Xiao Wu1, Shan Xu2, Yao Zhang2, Jianzhong Sun2, and Peiyu Huang2
1Radiology, The Second Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China, 2The Second Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China
Synopsis
Keywords: Data Acquisition, Machine Learning/Artificial Intelligence, deep-learning ,susceptibility-weighted imaging,magnetic resonance imaging
In
this study, we validated a deep-learning-based method for accelerating
susceptibility-weighted imaging (SWI) in 31 clinical subjects. Compared to the
fully sampled images, the accelerated SWI images had less noise and imaging
artifacts. Although the images had decreased sharpness, the anatomical details
of the lesions were mostly kept, and we had not observed false
negative\positive lesions. This method could be useful for clinical situations
that need timely imaging results.
INTRODUCTION
Susceptibility-weighted
imaging (SWI) is widely used in clinical settings to characterize hemorrhage
and mineralization1-3. Nonetheless, its application might be limited
by the long acquisition time, especially in patients who cannot tolerate long
scanning times or are in urgent need of treatment (e.g., stroke). Recently,
some deep-learning (DL) based SWI acceleration methods4, 5 have been proposed, but few have performed
validation in real clinical settings.METHODS
The
study protocol has been approved by the ethics committee of the second affiliated
hospital of Zhejiang University School of Medicine. A total of 31 subjects were
consecutively enrolled in the present study. A DL
accelerated SWI sequence was performed on each subject using a 24-channel
head-neck coil on a uMR790 scanner (United Imaging, Shanghai, China). The
parameters for the DL-SWI acquisition were TR=28.2ms, TE=20ms, Flip anlge=15°, matrix=336*292, voxel size=0.685*0.685*2mm3,
acceleration factor=5. As a reference, the routine parallel imaging (PI)
accelerated (AF=2) SWI scan with otherwise identical scanning was also
performed. The acquisition times for both DL and PI accelerated SWI sequences
were 1min46s and 4min45s, respectively. Two radiological technicians, XW (10
years of experience) and SX (9 years of experience) independently performed the
image quality assessments on all images by a double-blinded procedure. Firstly,
the observer evaluated the following image features: level of noise, lesion
conspicuity (when present), anatomic details (brain structures, tissue
boundaries), artifacts (motion and parallel imaging artifacts), and overall
image quality. The scoring was from 1 to 5 for each index, representing quality
from poor to excellent. For subjects with microbleeds, the observers also
counted the total number of microbleeds. Secondly, both SWI image sets were put
side-by-side and the observers compared the two images in each subject. With
scores 1 to 5, the DL-accelerated SWI images were rated as worse (1-2), equal
(3), or better than the PI SWI counterparts (4-5). The inter-rater correlation
coefficient was used to assess the consistency between the two observers. The
Wilcoxon signed-rank tests and student’s t-tests were used to test the
difference in ordinal and continuous variables, respectively. A difference
student t-test was used to investigate the noninferiority of the DL-accelerated
images (noninferiority margin, -0.5).
RESULTS
The
demographics can be seen in Table 1. Compared to the fully sampled images, DL-accelerated
images had significantly higher scores in terms of noise, artifacts, and
overall image quality score (p < 0.001, p < 0.001, p = 0.046,
respectively). The fully sampled images have higher scores than the DL-accelerated
images in terms of the sharpness of the image (p < 0.001). There was no
significant difference between fully sampled and accelerate images in terms of
lesion conspicuity (p = 0.564[HP1] ).
Additionally, the number of microbleeds identified on the DL-accelerated images
completely matched the fully sampled image. Noninferiority testing showed that
the DL-accelerated images (p=0.407; p=0.385).
DISCUSSION
Based
on our preliminary observations in a clinical setting, the DL accelerated SWI
showed increased signal-to-noise and fewer imaging artifacts than the routine
PI accelerated SWI with identical scan settings. The anatomical details of the
lesions were mostly kept, and we had not observed any false-negative or false-positive
lesions. The potential caveat of this DL accelerated SWI method is the
decreased sharpness at the current resolution setting (i.e. 0.685*0.685*2mm3).
As the DL reconstruction effectively filters out the highest frequency
component collected, due to its Poison Disc style sampling design with more sparse
sampling at the outer k-space part, the filtered spatial frequency components at
the current resolution setting are those most sensitive to the human eye. This
situation can be effectively improved by scanning with higher resolution such
as 0.5x0.5x2mm3, which will be the interest of our future study. However,
we found that the current reduction in sharpness did not interfere with
clinical diagnosis, but with the benefits of greatly improved patient comfort
and scan efficiency due to the significantly reduced scan times. CONCLUSION
With
significantly reduced scanning time, less imaging artifacts, and mostly kept
anatomical details, SWI imaging with DL acceleration could be useful for
clinical situations that need timely imaging results.Acknowledgements
No acknowledgement found.References
1. Thomas B, Somasundaram S, Thamburaj K, et
al. Clinical applications of susceptibility weighted MR imaging of the brain -
a pictorial review. Neuroradiology 2008;50:105-116.
2. Halefoglu AM, Yousem DM. Susceptibility weighted imaging:
Clinical applications and future directions. World J Radiol 2018;10:30-45.
3. Cheng AL, Batool S, McCreary CR, et al. Susceptibility-weighted
imaging is more reliable than T2*-weighted gradient-recalled echo MRI for
detecting microbleeds. Stroke 2013;44:2782-2786.
4. Duan C, Xiong Y, Cheng K, et al. Accelerating
susceptibility-weighted imaging with deep learning by complex-valued
convolutional neural network (ComplexNet): validation in clinical brain
imaging. Eur Radiol 2022;32:5679-5687.
5. Kim EH, Choi MH, Lee YJ, Han D, Mostapha M, Nickel D. Deep
learning-accelerated T2-weighted imaging of the prostate: Impact of further
acceleration with lower spatial resolution on image quality. Eur J Radiol
2021;145:110012.