David Lohr1, Wiebke Schlötelburg2, Maxim Terekhov1, and Laura Maria Schreiber1
1Chair of Molecular and Cellular Imaging, Comprehensive Heart Failure Center (CHFC), University Hospital Würzburg, Würzburg, Germany, 2Department of Nuclear Medicine, University Hospital Würzburg, Würzburg, Germany
Synopsis
Keywords: Myocardium, High-Field MRI
In this study we
aim to analyze blood tissue contrast (BTC), SNR, and reproducibility in
measurements of cardiac function at 7T. Furthermore, we assess shim quality and
its impact on image quality as well as T
2*. Measurements are performed in n=7 healthy volunteers using a 7T system.
Diastolic BTC and SNR in scan1 and scan2
were 2.14±0.24 and 2.22±0.33 as well as 96±35 and 105±29, respectively.
Reproducibility assessed via intra class correlation coefficient was found to
be excellent for ejection fraction (0.93), myocardial mass (0.99), stroke
volume (0.91), end-systolic volume (0.94), and end-diastolic volume (0.93).
Itroduction
Cardiac magnetic
resonance (CMR) imaging at 1.5 and 3T is an established technique and the gold
standard for the assessment of cardiac function and ventricular mass in
clinical routine1. While CMR studies in humans at 7T have shown general feasibility of
cardiac function2-6 and increased SNR7, it has not been demonstrated that these metrics can
be consistently derived with reasonable accuracy. In this study we thus
aim to analyze blood tissue contrast (BTC), SNR, and reproducibility in measurements
of cardiac function at 7T. Furthermore, we assess shim quality and its impact
on image quality as well as T2*.Methods
All experiments were approved by the
local ethics committee and written informed consent was given by each volunteer
prior to MRI measurements. Data was acquired in n=7 volunteers (n=6 female)
with body weight: 51-91kg and height: 161-185cm.
All MRI measurements are
performed using a 7T MAGNETOM™ Terra system (Siemens Healthineers, Erlangen)
and a novel cardiac transceiver array (Rapid Biomedical, Rimpar)8 comprising an
anterior and a posterior component. The array was used in pTX mode with phase
and amplitude settings as provided by the vendor.
The MRI protocol is illustrated
in Figure 1. After completion of the measurement protocol, the patient table
was moved to home position, the anterior array removed and volunteers were asked to leave the patient table for a duration of ~5 minutes. Afterwards,
volunteers were asked to lay down on the patient table again and all hardware
was reconnected. Steps 1-3 of the MRI protocol were performed again (due to
time constraints B0/T2* was acquired in
only n=6 volunteers (n=5 female).
Sequence parameters for cine acquisition
were: TE/TR: 0.7/~45ms, echo-spacing: 7ms, bandwidth: 915Hz/Px, FOV: 400x378mm2,
interpolated in-plane resolution: 0.7x0.7mm2, 30 cardiac phases.
Sequence parameters for the
multi-echo gradient echo sequence were TE: [1.5, …, 15]ms, TR: 328ms, FA: 40°, bandwidth:
1955Hz/Px, FOV: 360x300 mm2, interpolated in-plane resolution:
0.7x0.7 mm2.
Cardiac function for scan1 and
scan2 were evaluated three days apart by the same observer using Medis Suite MR
(Medis, Leiden). SNR9 and blood tissue
contrast based on respective Medis contours as well as data analysis with respect
to B0 (ROMEO10 software) and T2*
was done using Matlab (Mathworks, Natick).Results
The average flip angle selected
as optimal for subsequent CINE scans was 21±3°. Short axis cine data as
depicted in Figure 1 & Figure 2A showed good BTC. The observed increase in
blood pool signal from base to apex was present in all scans.
Average BTC values of the LV for
scan1 and scan2 are listed in Table 1 and individual values for all scans are
plotted in Figure 2B. Overall BTC was consistent between scans. While there was
a statistically significant difference in systolic BTC between scan1 and scan2,
the mean difference was only 0.22. Myocardial SNR is plotted in Figure 2C. No
significant differences were found between scan1 and scan2.
Figure 3A shows representative B0
maps of a single volunteer for the three different shim
volumes applied. Shimming based on a slab volume appears to reduce peak B0
values, improving B0 homogeneity in the left ventricle. Similar observations
were made for the slice volume, but less consistently. The different shim
volumes had no significant impact on mean T2* values
(Figure 3B-C). However, with p=1.5*10-3
(paired t-test), overall T2* was found to be
significantly lower in diastole (13.8±1.55ms) than
in systole (15.1±0.9ms).
Figure 4a depicts representative
SA CINE images of both scans for one volunteer. Anatomical details in both
acquisitions are very similar (yellow arrows), while slight differences in B1
distribution (blue arrows) are present after repositioning the volunteer and
the coil prior to the second scan. Correlation plots (Figure 4B-D) as well as
Pearson correlation coefficients, and ICCs (Table 1) demonstrate excellent
agreement between metrics of cardiac function for all volunteers. No
significant differences were found between metrics derived from scan1 and
scan2.Discussion
In this study we
show that 7T CMR using a novel transceiver array provides CINE images with
consistent BTC and SNR. Compared to literature
reports at 1.5T (1.5±0.4 at end-diastole and 1.4±0.3 at systole), we
demonstrate markedly improved BTC (2.14 ± 0.24 at end-diastole, 1.80 ± 0.13 at
systole).11 As indicated by ICCs>0.9,12 metrics of
cardiac function showed excellent agreement between scan and rescan, despite
minor B0 and B1 related artifacts. With n=7 the current
data set is rather small and more measurements will be required to demonstrate
that our results are valid for a broad spectrum of volunteers and patients.
