Vivian S. Nguyen1,2, Sudarshan Ragunathan3, Timothy J. Carroll4, Adil Javed5, Marcella K. Vaicik1, John G. Georgiadis1, and Keigo Kawaji1,2
1Biomedical Engineering, Illinois Institute of Technology, Chicago, IL, United States, 2Medicine - Cardiology, University of Chicago Medical Center, Chicago, IL, United States, 3Hyperfine Inc., Guilford, CT, United States, 4Radiology, University of Chicago Medical Center, Chicago, IL, United States, 5Neurology, University of Chicago Medical Center, Chicago, IL, United States
Synopsis
Keywords: Relaxometry, Low-Field MRI
T1
maps provide quantitative relaxometric measures of tissue responses that may carry
correlative indices of specific disease indications. On the recently
commercialized 64mT system (Hyperfine Inc.), little has been reported regarding
both clinical T1 relaxometric normal values in presence of Gadolinium-based
contrast and relevant acquisition schemes with Gd contrast enhancement. In this
study, we examine an array of Gadolinium-doped vials that represent 10-5000
fold dilution in the vasculature using an Inversion-Recovery FSE acquisition to
gain key insights on: a) anticipated normal values in presence of Gd, b) conventional
fitting algorithm performance, and c) potential inferences on inversion timing
selection.
Introduction
T1 mapping at 64mT field
strength has considerable potential to bring forth tissue characterization
methods that are already established on the higher field clinical MRI systems.
Such gained knowledge may provide important knowledge pertaining to optimal
pulse sequence parameters, T1-based contrast dosage, and protocol timing as new
techniques emerge specifically on this low-field system. Of note, voxel-wise
parametric fitting according to the relaxometric signal evolution as described
by the Bloch equation provides a choice post-processing – and allow for robust
measurement yields of tissue T1 with the inversion times and matched signal
intensity values. Despite this, relatively little is delineated about key acquisition
timing considerations (i.e. normal values for inversion times and number of
samples) best suited for this low-field system. In this study, we provide an
initial combinatorial approach based on an exhaustive acquisition of a
Gadolinium-doped vial array phantom.Materials and Methods
Phantom Preparation: Figure 1a shows the prepared Gd-based phantom using
Gadodiamide (Omniscan) (GE Healthcare, Waukesha WI). [Gd] was diluted from a
485mM stock solution at varied concentrations with range [0.1mM, 1.0mM] at
0.1mM increments, [1.1mM, 1.9 mM] at 0.2mM increments, and additionally at 2,
5, 10, 15, 20, and 40mM respectively, representing dilution factors between ~10
and 5000 that would be seen during both conventional bolus chase and delayed
accumulation in vasculature. A Distilled H2O reference vial, and a
duplicate 0.7mM vial were also included into the array for both quality control
and measurement reproducibility purposes.
Figure 1b shows the typical
signal intensity plot measurements depicted by the default product brain T1w
parameters at 64mT (parameters not shown), along with relative SI
measurements obtained on a higher field clinical system at 3.0T using product
brain T1w sequence.
Data Acquisition:
Data for T1 estimation were obtained using
Inversion-Recovery Fast Spin Echo (IR-FSE) with inversion times (TI) ranging
from 85 – 966 ms on a 64mT Hyperfine Swoop (Guilford CT) scanner with an
8-channel head coil array. Nineteen TIs examined in this study were as follows:
TIN=19 = [85, 132, 152, 248, 287, 309,
392, 416, 454, 534, 576, 617, 698, 732, 787, 831, 876, 941, 966] ms.
T1 map generation and
processing: A reference T1 map was generated
using a standard 3-parameter fit using all available inversion timepoint 3D
volumes (n=19). Another 13 were implemented modeling other schemes1.
A combinatorial subset commencing from applying the 3-point T1 fits on up to
half of these volumes to three volumes were examined. The first 40 were
generated using pseudo linear distribution across the full TI range of ~800ms,
and smaller 400-500ms TI ranges in the full acquisition interval. A difference
map of yielded T1 measures against the full map was derived and assessed using
a Region-of-Interest analysis. Derived
T1map variations against the full 19-TI fitted T1 map were analyzed using
region-of-interest analysis. Our observed findings are provided below.Results
The
SI profile plot concurs with both literature2 the normal range of
[Gd] assessed on our higher field system, yielding peak signals at [Gd]=1.5~1.7mM
vials and a similar noticeable drop above [Gd]=2mM at both examined field
strengths using T1w sequence tuned to the brain. Both high and low-field T1w
measures concurred, with visually noticeable differences in terms of linearity
over smaller [Gd] range subsets (e.g. in upward signal increase between [1.5,
1.7] mM). Pearson’s correlation yielded R2 = 0.83 for this
comparison.
Figure
2 shows the 19-TI-driven T1 map. Vials with [Gd] above 1.5mM yielded greater
residual error during 3-parameter fitting. Figure 3 shows the relative comparisons
of 53 examined combinatorial subsets pseudo-linear inversion timepoints, where we
highlight the best 5-parameter fit (every 100ms), a robust 3-parameter fit
(every 150 ms), and a third custom fit (i.e. 2-3x short TIs, one long TI). The depicted examples were determined by
overall measurement differences between voxelwise T1s against the reference
comparison. We observed notable discrepancies in fitted T1 measures despite the
post-processing was performed on matched number of inversion times but with
50ms offsets; likewise similar trends can be found when select inversion time
volumes are permuted at the front or back of the TI window (Figure 3b vs 3d; 3c
vs 3e).Discussion/Conclusion
We report
the findings of over 50 combinations of T1 fitting across a range of n=21 Gd-doped vials
with concentration varying between [Gd] = [0.1, 40]mM. Of note, we caution that
3-parameter fitting yielded non-converging, residual errors at [Gd] above
1.5mM. Our findings may offer insights
towards an initial set of candidate acquisition parameters that would be best
suited for 3-point fitted IR-FSE T1 relaxometry mapping consistent with
exhaustive sampling.Acknowledgements
This project was supported by NIH K25
HL141634 and the R.A. Pritzker Chair.References
1. Piechnik SK, Ferreira VM, Dall'Armellina E, Cochlin LE, Greiser A, Neubauer
S, Robson MD. Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for
clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold.
J Cardiovasc Magn Reson. 2010 Nov 19;12(1):69. doi: 10.1186/1532-429X-12-69.
PMID: 21092095; PMCID: PMC3001433.
2. Bleicher AG, Kanal E. A serial dilution study of
gadolinium-based MR imaging contrast agents. AJNR Am J Neuroradiol. 2008
Apr;29(4):668-73. doi: 10.3174/ajnr.A0905. Epub 2008 Jan 9. PMID: 18184840;
PMCID: PMC7978187.