Paul Kokeny1, Qiuyun Xu1, Sara Gharabaghi1, Sean Sethi1,2, Yu Liu3, Youmin Zhang3, Peng Liu3, Naying He3, Fuhua Yan3, and E. Mark Haacke1,2,4
1SpinTech MRI, Bingham Farms, MI, United States, 2Radiology, Wayne State University, Detroit, MI, United States, 3Shanghai Jiao Tong University School of Medicine, Ruijin Hospital, Shanghai, China, 4Neurology, Wayne State University, Detroit, MI, United States
Synopsis
Keywords: Image Reconstruction, Quantitative Imaging
By using the inherent relationship between
proton spin density (PSD) and T1, we propose a new image-processing approach to
reduce noise called CROWN (Constrained Reconstruction of
White Noise). Firstly, we established a linear relationship between these
two parameters, then applied a cost function to constrain simulated
Strategically Acquired Gradient Echo (STAGE) PSD map and T1 map data in the
presence of noise. Secondly, we applied this approach to
in vivo STAGE
images to reduce noise and improve SNR without the loss of image detail. CROWN
has the potential to make higher resolution or faster imaging viable with
improved SNR.
Introduction
There is a natural trade-off between scan time,
resolution and signal-to-noise ratio (SNR). Although many filters have been
developed to improve the SNR in MRI data, they all lead to some degradation of
the image which is recognizable as a remnant blurring relative to the original
data. Our goal is to show that, with the appropriate constraints applied to the
proton spin density (PSD) and T1 maps, one can obtain enhanced SNR without the
loss of detail. We call this approach CROWN which stands for “Constrained
Reconstruction of White Noise.” CROWN has been specifically designed for use
with multi-flip angle approaches such as strategically
acquired gradient echo (STAGE) imaging [1,2].Methods
The basic concept behind
CROWN is to use the following relationship between PSD (ρ=1/β) and T1 (1/R1) to
constrain the data in a way that reduces noise:
$$R1 = a β + b$$
This relationship can be
understood in terms of tissue water content, which drives both T1 and PSD values.
Often the PSD maps can be quite noisy, especially after correcting for T2*. CROWN
has the distinctive property of increasing the SNR in the PSD image. First, we
write the following cost function to calculate the estimated point (R1est,
βest) from the measured data (R1', β'):
$$C(β, R1) = min((β-β')^2
+ (R1 – R1')^2)$$
Second, substituting R1 from
Eq. (1) yields:
$$C(β, R1) = min((β-β')^2
+ ((aβ + b) - R1')^2)$$
And, finally, taking the
derivative of C(β, R1) with respect to β, and setting it equal to zero yields:
$$β_{est} = (β'+ a(R1'-b))/(1+a^2)$$
and
$$R1_{est}=a β_{est}+b$$
These new estimates (R1est,
βest) represent the CROWN output. They represent projections of
(R1',β') onto (R1,β) and they fall on the line represented by Eq. (1). Practically,
the coefficients
a and b for Eq. (1) need to be determined. The values used in this work are
based on the tissue properties of white matter (WM), gray matter (GM), and
cerebrospinal fluid (CSF) at 3T, as derived from the literature [3,4]. For WM,
T1=850ms, PSD=0.68, and T2*=53ms. For GM, T1=1611ms, PSD=0.84, and T2*=60ms.
For CSF, T1=4500ms, PSD=1, and T2*=2000ms. These three tissues are used to provide
three sample points on a plot of R1 versus β and a linear fit used to determine
the coefficients a and b by forcing the line to pass through the point
representing CSF. For example, for TE = 0ms, i.e., no T2* effects, the linear
relation is given by R1 = 2.03/sec β - 1.81/sec. To take T2* effects into
account, which was necessary for the in vivo
data shown below, values for PSD should be adjusted by the factor E2 =
exp(-TE/T2*), and linear regression re-performed using the new coefficients for
different TEs.
To evaluate the efficacy of CROWN, a test image was
built out of a set of embedded squares with each annular-like region
representing one tissue type (Figure 1). A TR of 25ms and TE of 0ms were used
to generate both 6° and 24° spoiled gradient-echo data. Gaussian noise was set
to be 10% of the signal from region 1 of the 6° magnitude image. PSD and T1
maps were generated using STAGE processing [1,2] and then CROWN was applied to
reduce the noise. Finally, both 6° and 24° images were simulated using the
improved PSD map. Two sets of in vivo STAGE data with identical
parameters were collected on a 3T Siemens Prisma scanner (Siemens Healthcare,
Erlangen, Germany). The coefficients used for CROWN at TE=5ms were a=1.52 and
b=-1.3. If the scanners remain stable between acquisitions, the only difference
between the images is background noise. The PSD maps (before and after CROWN) from
each acquisition were subtracted and noise was measured at the center and edge
of the brain using SPIN software (SpinTech MRI, Inc., Bingham Farms, MI). Results
As shown in the simulations (Figure 2) and the real
data (Figure 3), the noise was significantly reduced after CROWN processing for
the PSD map (see Table 1 for increases in SNR post CROWN for each simulated
region). After CROWN processing the SNR increased considerably for all tissues
but more so for the short T1 tissues. For the in vivo dataset, CROWN reduced the noise in the PSD map by a factor
of 1.77 at the center of the brain where receive coil sensitivity is generally
the poorest and 1.66 at the edge of the brain (Table 2).Discussion and Conclusions
CROWN offers a powerful new means to improve SNR
without modifying or blurring the image structures as is the case in most other
methods that purport to improve SNR. CROWN can lead to improved SNR not only in
the original PSD estimates but also in forward simulated MRI images. The major
limitations of CROWN are: the validity of the relationship between R1 and the
spin density; only PSD images are improved, and the method only works when PSD
and T1 maps are available. CROWN may have an immediate impact on improving SNR for
data collected with high parallel imaging acceleration factors, high resolution
imaging, or when the data are collected with radiofrequency receive coils with
a small number of channels. Acknowledgements
No acknowledgement found.References
1. Chen,
Y., Liu, S., Wang, Y., et al., STrategically Acquired Gradient Echo (STAGE)
imaging, part I: Creating enhanced T1 contrast and standardized susceptibility
weighted imaging and quantitative susceptibility mapping. Magn Reson Imaging,
2018. 46: p. 130-139.
2. Wang, Y., Chen, Y., Wu, D., et al., STrategically
Acquired Gradient Echo (STAGE) imaging, part II: Correcting for RF
inhomogeneities in estimating T1 and proton density. Magn Reson Imaging, 2018.
46: p. 140-150.
3. Wright,
P.J., et al., Water proton T1 measurements in brain tissue at 7, 3, and 1.5 T
using IR-EPI, IR-TSE, and MPRAGE: results and optimization. MAGMA, 2008.
21(1-2): p. 121-30.
4. Dieringer,
M.A., et al., Rapid parametric mapping of the longitudinal relaxation time T1
using two-dimensional variable flip angle magnetic resonance imaging at 1.5
Tesla, 3 Tesla, and 7 Tesla. PLoS One, 2014. 9(3): p. e91318.