Sven Nouwens1, Maarten Paulides2,3, and Maurice Heemels1
1Mechanical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands, 2Electrical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands, 3Radiotherapy, Erasmus MC, Rotterdam, Netherlands
Synopsis
Keywords: Sparse & Low-Rank Models, Thermometry, susceptiblity aritfact correction
Proton
resonance frequency shift-based MR thermometry is widely used to non-invasively
monitor thermal therapies in vivo. However, further clinical integration in
deep hyperthermia is hampered by intestinal air-motion induced susceptibility
artifacts. We developed a sparse regression approach to delineate
susceptibility artifact sources. The resulting mask is then used to correct the
artifact using existing methods from quantative susceptibility mapping. We
verified our approach by a heated phantom experiment equipped with a moveable
air volume and temperature probes. Here, we found a reduction in the mean
absolute error from 1.6 degrees Celsius to 0.4 degrees Celsius, near the
air-motion artifact.
Introduction
Proton
resonance frequency shift (PRFS) MR thermometry is widely used to
non-invasively monitor thermal therapies in vivo. However, (air-)motion
artifacts frequently corrupt the measurements, hampering further integration of
PRFS thermometry into clinical practice. Recently, computational methods from Quantitive
Susceptibly Mapping were proposed to correct for air-motion artifacts1.
However, these techniques, e.g., the PDF method2, can, in cases, also
inadvertently remove the temperature induced phase shift, in addition to the
motion artifact. Recent work aims to alleviate this problem by exploiting either
model-based temperature estimators or total field inversion3, 4.
Regardless, delineating susceptibility sources, i.e., mask selection, remains a
crucial step in the artifact removal process. Mask selection based on
anatomical scans is straightforward but is prone to errors as not all regions with
a low MR signal correspond to air motion. We propose to compute a mask directly
from the measured temperature using sparse regression. More specifically, we
compute the mask by exploiting the knowledge that there are limited number of
distinct susceptibility sources into a regularized optimization problem. In other
words, we can exploit our prior knowledge regarding air-motion artifacts to
separate temperature induced phase shifts from those caused by susceptibility artifacts.
We verified our approach using a heated phantom that is equipped with a
moveable air volume and temperature probes, see Figure 1 and for more details3.Method
We model
the measured (B0-drift corrected) PRFS thermometry as
$$T_{prfs}(r) = T(r) + D(r)*\Delta\chi(r).\tag{1}$$ Here, $$$T_{prfs},\ T,\ D,\ \Delta\chi$$$, and $$$*$$$ denote the
measured temperature, the true temperature, a dipole kernel1,2, the magnetic
susceptibility difference with respect to the baseline scan, and the
convolution operator, respectively. Recall that we assume $$$\Delta\chi$$$ is sparse,
i.e., there are only a few locations where the magnetic susceptibility changed
with respect to the baseline scan.
The PDF method estimates
susceptibility sources by solving the following optimization problem2,
$$\Delta\chi_{PDF}(r)=\arg\min_{\Delta\chi\in\Omega_{mask}}\int_{\Omega_{patient}}\big(T_{prfs}(r)-T_{est}(r)-D(r)*\Delta\chi(r)\big)^2dV.\tag{2}$$ Here, $$$T_{est}$$$ can be used to remove
prior (model-based) temperature information from the PRFS measurement1,3.
For simplicity, we choose $$$T_{est}=0$$$ for the
remainder of this paper. Note that in (2), sparsity is enforced by constraining
$$$\Delta\chi$$$ to the mask $$$\Omega_{mask}$$$. This naturally leads to the following tradeoff: If
the mask is too large, we risk removing the temperature induced phase shift. If
the mask is too small, we do not remove the motion artifact completely.
