Chloé Najac1, Kirsten Koolstra2, Tom O’Reilly1, and Andrew Webb1
1C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center, Leiden, Netherlands, 2Philips, Best, Netherlands
Synopsis
Keywords: Low-Field MRI, New Trajectories & Spatial Encoding Methods
Low-field MRI
systems (with B
0<0.1T) for point-of-care applications are becoming
increasingly widespread. Imaging at low-field remains challenging due to the low
intrinsic SNR. We evaluated speeding up 3D imaging using radial-based Cartesian
undersampled phase-encodings (PEs). In phantoms, we tested different
undersampling schemes and compared them to full in-out Cartesian PE in terms of
peak SNR (PSNR) to take account of spatial resolution and image SNR. Results
suggests that a radial-based Cartesian PE trajectory with an overall
acceleration factor of two can be implemented while preserving image quality (PSNR~69dB
with R=2 vs. ~70dB with R=1.3 in a discrete spatial resolution phantom).
Introduction
Point-of-care
(POC) imaging with low-field MRI (<0.1T) has been suggested as an important
element in increasing accessibility to healthcare in low and middle income countries1.
The main challenge is the intrinsic low SNR2, resulting in trade-off
between scan time and image resolution and limiting the minimum detectable size
of different pathologies. 3D turbo-spin echo (TSE) readouts are most commonly
used2 due to high k-space efficiency and relative
insensitivity to B0 inhomogeneities. Nevertheless, the current lack
of parallel imaging capability and limited gradient strengths mean that acquisition
times are longer than those used on conventional clinical systems. We evaluated
the feasibility to reduce acquisition time, while preserving detection-size
limits, using a 3D TSE with radial-based Cartesian k-space in the phase-encode
(PE) directions (Fig.1). Using compressed-sense reconstruction to correct for undersampling, we found in phantoms that
acquisition time can be reduced by a factor of 2 while preserving image quality. Materials and Methods
Hardware: We used a portable 46 mT
Halbach array-based magnet (outer/inner diameter=60.0/30.1cm, length=49.2cm,
weight=111kg), with custom-built RF amplifier, Bruker gradient amplifiers and
a Magritek Kea2 spectrometer3. Imaging was performed using a
solenoid or an elliptical spiral-solenoid head coil1-3.
Cartesian-Radial
PE trajectory implementation: We implemented a 2D radial-cartesian PE trajectory4. The trajectory follows
the Cartesian grid but is sampled in a radial pattern. We used an elliptical
circumscribed kPE,1-kPE,2 matrix (i.e. not
acquiring the corners of k-space), fully sampled in the center and
undersampled outside a certain center diameter (Fig.1). The trajectories
were created in Python code, and exported as text files to the spectrometer. As
represented in Fig.1, we tested different center-diameters (dcenter,
varying from 20 to 100% of the full ellipse dimension) and different
undersampling factors outside the center area (rout, varying from 2
to 8). Data were compared with a conventional in-out 2D PE trajectory (with and
without elliptical windowing).
Discrete
spatial
resolution phantom: We
used a phantom containing capillaries with inner diameters varying from 1mm to
5mm. Capillaries with 1mm diameter were spaced 1mm apart, those with 2mm
diameter 2mm apart, etc. Capillaries were filled with oil, resulting in short T1
and T2 relaxation values. Data were acquired with all trajectories
described above and using the following acquisition parameters: TR/TE=600/15ms,
echo-train-length (ETL)=6, resolution (FE,PE1,PE2)=1x1x1mm3,
acquisition bandwidth (BW)=20 kHz. Following Fourier transform along the frequency
encoding direction (FE), data from 30 slices were averaged to increase SNR,
giving an effective resolution of 1x1x30mm3.
Morphometric
brain-like (BrainLo) phantom: BrainLo was 3D printed with Polylactic Acid and filled with
different solutions of agarose/copper-sulphate doped-water/deuterated-water to
imitate brain tissue relaxation properties. Data were acquired with all
trajectories described above (except dcenter=20% and rout=8)
using the following acquisition parameters: TR/TE=600/14ms, ETL=6, resolution(FE,PE1,PE2)=1x2x2mm3,
BW=20 kHz, no signal averaging. Following Fourier transform along frequency
encoding direction (FE), data from 11 slices were averaged to increase SNR,
giving an effective resolution of 2x2x11mm3.
Undersampling
correction:
To correct for undersampling, we used Split Bregman (SB)5 as a
non-linear minimization scheme. The regularization parameters were tuned
empirically and set to μ=5 and λ=20.
Comparison: We compared all undersampled
datasets (in-out with elliptical windowing and radial-based Cartesian PE
trajectories) with the fully sampled in-out dataset ($$$u_{ref}$$$). For each reconstruction ($$$u$$$), we calculated the peak SNR (PSNR)
defined as $$$20*log_{10}(\frac{max(u_{ref})}{RMSE(u_{ref},u)})$$$. The PSNR represents a measure of both
the sharpness and SNR of the image. We measured the PSNR across the entire
image (BrainLo phantom), as well as across regions with different capillary
diameter in the resolution phantom (Fig.2).Results and Discussion
Trajectory
implementation:
Fig.1 and
2 illustrates
how k-space was sampled with the various undersampling
combinations (different dcenter and rout), the total
acceleration factor R and resulting acquisition time. Using these combinations,
we could accelerate our acquisition by factors varying from 1.3 to 8.1.
Resolution
phantom:
Fig.3 shows images after undersampling correction. Capillaries with 1mm inner
diameter could not be clearly detected with any trajectory including full k-space
coverage. Capillaries with diameter 2mm and higher could be separated up to an
overall reduction factor of 3-4. When measuring the PSNR (Fig.4) across
the entire image, there was a very small loss with acceleration factors up to
2. Looking at capillaries of different sizes, we found that as expected image
loss was greater for the smaller capillaries with PSNR being reduced by 2dB
with acceleration factors up to 2 and by 10 dB with acceleration factor 8.
There were little differences in PSNR when comparing the two different undersampling
approaches (reducing the fully sampled center-diameter or increasing the undersampling
factor outside the diameter) as long as the overall acceleration factor R was
similar.
BrainLo
phantom:
Fig.5 illustrates the effects on the BrainLo phantom. Although PSNR
values remained relatively affected across acceleration factor (~60 dB), images
with undersampling all appeared less sharp. Conclusion
We have shown the
potential to use a radial-based Cartesian trajectory with undersampling scheme
to reconstruct 3D TSE images on a POC 46mT Halbach MRI scanner. Results in resolution phantom suggest that acceleration factors up to 2 are feasible, while preserving image
sharpness and SNR, for feature sizes of ~2mm. Results in a brain-mimicking
phantom showed relatively similar results. In future, we will evaluate improvement
when combining the radial-based PE with partial Fourier.Acknowledgements
This project
has received funding from Horizon 2020 ERC Advanced PASMAR 101021218 and the Dutch
Science Foundation Open Technology 18981.References
1Anazado, NMR in Biomed. (2022) ;
2O’Reilly et al., MRM (2021) ; 3O’Reilly et al., MRM
(2020); 4Busse et al., MRM (2008); 5Koolstra et al.,
MAGMA (2021)