Sajad Mohammed Ali1, Nirbhay Yadav2, Ronnie Wirestam1, Munendra Singh3, Hye-Young Heo3, Peter van Zijl2, and Linda Knutsson4
1Medical Physics, Lund University, Lund, Sweden, 2Radiology, F.M. Kirby Research Center, Johns Hopkins University, Kennedy Krieger Institute, Baltimore, MD, United States, 3Radiology, Johns Hopkins University, Baltimore, MD, United States, 4F.M. Kirby Research Center, Radiology, Medical Radiation Physics, Kennedy Krieger Institute, Johns Hopkins University, Lund University, Baltimore, MD, United States
Synopsis
Keywords: Data Analysis, Machine Learning/Artificial Intelligence, Lorentzian curve fitting
Water
saturation shift referencing (WASSR) Z-spectra can be used to correct shifts
due to B
0-field inhomogeneities, for magnetic susceptibility mapping
and analysis of relaxation effects. The spectra follow a Lorentzian shape with
discrete values. Hence, a Lorentzian fit to retrieve the shape parameters
(amplitude A, line width LW and frequency shift Δf
H2O ) simplifies analysis. Conventionally, the
least-squares (LS) method is used for such fitting despite being time consuming
and sensitive to the unavoidable noise in
vivo. We propose a deep learning-based Lorentzian-fitting neural network
(LoFNet) that demonstrated improved robustness against noise and sampling
density in combination with reduced time consumption.
Introduction
By irradiating a sample with radiofrequency (RF) pulses over a discrete number of offset frequencies and taking the normalized water signal as a function of the saturation frequencies a Z-spectra is formed6. If the irradiated sample contains exchangeable protons in solute molecules the water signal in the Z-spectrum will decrease at the resonance frequency of the solute due to saturation transfer1,6–11. At the resonance frequency of water, the signal will be further reduced due to direct saturation (DS)6 which will result in a Lorentzian shape12. By using a sufficiently low B1 and short saturation duration the DS effect in the Z-spectrum can be isolated1 yielding a water saturation shift referencing (WASSR) Z-spectrum. WASSR Z-spectra are useful for correcting shifts due to B0-field inhomogeneities1–4, for magnetic susceptibility mapping5 and analysis of relaxation effects4. By fitting a WASSR Z-spectrum to a Lorentzian curve, the shape parameters (amplitude A, line width LW and frequency shift $$$\Delta f_{H_2O}$$$) are directly obtained thus simplifying the analysis. The conventional approach is to use a least-squares Lorentzian fitting. However, this method is time consuming and sensitive to the unavoidable in vivo noise. To overcome these shortcomings, we propose to use a deep learning based Lorentzian fitting neural network (LoFNet).Theory
The
normalized WASSR Z-spectrum follows a Lorentzian line shape defined as:
$$Z(\Delta f_{RF})=\frac{S_{sat}(\Delta f_{RF})}{S_0}=1-\frac{A\cdot LW^2}{LW^2+4(\Delta f_{RF}-\Delta f_{H_2O})} (1)$$
where $$$S_{sat}(\Delta f_{RF})$$$
is the signal intensity at the irradiation
offset from the water proton frequency, $$$S_0$$$
is the intensity without saturation and $$$\Delta f_{RF}$$$
is the saturation frequency offset (in Hz). A is the normalized signal amplitude, LW the linewidth (in Hz) and $$$\Delta f_{H_2O}$$$ is the water proton frequency offset (in Hz).Methods
A training dataset consisting of 5,000,000
sample-label pairs was simulated using Eq. (1) and noise was added before
transforming each sample to be governed by a Rician probability density
function. The shape parameters (A, LW and $$$\Delta f_{H_2O}$$$) were randomly chosen from pre-defined intervals for
each sample and saved as a corresponding label to the simulated sample.
Similarly, a test dataset was simulated consisting of 10,000 sample-label
pairs. An additional group of test data was simulated each with 10,000
sample-label pairs but with increasing noise level from 0 to 4.5 % with 0.5 %
increments.
The study was approved by the
Institutional Review Board (IRB) and written informed consent was obtained from
each subject. WASSR Z-spectra from 3 T scans were extracted from white matter
(WM), gray matter (GM), cerebrospinal fluid (CSF) and tumour (T) regions
yielding in vivo datasets.
A neural network architecture was constructed and
the hyperparameters optimized using the Python library HyperOpt13. The optimized network LoFNetHP was
trained on 90 % of the training dataset and the remaining part was used as a
validation set during training. The trained model was evaluated on the in vivo and simulated datasets.
Prediction errors (PEs), robustness against increased noise and robustness
against reduced sampling density were compared to those of the LS-method. In
addition, a comparison of the time consumption for the two methods was made. Metrics
used were root-mean-squared error (RSME) and mean-absolute error (MAE).Results
The
LoFNetHP and the LS-method produced comparable PEs on all in vivo
datasets (WM, GM, CSF, T) with no statistically significant difference $$$p=0.35$$$ and $$$p=0.27$$$ for RMSE
and MAE, respectively. Figure 1 shows three representative fits on samples from
WM, GM, and CSF. Both methods showed good fits to the discrete sample
values and are overlapped by visual inspection. A closer analysis (Figure 1,
bottom row) shows small differences between the two methods on the
order of 1 % or less. However, a significant difference ($$$p=0.048$$$) in time consumption was observed, where LoFNetHP
was on average 70 times faster.
