Rohini Vidya Shankar1, Li Huang1, Radhouene Neji1,2, Grzegorz Kowalik1, Alexander Paul Neofytou1, Ronald Mooiweer1,2, Tracy Moon3, Nina Mellor3, Reza Razavi1, Kuberan Pushparajah1, and Sébastien Roujol1
1Biomedical Engineering, King's College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare Limited, Camberley, United Kingdom, 3Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom
Synopsis
Keywords: Heart, MR-Guided Interventions, Tracking, Real-time
MR-guided cardiac catheterization procedures
currently employ passive tracking approaches to follow the gadolinium-filled
catheter balloon during catheter navigation. This requires frequent manual
tracking and repositioning of the imaging slice, especially when the catheter
moves out-of-plane during the procedure. In this study, we developed a novel
MRI guidance approach that enables automatic real-time tracking of the catheter
balloon and repositioning of the imaging slice for continuous visualization of
the balloon during catheter navigation. We first demonstrate the proposed
approach in a phantom and subsequently present an initial evaluation in
patients.
Introduction
MRI is an attractive alternative to X-ray
fluoroscopy for the guidance of cardiac catheterization procedures
due to high soft tissue contrast, superior hemodynamic data, and lack of ionizing
radiation.1–3 Current passive tracking approaches using gadolinium
(Gd)-filled balloon wedge catheters require frequent manual slice
tracking and manipulation of the imaging plane to follow the catheter during
catheter navigation. It was previously reported that the catheter was
out-of-plane in >30% of dynamic frames during navigation.4 In
this study, we present further technical developments to a novel cardiac MRI
sequence5 that enables automatic real-time tracking and
visualization of the balloon during catheter navigation. The proposed technique
is first evaluated in a phantom and subsequently demonstrated in patients
undergoing MR-guided cardiac catheterization.Methods
(1) Framework:
The proposed prototype sequence5 consists of two imaging modes: Calibration and Runtime (Figure 1). For
optimal/improved
contrast between the balloon and cardiovascular anatomy, non-selective partial
saturation (pSAT)6 and fat suppression pulses are used. The sequence
starts with the Calibration mode
where a fixed stack of contiguous slices (n=10–20, 10 mm thickness, pSAT=90°) is
acquired in <2–3
s. Real-time post-processing of this slice stack using intensity thresholding
and pattern matching is performed within the prescribed shim box to compute the
initial 3D coordinates of the balloon. The sequence then automatically switches
to the Runtime mode where three contiguous
slices in the orientation of interest (10 mm thickness, pSAT=30–50°) are acquired continuously. Initially,
these three slices are automatically adjusted based on the 3D coordinates
obtained from the Calibration mode to
intersect the balloon in the central slice. During Runtime, the balloon position is continuously estimated from
real-time post-processing of the three slices using intensity thresholding,
pattern matching, and spatiotemporal constraints. Post-processing is restricted
to the volume of interest covered by the prescribed shim box to exclude
hyperintense regions like fat or interfaces and avoid false identification of
structures/spots. If the balloon is detected in one of the outer slices, the three
slices are repositioned to ensure that the catheter is in the central slice. Furthermore,
the sequence switches back to the Calibration
mode if the balloon is lost for >5 dynamics, for e.g. due to a sudden leap
of the catheter beyond the three-slice through-plane range in the Runtime mode.
