Haonan Zhang1, Qingwei Song1, Jiazheng Wang2, and Ailian Liu1
1Department of Radiology, the First Affiliated Hospital of Dalian Medical University, Dalian, Dalian, China, 2Philips Healthcare, Beijing, China, Beijing, China
Synopsis
Keywords: Brain Connectivity, Brain
Compared with
echo planar imaging diffusion weighted iamging (EPI-DWI), turbo spin echo
diffusion weighted imaging (TSE-DWI) can significantly reduce magnetic
sensitivity artifacts in skull base imaging. However, the longer scan time
limits its clinical promotion. The purpose of this study is to investigate the
effect of the compression sensing acceleration factor on the image quality of
TSE-DWI in the skull base area.
Introduction
Diffusion
weighted imaging (DWI), as one of the most commonly used MR functional
sequences for the diagnosis of brain diseases, can non-invasively quantify the
diffusion characteristics of water molecules using ADC1. The
traditional DWI sequence uses echo planar imaging (EPI) technology, which can
complete whole brain imaging in a short time. However, due to the variation of
the magnetic field in the skull base area, it is prone to susceptibility artifacts
and misdiagnosis2. DWI imaging based on turbo spin echo sequence
(TSE) can eliminate magnetic sensitivity artifacts, but the scan time is
relatively long, which limits its clinical promotion3. Compressed
sensing (CS) can significantly shorten the scan time through sparse sampling,
while ensuring the image quality to meet the diagnostic requirements4.
The purpose of this study is to explore the effect of the compressed sensing
acceleration factor on the image quality of the fast spin echo diffusion
weighted imaging in the skull base area.Materials and methods
The
study population consisted of 22 healthy subjects(8 females, 53.4±18 years), 22
subjects with dentures(8 females, 57.8±19.7 years) and 11 subjects with skull
base lesions(8 females, 45.1±19.7 years), defined as group A, B, and C,
Respectively. All subjects underwent the EPI-DWI and TSE-DWI scan of Skull Base
on a 3.0 T MR scanner (Ingenia CX, Philips Healthcare, Best, the Netherlands). TSE-DWI
adopted a routine clinical setup of SENSE acceleration factor 2.6, and CS
factors of 2.6, 3.6, 5.6 and 7.6. Other scan parameters were shown in Table 1. In
the b=1000 image, the largest level of the cerebellum and the upper and lower
levels of the cerebellum and the brain stem and pons on the left and right
sides of the brain parenchymal regions delineate the regions of interest to
measure the ADC value and signal intensity. In group C, ROI was delineated to
measure the ADC value and signal intensity at the largest layer of the b=1000
image lesion and the upper and lower layers. The signal intensity and noise
intensity of the dentate nuclei on both sides are measured at the largest level
of the dentate nucleus, based on which Signal to Noise Ratio (SNR) and Contrast
to Noise Ratio (CNR) were calculated for all volunteers. The two observers used
a four-point scoring method to evaluate the quality of the images. The Kappa
test was adopted to evaluate the consistency of the scores from the two
radiologists. If the consistency
was in good agreement, the corresponding images would be adopted for further
analysis by senior physicians. In the following analysis, the Friedman
test was used to test the differences of the cerebellum, brainstem and pons,
ADC value, SNR, CNR, and subjective scores on the two sides of TSE-DWI in
groups A, B, and C, and the Wilcoxon test was employed to make a pairwise comparison. The optimal AF is selected
subsequently for comparison with EPI-DWI.
This study has been approved by the local IRB.Results
The scores by
the two observers were in good agreement (kappa=0.772,
0.831, 0.828). Comparing TSE-DWI with different AF, there was no significant
difference in ADC between the right cerebellum and pons in group A, and the ADC
on the right cerebellum and left in group B. Pairwise comparison, when AF is
CS5.6 in group A, the left cerebellar CNR, right pons SNR, CNR and image scores
have statistically significant differences compared with SENSE2.6 (p<0.05);
at same AF in group B, the left cerebellum and pons brainstem SNR, CNR, right
cerebellum and pons brainstem CNR and image scores were significantly different
than SENSE2.6 (p<0.05); when AF is CS5.6 in group C, there was a
statistically significant difference between CNR and SENSE2.6 (p<0.05).
Subsequent comparison between TSE-DWI CS3.6 and EPI, there was a statistical
difference in the scores of groups B and C(p<0.05), but the score of TSE-DWI
CS3.6 was higher than that of EPI-DWI.Conclusions
Compared
with EPI-DWI , TSE-DWI can significantly improve the geometric distortion
artifacts of the skull base. The TSE-DWI combined with CS gradually shortens
the scanning time as the acceleration factor increases. When CS=3.6, compared
with SENSE=2.6, the scan time is shortened by 21.92%. For subjects with
dentures, it is recommended to use CS=3.6 to scan the skull base with TSE-DWI
sequence.Acknowledgements
No acknowledgement found.References
1. Kolff-Gart, AS; Pouwels, PJ; Noij, DP, et al.
Diffusion-weighted imaging of the head and neck in healthy subjects:
reproducibility of ADC values in different MRI systems and repeat sessions.AJNR
Am J Neuroradiol,2015,36(2):384-390.
2. Wan Q, Lei Q, Wang P, et al. Intravoxel Incoherent Motion
Diffusion-Weighted Imaging of Lung Cancer: Comparison Between Turbo Spin-Echo
and Echo-Planar Imaging. J Comput Assist Tomogr,2020,44(3):334-340.
3. Elefante, A; Cavaliere, M; Russo, C, et al. Diffusion
weighted MR imaging of primary and recurrent middle ear cholesteatoma: an assessment by readers with different
expertise. Biomed Res Int,2015,2015:597896.
4. Bratke, G; Rau, R; Weiss, K, et al. Accelerated
MRI of the Lumbar Spine Using Compressed Sensing: Quality and Efficiency. J
Magn Reson Imaging,2019,49(7):e164-e175.