Christoph Stefan Aigner1, Jean Pierre Bassenge1,2, Sebastian Dietrich1, Max Lutz1, Felix Krüger1, and Sebastian Schmitter1,3,4
1Physikalisch-Technische Bundesanstalt (PTB), Berlin and Braunschweig, Germany, 2Working Group on Cardiovascular Magnetic Resonance, Experimental and Clinical Research Center, a joint cooperation between the Charité Medical Faculty and the Max-Delbrück Center for Molecular Medicine, Berlin, Germany, 3Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 4University of Minnesota, Center for Magnetic Resonance Research, Minneapolis, MN, United States
Synopsis
Keywords: RF Pulse Design & Fields, Parallel Transmit & Multiband
3D ultrahigh field (UHF) imaging of the body, particularly
the heart, is highly challenging due to the inhomogeneous transmit field (B
1+) leading to spatially
varying flip-angle (FA) patterns. This study demonstrates that pre-computed universal
pulses (UPs) allow for calibration-free 3D cardiac FA homogenization at 7T,
despite inter-subject variations because of sex, age, BMI, and coil-placement
differences. Furthermore, the performance is robust across multiple MRI operators or
re-scans after one year. UPs are, therefore, suited to significantly simplify UHF cardiac imaging
by removing the need for lengthy calibration prescans.
Introduction
Ultrahigh field (UHF; >3T) MRI of the human body is
limited by variations in spatial flip angle (FA) patterns.1 In
contrast to subject-tailored parallel transmission (pTx) methods2,3,
requiring subject-specific B1+-maps4,
we demonstrated that calibration-free cardiac universal pulses (UPs) are
applicable in the body, thus saving valuable scan time.5 When
generating the B1+-library for cardiac UPs calculation,
we found substantial variations in the channel-wise B1+-distributions
among subjects due to i) different body sizes and shapes and ii) different
placements of the pTx coil on the body. Due to this effect, our UP
library for the heart required more subjects than the brain UP counterpart.6
Interestingly, however, UPs calculated for the heart resulted in good FA homogeneity within the heart. UPs also performed well in surrounding
tissue, such as the aorta, which was unexpected but consistently observed in all
volunteers.5 This indicates a certain robustness of the UPs also for
different placements of the RF coil, which is needed for clinical applications.
In this work, we analyze the reproducibility and variability of three different
excitations in seven subjects: default excitations, subject-tailored pTx and
calibration-free UPs. To this end, a different MRI operator re-scanned the
seven subjects after one year.Methods
MRI was performed on a Magnetom 7T scanner (Siemens
Healthineers, Erlangen, Germany) with a certified commercial 32-element body
array (MRI.TOOLS, Berlin, Germany) driven in 8Tx/32Rx mode. 36 channel-wise 3D B1+-datasets
were acquired under free-breathing2 and were divided into three
groups: i) UP-library (group1: 14M/8F, 21-66years, 19.8-28.3kg/m2), ii)
unseen test-cases (group2: 3M/4F, 25-56years, 19.5-35.3kg/m2) and
iii) re-scans of unseen test-cases (group3: same group as group2; re-scanned
after one year). Subject-tailored 4kT-point pulses (TP) were designed for each
dataset of group2, and a 4kT-point UP was calculated using 22 B1+-datasets
of group1.5 The TPs/UP were then re-applied to group3 approximately
one year later. The posterior half of the coil was centred on the volunteer's
heart in the head-to-foot direction, while the anterior half was placed 2cm
from the volunteer's chin based on anatomical landmarks and volunteer feedback.
A different MRI operator performed the coil placement of group3.
The reproducibility and variability of three
excitations were assessed among groups2&3: the default excitation (a
default shim for the heart set by the vendor), TP and UP. Reconstruction of B1+-datasets,
manual slice-by-slice selection of 3D cardiac ROIs, design of RF pulses, and generation
of pulse files was performed on a separate workstation. The pulses' performance
was analyzed using FA predictions and coefficient-of-variations (CV) in the
heart volumes.3 The source code and the B1+-datasets
can be downloaded from https://github.com/chaigner/UP_body.Results and Discussion
Fig.1 shows relative channel-wise 3D B1+-datasets
of the thorax and the FA prediction of three different pTx settings (default,
tailored and UP) in the human heart for one representative B1+-datasets
of group1 (UP-library).