(2) Experimental evaluation: The proposed framework was evaluated
in a 3D printed heart phantom and in two patients undergoing MR-guided cardiac
catheterization. All imaging experiments were performed on a 1.5T MRI scanner
(MAGNETOM Aera, Siemens Healthcare, Erlangen, Germany). The balloon of the
wedge catheter was filled with 1% gadolinium (Dotarem®) for positive contrast
visualization. The 2D single-shot bSSFP acquisition parameters were: TR/TE=2.44/1.02
ms, FA=50°, FOV=450×450 mm2, resolution=1.4×1.4 mm2,
slice thickness=10 mm, bandwidth=1010 Hz/px, GRAPPA factor=2, partial
Fourier=5/8. The Calibration stack
was prescribed in the coronal orientation for fast screening of the
cardiovascular system and remained fixed during the procedure. The Runtime slices can be prescribed freely
in the orientation of interest for optimal visualization of the desired
structures.Results
Figure 2(A) shows still frames
of the phantom experiment with the balloon successfully detected in the Calibration stack (slice 8) followed by
the three Runtime slices correctly
centred on the balloon. Figure 2(B) shows
the Runtime slices acquired at later
dynamics demonstrating automatic slice repositioning to follow the balloon in
the central slice. Figure 3 shows an
animation where the catheter was continuously tracked while it was navigated
through the phantom. Figures 4 (still
frames) and 5 (animation) demonstrate the application of the proposed approach
in one patient. As before, the balloon was initially detected in the Calibration mode and then tracked with
slice repositioning during Runtime
while the catheter was manipulated. In the Calibration
mode, the balloon was automatically identified with 100% accuracy both in the
phantom and in vivo. In the Runtime mode, the detection accuracy was
94.8% (88.5% in the central slice) in the phantom and 77.6±23.8% (59.1±2.8% in the central
slice) in vivo.Discussion
The proposed sequence can provide a more
robust guidance approach as it avoids manual slice repositioning and
manipulation of the imaging plane during catheter navigation. This sequence can
also facilitate the fast detection of out-of-plane catheters, which are
time-consuming to locate when manually tracked and may prolong the procedure.
A current limitation
of the proposed sequence is decreased temporal resolution due to the
acquisition of three contiguous slices in the Runtime mode. Future work would focus on implementing acceleration techniques
to achieve higher framerates and performing a larger clinical study to
establish its advantage over existing techniques.Conclusion
The proposed guidance approach enables continuous and robust automated slice
repositioning and catheter tracking during MR-guided cardiac catheterization.Acknowledgements
This work was supported by the Engineering and
Physical Sciences Research Council (EPSRC) grant (EP/R010935/1), the British
Heart Foundation (BHF) grants (PG/19/11/34243 and PG/21/10539), the Wellcome
EPSRC Centre for Medical Engineering at King’s College London (WT
203148/Z/16/Z), the National Institute for Health Research (NIHR) Biomedical
Research Centre based at Guy’s and St Thomas’ National Health Service (NHS)
Foundation Trust and King’s College London, and Siemens Healthineers. The views
expressed are those of the authors and not necessarily those of the NHS, the
NIHR or the Department of Health.References
[1] Razavi R, Hill DL, Keevil SF, et al.
Cardiac catheterisation guided by MRI in children and adults with congenital
heart disease. Lancet. 2003;362(9399):1877-1882.
[2] Pushparajah K, Chubb H, Razavi R.
MR-guided Cardiac Interventions. Top Magn Reson Imag. 2018;27(3):115-128.
[3] Ratnayaka
K, Faranesh AZ, Hansen MS, et al. Real-time MRI-guided
right heart catheterization in adults using passive catheters. Eur Heart J.
2013;34(5):380-389.
[4] Velasco Forte MN, Roujol S, Ruijsink B, et
al. MRI for Guided Right and Left Heart Cardiac Catheterization: A Prospective
Study in Congenital Heart Disease. J Magn Reson Imag. 2020:53(5):1446-1457.
[5] Shankar RV, Neji R, Duong P, et al.
Real-time Automatic Tip Tracking of Gadolinium-filled Balloon Wedge Catheter
during MR-guided Cardiac Catheterization. Proc Intl Soc Magn Reson Med 27.
Montreal (QC), Canada. 11-16 May 2019. P3844.
[6] Velasco
Forte MN, Pushparajah K, Schaeffter T, et al. Improved passive
catheter tracking with positive contrast for CMR-guided cardiac catheterization
using partial saturation (pSAT). J Cardiovasc Magn Reson. 2017;19(1):60.