Xinzeng Wang1, Daniel Litwiller2, Arnaud Guidon3, Tim Sprenger4, and Robert Marc Lebel5
1GE Healthcare, Houston, TX, United States, 2GE Healthcare, Denver, CO, United States, 3GE Healthcare, Boston, MA, United States, 4GE Healthcare, Stockholm, Sweden, 5GE Healthcare, Calgary, AB, Canada
Synopsis
Keywords: Image Reconstruction, Artifacts, Partial Fourier
A partial Fourier
acquisition has been widely used for fast MR imaging. To reduce the truncation
artifacts in partial Fourier image, Homodyne reconstruction is often used, and
it exploits the conjugate symmetry in real-valued signal to recover the full
k-space. However, the MR signal is complex-valued. Artifacts are commonly
observed in Homodyne images in the regions of rapid phase change due to the
interference of imaginary components of adjacent pixels. In this work, we
proposed a modified Homodyne reconstruction to reduce the conventional Homodyne
artifacts and truncation artifacts by using a high-resolution phase from a
pre-trained deep-learning network.
Introduction
A partial Fourier acquisition has been widely used for fast MR
imaging. Many partial Fourier reconstruction methods have been proposed to
minimize the truncation artifacts and image blurring due to the truncation in
k-space, such as Conjugate Synthesis, Homodyne, POCS, etc (1-4). These
reconstruction methods exploit the conjugate symmetry because real functions
have conjugate symmetry in the frequency domain. However, inhomogeneous B0
field, motion, unwanted phase shift, etc. introduce phase into the MR images,
violating the real-valued assumption. Therefore, partial k-space reconstruction
always requires some type of phase estimation that is not affected by the
aforementioned truncation artifacts. However, the conventional reconstruction
methods use a low-resolution phase for phase correction, resulting in gross
artifacts in regions of the image where the phase from the low-pass portion of
kspace fails to adequately resolve rapid local variations.
To overcome these challenges and mitigate these artifacts, this
work describes a modified partial Fourier reconstruction method using a
high-resolution phase generated by a pre-trained deep learning network. The
proposed method was evaluated with out-of-phase image, functional MRI (fMRI)
and diffusion weighted MRI (DWI) images. Methods
The Conventional Homodyne reconstruction method uses a
low-resolution phase for phase correction after weighting convolution. In
regions with rapid phase variations, the phase correction could not completely
suppress the interferences of the imaginary component in the nearby pixels (1),
resulting in Homodyne artifacts.
In contrast, the proposed modified Homodyne reconstruction
uses a high-resolution phase map for phase correction before the weighting
convolution. The high-resolution phase map was estimated using a deep-learning
network, which was trained from a database of over 10,000 images. After a more
accurate phase correction, the interference of the nearby pixels would be well
suppressed, therefore reducing Homodyne artifacts.
A digital phantom was used to demonstrate the feasibility and
improvements of the modified Homodyne reconstruction method. Then, out-of-phase
liver images, fMRI images and diffusion weighted images were acquired using
half-NEX on a GE 3T MRI scanner (Discovery MR750, GE Healthcare, Waukesha, WI)
with IRB approval and written informed consent to compare the conventional
Homodyne reconstruction and the modified Homodyne reconstruction methods.Results and Discussions
In the digital phantom, each component was assigned a different
phase (Figure 1b) to simulate phase variations in real MR imaging. The k-space
was truncated to simulate partial Fourier acquisition with a partial NEX of
0.625. The image reconstructed using zero-filling showed the ringing artifacts
(Figure 1c), which are reduced by the Homodyne reconstruction (Figure 1d). The
conventional Homodyne reconstruction generated artifacts (Homodyne artifacts)
in the regions of rapid phase changes, as shown in Figure 1d. This is mainly
caused by the interference of the imaginary component in the adjacent voxels.
The modified Homodyne reconstruction corrects the phase with a high-resolution
phase map generated from a pre-trained DL network, eliminating the interference
of imaginary components in the adjacent voxels. Both the truncation artifacts
and the traditional Homodyne artifacts were well minimized using the modified
Homodyne reconstruction.
To evaluate the Homodyne reconstruction methods in regions of
rapid phase change, the out-of-phase images were acquired in a volunteer with a
partial NEX of 0.65. Water and fat signals have ~180 degrees phase difference
in out-of-phase images, generating the India ink artifact at the fat-water
interface, as shown in Figure 2a. The India ink artifact could be used in
diagnosis to confirm the fatty nature of some lesions. However, part of the
dark boundary disappeared in the image that was reconstructed using conventional
Homodyne due to the interference of the imaginary component in adjacent voxels,
as shown in Figure 2b (red arrows). In contrast, modified homodyne
reconstruction minimizes the interference with a robust high-resolution phase
correction and improves the definition of water-fat boundaries (Figure 2c,
green arrows).
Functional MRI (fMRI) images were often acquired with partial
Fourier to accelerate the acquisition. However, bulk motion, pulsation motion
and field inhomogeneities could introduce high-frequency phase in fMRI images.
With the conventional homodyne artifacts, Homodyne artifacts with dark
boundaries or pixels are commonly seen in fMRI images, as shown in Figure 3a
(red arrows). These artifacts were well suppressed using the modified Homodyne
reconstruction, as shown in Figure 3b.
Diffusion weighted imaging
also uses partial Fourier to reduce the echo train length and distortion.
However, the movement of the eyeball, CSF and blood during the diffusion
preparation could introduce rapid phase change, resulting in a signal loss in
these regions (red arrows, Figure 4a, 4c). However, this kind of artifact was
minimized using the proposed modified Homodyne reconstruction.Conclusion
A high-resolution
phase generated from the pre-trained deep learning model enables robust phase
correction for Homodyne reconstruction. With a robust and accurate phase
correction, the proposed modified Homodyne swapped the order of phase
correction and weighting convolution, reducing the truncation artifacts and the
traditional Homodyne artifacts due to the interference of imaginary components
in adjacent voxels.Acknowledgements
No acknowledgement found.References
[1] Partial k-space Reconstruction by John
Pauly, https://users.fmrib.ox.ac.uk/~karla/reading_group/lecture_notes/Recon_Pauly_read.pdf
[2] Noll DC, Nishimura GD, Macoviski A. Homodyne
detection in magnetic resonance imaging. IEEE Trans Med
Imaging 1991; 10:154–163.
[3] Haacke EM, Lindskog ED, Lin W. A fast, iterative,
partial-Fourier technique capable of local phase recovery. J Magn
Reson 1991; 92:126–145.
[4] Cuppen J, van Est A. Reducing MR imaging time by
one-sided reconstruction. Magn Reson Imaging 1987; 5:516–527