Daniel I Gendin1 and Jiaen Liu2
1Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, United States, 2Advanced Imaging Research Center and Radiology, UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
Keywords: Image Reconstruction, Brain
Head motion can lead to artifacts in MR images. One source
of motion artifacts is dynamic nonlinear variation in the background magnetic
field (B
0). Fully accounting for this variation is computationally
intractable. We proposed an efficient method of performing this correction by
making use of principal component analysis and considering only a few dominant
modes. We applied our method to measured data and
compared the results to the case with no correction as well correction using a
previous K-means based method.
Introduction
MRI is prone to motion-induced artifacts. One potential
source for these artifacts is the fact that head motion can lead to spatially
nonlinear dynamic changes in the background polarizing magnetic field (B0)
due to the static B0 shim setting and motion-induced spatial changes
of magnetic susceptibility in the head and torso1. The resulting
artifacts become much more pronounced at higher field strengths (e.g. 7T) and
in methods focusing on the susceptibility effect such as T2*-weighted
MRI. Therefore, in order to fully take advantage of high-field MRI’s unique
ability to image fine scale structures, the effect of motion-induced nonlinear
B0 fluctuation must be corrected for. Several works have attempted
to address this issue2-4. Performing the correction on a
shot-by-shot basis is computationally expensive. Here, we proposed an efficient
method based on principal component analysis (PCA) of the B0 effect
on MRI signal5 and compared with another method using K-means
clustering4.Methods
Principle of PCA-based B0 correction
The forward equation for how MR signal is affected by the B0
field (unit hertz) is:
$$s(k(t))=M\cdot\mathcal{FT}\{e^{2i\pi\,TE\,B_0(r,t)}\rho(r)\}.$$
Here, $$$s$$$ is the signal, $$$\rho$$$ is the underlying
image, $$$TE$$$ is the echo time, $$$\mathcal{FT}$$$ is the Fourier transform,
and $$$M$$$ is the motion operator (for simplicity, B1 sensitivity
is omitted here). Directly solving this equation would require as many FTs as
there are B0 measurements, which is computationally intractable. To
efficiently perform the correction, we approximated the exponential phase
factor (EPF) from the equation above with a sum of a few principal components (as
proposed by Wilm et. al5).
$$\text{EPF}=e^{2i\pi\,TE\,B_0(r,t)}\approx\sum_{j=1}^Lb_j(r)c_j(t).$$
Here, $$$b_j(r)$$$ is the jth principal mode or component,
$$$c_j(t)$$$ is the weight of mode j at time t, and $$$L$$$ is the number of
components that was used to perform the approximation.
Example values of the EPF and its corresponding principal
modes are presented in Figure 1.
Plugging this decomposition into the forward equation, we
obtained:
$$s(k(t))=M\cdot\sum_{j=1}^Lc_j(t)\mathcal{FT}\{b_j(r)\rho(r)\}.$$
The solution of the above equation now only requires L FTs.
MRI Experiment
We
performed two sets of experiments to evaluate the proposed correction algorithm.
In Experiment I, eight healthy subjects (age:46±20, 4 males) were recruited and scanned on a 3T
MRI (Prisma, Siemens). Subjects were instructed to stay relaxed, and the B0
changes in the brain during eight-minute time windows were measured using a 3D
EPI sequence (resolution:6x6x14mm3, FOV:240x180x144mm3,
TE:4.2ms, and volume TR:0.65s). The scans were repeated 6 times for each
subject. B0 changes in the head frame were estimated from the
co-registered EPI phase data. The obtained data was used to evaluate the
validity of our approximation to the EPF. The data was also used to perform
simulations where the corruption to a known image was simulated using the
measured B0 changes (TE:60ms) and the correction was applied to the
corrupted image. In Experiment II, we considered 3D T2*-weighted MRI
at 3T (isotropic 2mm resolution, TE:32ms, and FOV:240x180x120mm3). Two
(male) subjects were recruited. Subjects were instructed to move their head,
and a separate scan served as the reference when subjects stayed still. 3D
EPI-based navigators were acquired with spatial resolution of 6x5.6x6mm3 and
temporal resolution of 0.5s (for details see Liu et.al4). The
T2*-weighted images were reconstructed with different corrections (see Figure 4).
The reconstruction performance was evaluated based on the normalized root mean
square error (nRMSE) against the reference scan.Results
The results for an example simulation are presented in
Figure 2. We observed that the reconstructions were improved by correcting for
the nonlinear B0 fluctuations. The correction improved with more PCA
modes and reached a point after which including additional modes did not
significantly improve the reconstruction.
We also directly evaluated the validity of our approximate EPF
by computing the normalized Frobenius norm of the difference between the full and
approximate EPFs. The results for this comparison for all subjects were summarized
in Figure 3. The approximation was performed using both PCA and K-means. For a small
number of modes/means, K-means slightly out-performed PCA, but as the number of
modes/means increased, PCA performed better than K-means.
The results for experiments where subjects were asked to
intentionally move, are presented in Figures 4 and 5. We observed that
performing a nonlinear correction improved the reconstruction as compared to only
a linear correction. The difference between PCA and K-means based
reconstructions was not significant from an nRMSE perspective, but we did
observe some qualitative improvements in PCA based reconstructions (denoted by
arrows). We noted that motion (especially pitch) was greater for subject 1 than
2, we correspondingly observed more artifacts in subject 1. We observed that the
overall reconstruction times for PCA and K-means based reconstructions were
similar when the number of modes and means was the same.Discussion and Conclusions
In this work, we showed that PCA can be used to efficiently
correct for motion-induced nonlinear B0 changes. PCA-based
correction performed similarly to the previously established K-means based
approach4 from an nRMSE perspective, with some qualitative
improvements observed in PCA-based reconstructions. Systematic evaluation of
this approach and other methods, such as the K-means approach, in a larger
subject population is warranted in future studies to further establish its
efficacy in high field and B0-sensitive applications, such as
T2*/susceptibility-weight MRI.Acknowledgements
The authors
gratefully acknowledge funding through the UT Southwestern faculty startup fund. We would also like to thank Yujia Huang for his help with data acquisition.References
1. Liu, J., de Zwart, J. A.,
van Gelderen, P., Murphy‐Boesch,
J., & Duyn, J. H. (2018). Effect of head motion on MRI B 0 field distribution. Magnetic
resonance in medicine, 80(6), 2538-2548.
2. Brackenier, Y., Cordero‐Grande, L., Tomi‐Tricot, R.,
Wilkinson, T., Bridgen, P., Price, A., ... & Hajnal, J. V. (2022). Data‐driven motion‐corrected brain
MRI incorporating pose‐dependent B0 fields. Magnetic Resonance in Medicine.
3. Dong, Z., Wang,
F., & Setsompop, K. (2022). Motion‐corrected
3D‐EPTI with efficient 4D navigator acquisition
for fast and robust whole‐brain
quantitative imaging. Magnetic Resonance in Medicine.
4. Liu, J., van Gelderen,
P., de Zwart, J. A., & Duyn, J. H. (2020). Reducing motion sensitivity in
3D high-resolution T2*-weighted MRI by navigator-based motion and nonlinear
magnetic field correction. NeuroImage, 206, 116332.
5. Wilm, B. J., Barmet, C.,
& Pruessmann, K. P. (2012). Fast higher-order MR image reconstruction using
singular-vector separation. IEEE transactions on medical imaging, 31(7),
1396-1403.