In this work, we compute
the mask $$$\Omega_{mask}$$$ using sparse
regression,
$$\Delta\hat{\chi}(r)=\arg\min_{\Delta\chi}\int_{\Omega_{patient}}\big(T_{prfs}(r)-D(r)*\Delta\chi(r)\big)^2dV+\lambda\int_{\Omega_{patient}}|\Delta\chi(r)|dV.\tag{3}$$ From (3), we then
compute $$$\Omega_{mask}$$$ as $$$\Omega_{mask}=\{r\in\mathbb{R}^3\mid|\Delta\hat{\chi}(r)|\geq threshold\}$$$. Note that, $$$\lambda$$$ controls the
sparsity of the resulting mask estimate. We solve (3) efficiently using SR35.
It may seem strange at first not to use the susceptibility estimate from (3)
directly. However, it is known that sparse regression is excellent at finding
the support of a function, but not necessarily well-suited to computing the
best estimate5. For this reason, we threshold the minimizer (3) above
the noise floor, and then, use the obtained mask in (2) to compute the final
estimated susceptibly source. Finally, the corrected temperature is given by $$$T_{corrected}(r) = T_{prfs}(r)-D(r)*\Delta\chi_{PDF}(r)$$$.
The method is
verified using a heated
phantom equipped with a moveable air volume and temperature probes, see Figure 1 and for more details3.
Results
In Figure 2,
we show the anatomical scan, PRFS measurement, identified mask, and artifact
corrected temperature. Indeed, visually, sparse regression in combination with
the PDF method appears to be successful at removing the air-motion artifact (indicated
by the arrow). In Figure 3 and Table 1, we quantitively verify the artifact
correction scheme using four temperature probes that serve as a ground truth. Here,
the mean absolute error (MAE) at probe 1 decreased from 1.6 degrees Celsius to 0.4
degrees Celsius, indicating the susceptibility artifact is removed. Crucially,
the MAE at the remaining (uncorrupted) probes remains similar, indicating that
the temperature induced phase shift is not removed by our method. Conclusion
In this
work, we showed that sparse regression techniques are effective at computing
masks for air-motion correction. By computing the mask directly from the
measured thermometry, this method avoids thresholding anatomical scans, which
can be problematic. We demonstrated the efficacy of this method on a heated
phantom experiment equipped with temperature probes and a moveable air volume.
Here, we observed that the air motion artifact was successfully removed, while
uncorrupted regions were not affected by our post-processing.Acknowledgements
This research is supported by KWF Kankerbestrijding and NWO Domain AES, as part of their joint strategic research programme: Technology for Oncology II. The collaboration project is co-funded by the PPP Allowance made available by Health$$$\sim$$$Holland, Top Sector Life Sciences & Health, to stimulate public-private partnerships.References
1. Wu, M., Mulder, H. T., Baron, P., Coello,
E., Menzel, M. I., van Rhoon, G. C., & Haase, A. (2020). Correction of motion‐induced susceptibility artifacts and B0 drift during proton resonance
frequency shift‐based MR thermometry in the pelvis
with background field removal methods. Magnetic resonance in medicine, 84(5),
2495-2511.
2. Liu,
T., Khalidov, I., de Rochefort, L., Spincemaille, P., Liu, J., Tsiouris, A. J.,
& Wang, Y. (2011). A novel background field removal method for MRI using projection
onto dipole fields. NMR in Biomedicine, 24(9), 1129-1136.
3. Nouwens,
S. A. N., Paulides, M. M., Fölker, J., VilasBoas-Ribeiro, I., de Jager, B.,
& Heemels, W. P. M. H. (2022). Integrated thermal and magnetic
susceptibility modeling for air-motion artifact correction in proton resonance
frequency shift thermometry. International Journal of Hyperthermia, 39(1),
967-976.
4. Boehm,
C., Goeger‐Neff, M., Mulder, H. T., Zilles, B.,
Lindner, L. H., van Rhoon, G. C., ... & Wu, M. (2022). Susceptibility
artifact correction in MR thermometry for monitoring of mild radiofrequency
hyperthermia using total field inversion. Magnetic Resonance in Medicine,
88(1), 120-132.
5. Zheng,
P., Askham, T., Brunton, S. L., Kutz, J. N., & Aravkin, A. Y. (2018). A
unified framework for sparse relaxed regularized regression: SR3. IEEE
Access, 7, 1404-1423.