Fig.2 shows the predicted FAs and the CV in the heart
volume using subject-tailored 4kT-points pulses (TP). Note the substantial
differences in spatial FA variations resulting from a variation of the heart
size, heart position, BMI, and coil placement between the datasets. Applying
the same TP to group3 resulted in elevated FA spread (Fig.2B) and increased CV
values (Fig.2C) in all re-scans. The median CV of 6.3% (test-cases) increased to
19.73% (re-scans) demonstrating the need to re-acquire B1+-maps
and re-calculate the pTx-pulses in the TP case. Moreover, the TP resulted in substantial
FA dropouts in the heart in re-scan #35, likely caused by different coil
placement.
Fig.3 shows the FA prediction using the default RF
shim and the UP for a sagittal slice of the 3D volume. In all test-cases and re-scans,
the default RF shim resulted in FA dropouts in the FA predictions that were prevented
by applying the UP. This was also the case for re-scan (#35), with substantial
differences in the coil placements.
Fig.4A shows the quantitative FA predictions in the
heart volume using two pTx settings (default and UP). The default RF shim and the
UP achieve a median FA close to the target FA of 10° but result in a diverse FA
spread. As expected, the UP outperforms the default RF shim with higher minimal
FAs and a reduced FA spread from 6.2° to 1.8° (interquartile range). Fig.4B-C
show the CV evaluation in the heart volume. Across all 36 datasets, the UP reduces
the CV below 20% with a median CV of 12.4% (range=8.2%-18.6%) compared to the
default phase setting with a median CV of 42.8% (range=23%-64.8%). Across the
different B1+-dataset groups, the UP consistently
delivers CV values of approximately 12% (UP-library: 12.7%, test-cases: 11.8%
and re-scans: 11.5%).
Figure 5 shows exemplary 3D GRE images (without
cardiac gating) from four re-scans using the proposed UP. Close agreement was
observed between the B1+-predictions and the 3D GRE
images, demonstrating coil-placement robustness of calibration-free pTx in the
human heart.Conclusion
Despite substantial B1+-variations
between scan and re-scan, the work shows that pre-computed UPs provide
robustness not only against FA variations among subjects but also under changes in coil placements. This highly valuable property was not the case for
the static default shim or the TPs and makes the UP suitable for
calibration-free 3D body applications at 7T. Acknowledgements
We gratefully acknowledge funding from the
German Research Foundation SCHM 2677/2-1, SCHM 2677/4-1 and GRK2260, BIOQIC.References
[1] Ladd, E, Bachert, P, Meyerspeer, M, et al. Pros
and cons of ultra-high-field MRI/MRS for human application, Prog. Nuc. Magn.
Reson. Spec. 2018; 109:1-50. doi: 10.1016/j.pnmrs.2018.06.001
[2] Padormo, F, Beqiri, A, Hajnal, JV, and Malik, SJ.
Parallel transmission for ultrahigh‐field imaging. NMR Biomed. 2016; 29: 1145–
1161. doi: 10.1002/nbm.3313
[3] Aigner, CS, Dietrich, S, Schmitter, S.
Three-dimensional static and dynamic parallel transmission of the human heart at
7 T. NMR in Biomedicine. 2021; 34:e4450. doi: 10.1002/nbm.4450
[4] Dietrich, S, Aigner, CS, Kolbitsch, C, et al. 3D
Free-breathing multichannel absolute B1+ Mapping in the human body at 7T. Magn
Reson Med. 2021; 85: 2552– 2567. doi: 10.1002/mrm.28602
[5] Aigner, CS, Dietrich, S, Schaeffter, T, Schmitter,
S. Calibration-free pTx of the human heart at 7T via 3D universal pulses. Magn
Reson Med. 2021; 87: 70– 84. doi: 10.1002/mrm.28952
[6] Gras, V,
Vignaud, A, Amadon, A, Le Bihan, D and Boulant, N Universal pulses: A new
concept for calibration‐free parallel transmission. Magn. Reson. Med. 2017; 77:
635-643. doi: 10.1002/mrm.26148