oncol Biol Phys.
2016;94(5):1073-1087. doi:10.1016/j.ijrobp.2015.12.361
2. Datta NR, Rogers S, Klingbiel
D, Gómez S, Puric E, Bodis S. Hyperthermia and radiotherapy with or
without chemotherapy in locally advanced cervical cancer: a systematic
review with conventional and network meta-analyses. Int J Hyperth. 2016;32(7):809-821.
doi:10.1080/02656736.2016.1195924
3. Issels RD, Lindner LH, Verweij
J, et al. Effect of neoadjuvant chemotherapy plus regional hyperthermia on
long-term outcomes among patients with localized high-risk soft tissue
sarcoma the EORTC 62961-ESHO 95 randomized clinical trial. JAMA Oncol. 2018;4(4):483-492.
doi:10.1001/jamaoncol.2017.4996
4. Rieke V, Pauly KB. MR
thermometry. J Magn Reson Imaging.
2008;27(2):376-390. doi:10.1002/jmri.21265
5. Tan
J, Mougenot C, Pichardo S, Drake JM, Waspe AC. Motion compensation using
principal component analysis and projection onto dipole fields for
abdominal magnetic resonance thermometry. Magn Reson Med. 2019;81(1):195-207. doi:10.1002/mrm.27368
6. Zaporzan B, Waspe AC, Looi T,
Mougenot C, Partanen A, Pichardo S. MatMRI and MatHIFU: Software toolboxes
for real-time monitoring and control of MR-guided HIFU. J Ther Ultrasound. 2013;1(1).
doi:10.1186/2050-5736-1-7
7. Gorman
RB, McKenzie DK, Pride NB, Tolman JF, Gandevia SC. Diaphragm length during
tidal breathing in patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med.
2002;166(11):1461-1469. doi:10.1164/rccm.200111-087OC
8. Ryan LK, Foster FS. Tissue
equivalent vessel phantoms for intravascular ultrasound. Ultrasound Med Biol.
1997;23(2):261-273. doi:10.1016/S0301-5629(96)00206-2