The
LS-method exhibited a tendency of a rapid increase in PEs with increased sample
noise (same trend can be seen for both metrics), whereas LoFNetHP
showed marginal increase of PE up to 4.5 % (Figure 2). The insets of Figure 2 show
that the PEs for LS start to increase after 0.5 %. The two methods produced
comparable PEs up to approximately 1 % after which the PEs for LS took off
rapidly. With reduced number of sample points, the PEs increased for both
methods (Figure 3). However, the increase occurred earlier and is more
pronounced for LS.Discussion
The improved robustness against noise for
LoFNetHP compared to LS allows for higher noise levels and thus a
reduced MRI acquisition time of the WASSR spectra or alternatively, an
increased spatial resolution. This can be applied to, for example, quantitative
susceptibility mapping. The improved robustness of LoFNetHP against
reduced sampling density compared to LS is a good indicator of the method’s
generalizability to different sampling protocols. However, the sampling density
should not be reduced below 8 points to ensure sufficiently low PEs (Figure 3).Conclusion
In Lorentzian fitting of WASSR spectra,
the proposed deep learning method, LoFNetHP, showed improved robustness against
noise and against reduced sampling density as well as significantly reduced
time consumption compared to the conventional LS-method. These advantages increase
the efficiency of the analysis, ease SNR requirements and facilitate
generalizability.Acknowledgements
No acknowledgement found.References
1. Kim
M, Gillen J, Landman BA, Zhou J, van Zijl PCM. Water saturation shift
referencing (WASSR) for chemical exchange saturation transfer (CEST)
experiments. Magn Reson Med. 2009;61(6):1441-1450. doi:10.1002/mrm.21873
2. Müller-Lutz A, Matuschke F,
Schleich C, et al. Improvement of
water saturation shift referencing by sequence and analysis optimization to
enhance chemical exchange saturation transfer imaging. Magn Reson Imaging.
2016;34(6):771-778. doi:10.1016/j.mri.2016.03.013
3. Liu
G, Qin Q, Chan KWY, et al. Non-invasive temperature mapping using
temperature-responsive water saturation shift referencing (T-WASSR) MRI. NMR Biomed. 2014;27(3):320-331.
doi:10.1002/nbm.3066
4. Smith SA, Bulte JWM, van
Zijl PCM. Direct saturation MRI: theory and
application to imaging brain iron. Magn Reson Med. 2009;62(2):384-393. doi:10.1002/mrm.21980
5. Lim IAL, Li X, Jones CK,
Farrell JAD, Vikram DS, van Zijl PCM. Quantitative
magnetic susceptibility mapping without phase unwrapping using WASSR. NeuroImage.
2014;86:265-279. doi:10.1016/j.neuroimage.2013.09.072
6. van
Zijl PCM, Yadav NN. Chemical exchange saturation transfer (CEST): what is in a
name and what isn’t? Magn Reson Med. 2011;65(4):927-948.
doi:10.1002/mrm.22761
7. Zhou
J, Payen JF, Wilson DA, Traystman RJ, van Zijl PCM. Using the amide proton
signals of intracellular proteins and peptides to detect pH effects in MRI. Nat Med. 2003;9(8):1085-1090.
doi:10.1038/nm907
8. Gilad AA, McMahon MT,
Walczak P, et al. Artificial reporter gene providing MRI
contrast based on proton exchange. Nat Biotechnol. 2007;25(2):217-219.
doi:10.1038/nbt1277
9. van
Zijl PCM, Jones CK, Ren J, Malloy CR, Sherry AD. MRI detection of glycogen in
vivo by using chemical exchange saturation transfer imaging (glycoCEST). Proc
Natl Acad Sci U S A. 2007;104(11):4359-4364. doi:10.1073/pnas.0700281104
10. Ward
KM, Aletras AH, Balaban RS. A new class of contrast agents for MRI based on
proton chemical exchange dependent saturation transfer (CEST). J Magn Reson
San Diego Calif 1997. 2000;143(1):79-87. doi:10.1006/jmre.1999.1956
11. Vinogradov
E, Sherry AD, Lenkinski RE. CEST: from basic principles to applications,
challenges and opportunities. J Magn Reson San Diego Calif 1997.
2013;229:155-172. doi:10.1016/j.jmr.2012.11.024
12. Mulkern
RV, Williams ML. The general solution to the Bloch equation with constant rf
and relaxation terms: application to saturation and slice selection. Med
Phys. 1993;20(1):5-13. doi:10.1118/1.597063
13. Bergstra J, Yamins D, Cox D. Hyperopt: A Python Library for
Optimizing the Hyperparameters of Machine Learning Algorithms.; 2013:19.
doi:10.25080/Majora-8b375195-003