Tobey D Haluptzok1, Simon Schmidt1, Russell L Lagore1, and Gregory J Metzger1
1Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States
Synopsis
Keywords: RF Arrays & Systems, High-Field MRI
To more fully realize the benefit of UHF MRI, coil designs must be
optimized to for pTx and intrinsic SNR performance. In this abstract we compare
5 RF coil array block systems. The first comparison ran is pTx performance, the
second comparison is of intrinsic array SNR, and the third tests robustness to
top loading. The shielded and unshielded three-loop dipole elements performed better
than the single-loop dipole and the shieled helped with top load insensitivity.
From these results we conclude that the three-loop dipole coil blocks have superior
performance and that the shield doesn’t hurt performance.
Purpose
UHF MRI (>= 7T) has the promise of increased signal to noise and
contrast to noise ratios. However, to more fully realize the theoretical SNR
and CNR gains, coil arrays must be optimized for parallel transmit (pTx) and
intrinsic SNR. In this abstract we compare the pTx and intrinsic SNR
performance of five different RF coil block setups. Specifically, we propose
the addition of an RF shield that allows for on-board electronics and evaluate
the affect this RF shield has on pTx and intrinsic SNR performance. Methods
Five
different RF coil blocks were constructed and tested in a 7T MRI scanner. These
blocks consisted of a single loop-dipole (LD) block, a loop-dipole with an RF
shield (LD-SH), a 3loop-dipole (3LD), a 3loop-dipole with an RF shield
(3LD-SH), and a LD block from an existing coil array for reference(LD-R)1.
The RF shield was constructed from a two-sided flexible PCB with 9µm copper
thickness and 25µm substrate thickness. 12mm fingers separated by 1mm slots were etched in the shield to minimize eddy
currents2. Since the goal of
adding an RF shield is to allow for on-board electronics, a motherboard containing
preamplifiers and transmit receive (TR) switches was created and spaced 2cm
above the coil elements. For the non-shielded elements, one meter of coaxial
cable (K_02252-16, Huber+Suhner) was used to connect the RF antennas to the
same motherboard to simulate the existing body coil array setup with off-board electronics.
pTx Performance comparison: To compare the pTx performance, a
2D multi-slice B1 relative scan was acquired to obtain relative transmit (B1+)
and receive (B1-) profiles3. The channel-wise B1+ maps were
normalized to the same input power and used to calculate phase only efficiency
shims4 with 5x5x5cm3 and 4x4x15cm3 shimming
volumes at different depths.
The
differences in the RF coil blocks are evaluated based on the mean B1+
at different depths (Figure 1).
SNR Analysis: One of the theoretical advantages of on-board
electronics is the decrease in noise figure (NF) resulting from the elimination
of additional cabling between the coil and the preamplifier. The preamplifiers
used for this experiment is a SPF5122Z (Qorvo). To measure SNR, an AFI, proton
density weighted GRE, and noise scan were collected and used to calculate normalized
intrinsic SNR maps for each coil block5. Histograms
along with the total SNR and median voxel SNR were then used to compare the RF
coil blocks (Figure 2).
Additional
considerations: Along with the SNR and pTx performance of this coil block, another
important factor to consider is the robustness to different top loading
conditions. To quantitatively test this, the 3LD and the 3LD-SH were simulated
on a body phantom, both with and without a top load. The loops and dipole were
impedance matched to 50ohms at 297MHz in the simulation without a top load and
this matching circuit stayed the same for the top loaded simulation. The
reflection coefficient for each coil port was calculated with the results shown
in Figure 3. Results
The proposed 3LD-SH block had better pTx (Figure 1) and SNR (Figure 2)
performance than all other blocks tested. Likewise, both 3LD blocks performed
better than the single LD blocks in both pTx and SNR. Figure 3 indicates that
top loading the 3LD-SH block had less of an affect than top loading the
unshielded 3LD block i.e., there was a greater Q broadening and a larger
frequency shift on the unshielded block.Discussion
While the 3LD-SH block outperformed the other blocks, the SNR difference
between it and the 3LD is only minor. When looking at the normalized median SNR
in Figure 2b, there is only a 3% difference between the two. It is feasible
that this 3% difference could have resulted from the mask used to filter out
low signal regions i.e., the mask artificially favored the 3LD-SH block. A more
distinct difference is noted when the 3LD and LD blocks are grouped together.
When comparing these two groups, the 3LD blocks perform around 30% better pTx
performance and have around 15% higher median SNR. Along with the superior SNR
and pTx performance, having a three-LD block design will allow for parallel
imaging along the magnet bore direction. Future studies will involve evaluating
the SNR and pTx performance of a full 16-transmit 32-recieve 3LD-SH coil array.Conclusion
The pTx performance and intrinsic SNR was compared across 5 unique RF
coil blocks. The motherboard that housed the pre-amplifiers and TR switches
stayed the same for each of these blocks to eliminate active electronic
variability and the effect of adding an RF shield was tested. Our results show
that both three-LD blocks outperform the single-LD blocks in both pTx
efficiency and SNR, and that the addition of an RF shield doesn’t degrade coil performance.Acknowledgements
Funding was provided by NIH P41 EB027061 and NIH R01 EB029985.References
1. Ertürk,
A., Raaijmakers, A., Adriany, G.,
Uğurbil, K., & Metzger, G. A 16-channel combined loop-dipole
transceiver array for 7 Tesla body MRI. Magn
Reson Med, 2016; 77(2):884-894.
2. Lechner-Greite,
S., Hehn, N., Werner, B., et al. Minimizing eddy currents induced in the ground
plane of a large phased-array ultrasound applicator for echo-planar
imaging-based MR thermometry. J. Ther. Ultrasound, 2016; 4(1):1-14
3. Brunner, D., Pruessmann, K., SVD Analysis of Array
Transmission and Reception and its Use for Bootstrapping Calibration. Magn
Reson Med, 2016; 76(6):1730-1740.
4. Metzger,
G., Snyder, C., Akgun, C., Vaughan, T., Uğurbil, K., and Van de Moortele P.F.
Local B1+ shimming for prostate imaging with transceiver arrays at
7T based on subject-dependent transmit phase measurements. Magn Reson Med, 2008: 59(2):396-409.
5. Van
de Moortele, P. F., Akgun, C., Adriany, G., Moeller, S., Ritter, J., Collins,
C. M., ... & Uğurbil, K. (2005). B1 destructive interferences and spatial
phase patterns at 7 T with a head transceiver array coil. Magn Reson Med, 54(6), 1503-1518.