Our
results regarding B0 shimming and the measurement of T2*
suggest that the shimming volume ought to be a slab around the measured slice. Further
measurements will be required to assess the optimal thickness of this slab.
Diastolic and systolic T2* values (ratio=1.09) are
comparable to septal diastolic (13.7±1.1 ms) and systolic (15.0±2.1) T2*
values (ratio=1.12) reported at 7T13, indicating that T2*
analysis does not have to be limited to the septum. Conclusion
We demonstrate that 7T CMR
enables high reproducibility in the assessment of cardiac function and
ventricular mass in subsequent measurements.Acknowledgements
LM Schreiber receives research
support by Siemens Healthineers. The position of D. Lohr is partially funded by
this research support.References
1. Ponikowski P, Voors AA, Anker SD,
Bueno H, Cleland JGF, Coats AJS, Falk V, González-Juanatey JR, Harjola V-P,
Jankowska EA, Jessup M, Linde C, Nihoyannopoulos P, Parissis JT, Pieske B,
Riley JP, Rosano GMC, Ruilope LM, Ruschitzka F, Rutten FH, van der Meer P,
Group ESD. 2016 esc guidelines for the diagnosis and treatment of acute and
chronic heart failure: The task force for the diagnosis and treatment of acute
and chronic heart failure of the european society of cardiology (esc)developed
with the special contribution of the heart failure association (hfa) of the
esc. European Heart Journal. 2016;37:2129-2200
2. Stäb D, Al Najjar A, O'Brien K,
Strugnell W, Richer J, Rieger J, Niendorf T, Barth M. Cardiac magnetic
resonance imaging at 7 tesla. JoVE. 2019:e55853
3. Reiter
T, Lohr D, Hock M, Ankenbrand MJ, Stefanescu MR, Kosmala A, Kaspar M, Juchem C,
Terekhov M, Schreiber LM. On the way to routine
cardiac mri at 7 tesla - a pilot study on consecutive 84 examinations. PloS
one. 2021;16:e0252797
4. Ibrahim EH, Arpinar VE, Muftuler
LT, Stojanovska J, Nencka AS, Koch KM. Cardiac functional magnetic resonance
imaging at 7t: Image quality optimization and ultra-high field capabilities.
World journal of radiology. 2020;12:231-246
5. Prothmann M, von
Knobelsdorff-Brenkenhoff F, Töpper A, Dieringer MA, Shahid E, Graessl A, Rieger
J, Lysiak D, Thalhammer C, Huelnhagen T, Kellman P, Niendorf T, Schulz-Menger
J. High spatial resolution cardiovascular magnetic resonance at 7.0 tesla in
patients with hypertrophic cardiomyopathy – first experiences: Lesson learned
from 7.0 tesla. PloS one. 2016;11:e0148066
6. von Knobelsdorff-Brenkenhoff F,
Frauenrath T, Prothmann M, Dieringer MA, Hezel F, Renz W, Kretschel K, Niendorf
T, Schulz-Menger J. Cardiac chamber quantification using magnetic resonance
imaging at 7 tesla—a pilot study. European Radiology. 2010;20:2844-2852
7. Niendorf T, Schulz-Menger J, Paul
K, Huelnhagen T, Ferrari VA, Hodge R. High field cardiac magnetic resonance
imaging. Circulation: Cardiovascular Imaging. 2017;10:e005460
8. Terekhov
M, Lohr D, Reiter T, Elabyad IA, Hock M, Schreiber LM. New commercial 8tx/16rx
array for Clinical 7T Cardiac
mri: Initial experience. Proc. of the 29th Annual Meeting of ISMRM. 2021:1584
9. Constantinides CD, Atalar E,
McVeigh ER. Signal-to-noise measurements in magnitude images from nmr phased
arrays. Magnetic resonance in medicine. 1997;38:852-857
10. Dymerska B, Eckstein K, Bachrata B,
Siow B, Trattnig S, Shmueli K, Robinson SD. Phase unwrapping with a rapid
opensource minimum spanning tree algorithm (romeo). Magnetic resonance in
medicine. 2021;85:2294-2308
11.T hiele H, Nagel E, Paetsch I, Schnackenburg
B, Bornstedt A, Kouwenhoven M, Wahl A, Schuler G, Fleck E. Functional cardiac
mr imaging with steady-state free precession (ssfp) significantly improves
endocardial border delineation without contrast agents. Journal of Magnetic
Resonance Imaging. 2001;14:362-367
12. Koo TK, Li MY. A guideline of
selecting and reporting intraclass correlation coefficients for reliability
research. Journal of Chiropractic Medicine. 2016;15:155-163
13. Huelnhagen
T, Ku M-C, Reimann HM, Serradas Duarte T, Pohlmann A, Flemming B, Seeliger E,
Eichhorn C, A. Ferrari V, Prothmann M, Schulz-Menger J, Niendorf T. Myocardial
effective transverse relaxation time t2* is elevated in hypertrophic
cardiomyopathy: A 7.0 t magnetic resonance imaging study. Scientific Reports.
2018;8:3974