4569

Vector concomitant field mapping by phase contrast imaging
Seung-Kyun Lee1 and Afis Ajala1
1GE Global Research, Niskayuna, NY, United States

Synopsis

Keywords: Gradients, Gradients, concomitant field

We show that concomitant field-induced phase in phase-contrast imaging allows measurement of the transverse (x,y) components of the magnetic field generated by a gradient coil. Combined with conventional B0 mapping which gives the z-component, the method permits experimental determination of the full three-dimensional vector magnetic field of a gradient coil. We demonstrate the method on a high-performance head gradient (MAGNUS) system and discuss its application to concomitant-field artifact correction based on one-time calibration.

Introduction

In MRI, concomitant field refers to the transverse (Bx, By) components of the magnetic field that necessarily accompany spatial variation of the z-directional magnetic field (Bz) generated by the gradient coils during image encoding. The concomitant field has been most extensively studied for its effects on image artifacts [1], but it also contributes to peripheral nerve stimulation (PNS) [2] and implant heating [3]. Despite its importance, direct measurement of Bx, By fields is not easy because traditional field mapping methods measure the total magnetic field which is dominated by Bz. Here we show that a balanced bipolar gradient pulse can encode the transverse magnetic field into the image phase, and strategic acquisition of such data followed by azimuthal symmetry-based post-processing allows us to map out the full vector magnetic field of a gradient coil in 3D space.

Theory

Isolation of transverse field phase. Consider a gradient coil which produces an x-directional field BT as well as a z-directional one Bg at a particular voxel. The total magnetic field vector (Btot) at the voxel consists of the gradient field vector, main magnetic field B0, static off-resonance field Bs, and the eddy current field Be (Eq. (1)). The magnitude of Btot can be expressed as Eq. (2) where we retained terms up to the second order in the small parameter δ. Importantly, in Eq. (2) the longitudinal (Bz,all) and transverse (BT) field terms are separated. Eq. (3) shows the instantaneous Larmor frequency. When a bipolar gradient pulse is applied for time Δt, the voxel accrues phase due to Bs, Be, BT (Eq. (4)). The Be term reverses sign if the opposite-polarity pulse is applied (Eq. (5)), and the Bs term is isolated if the pulse is zeroed (Eq. (6)). The three phases φ+, φ-, φ0 can be combined to compute φc (Eq. (7), Fig. 2) which gives the desired magnitude of BT (Eq. (8)).
3D vector field calculation. The azimuthal (m=1) symmetry of a transverse gradient coil implies that two measurements of BT, on the coronal and sagittal midplanes, are sufficient to calculate the transverse magnetic field anywhere in a cylindrical volume. When this information is combined with the longitudinal field Bz that can be measured by phase induced by an unbalanced (unipolar) gradient pulse, we can reconstruct a 3D vector field map of the gradient coil (Fig. 3). For an axi-symmetric (m=0) longitudinal gradient coil (Gz), BT measurement on one (e.g. coronal) plane suffices to enable 3D reconstruction.

Experimental Methods

All experiments were performed with a high-performance head gradient coil (MAGNUS)[4] inserted in a whole-body 3T magnet, using a cylindrical phantom and a volume T/R coil. Figure 2 shows the schematic of the modified 2D gradient echo-based phase-contrast imaging sequence used to produce the three phase images. Representative scan parameters are TR/TE/Δt = 200/8.3/4.4 ms; FOV/ slice thickness/ isotropic pixel size = 260/5/1.35 mm. Real-time concomitant field correction [5] was disabled. Two more phase images were obtained for Bz mapping. The phase images were processed following the workflow of Fig. 3. The experiment was repeated for x, y, z gradient coils. The resulting vector field maps were compared with the gradient coil's electromagnetic (EM) design.

Results

The middle column of Fig. 4 shows the measured vector field maps of the x- and z-gradient coils. The maps visually agree well with the EM-design field maps shown on the left column. Our data enable direct visualization of the strong concomitant fields of an asymmetric (x) gradient coil, which is evident when compared with the z-component-only map (right column). The field maps for the y-coil were similar to the x-coil, rotated by 90 degrees in the xy-plane. Quantitative comparison showed that the measurement error (compared to the EM design) was less than ~5% within 5 cm from the isocenter, but gradually increased with distance (not shown). The discrepancy is possibly due to vibration-induced phase that is quadratic to the bipolar pulse amplitude, which is not cancelled by phase combination of Eq. (7).

Discussions and Conclusion

Full knowledge of the vector magnetic field generated by the gradient coils allows one to model spatio-temporal evolution of the total encoding magnetic field during imaging. This enables prediction of the concomitant-field-induced phase evolution at every voxel, which can inform prospective correction mechanisms (such as gradient pre-emphasis) or artifact correction in post-processing (Fig. 5). The proposed method can also be used to obtain full vector magnetic field data for electric field calculation, for the purpose of PNS simulation or eddy current assessment, without resorting to gradient coil EM design.
To date, experimental concomitant field measurement has fallen under two categories: (i) low spatial-order parameter fitting under linear gradient assumption [6], and (ii) dynamic Larmor frequency measurement by an NMR probe array for individual sequences [7]. The proposed method is advantageous in that it is applicable to nonlinear gradients and allows sequence-independent calibration. In conclusion, we have demonstrated an experimental method to measure the three-dimensional vector magnetic field produced by a gradient coil. The experimental data showed good agreement with the EM design and hold promise for data-driven correction of concomitant field artifacts in high-performance gradient systems.

Acknowledgements

This work was supported by CDMRP W81XWH-16-2-0054. This presentation does not necessarily represent the official views of the funding agency.

References

[1] M.A. Bernstein et al., Concomitant gradient terms in phase contrast MR: analysis and correction. Magn Reson Med 39(2):300-308 (1998)

[2] S.S. Hidalgo-Tobon, M. Bencsik, and R. Bowtell, Reducing peripheral nerve stimulation due to gradient switching using an additional uniform field coil. Magn Reson Med 66(5):1498-1509 (2011)

[3] A. Arduino et al., A contribution to MRI safety testing related to gradient-induced heating of medical devices. Magn Reson Med 88(2):930-944 (2022)

[4] T.K.F. Foo, et al., Highly efficient head-only magnetic field insert gradient coil for achieving simultaneous high gradient amplitude and slew rate at 3.0T (MAGNUS) for brain microstructure imaging Magn Reson Med 83:2356-2369 (2019)

[5] S. Tao, et al., Gradient pre-emphasis to counteract first-order concomitant fields on asymmetric MRI gradient systems. Magn Reson Med 77(6):2250-2262 (2017)

[6] N. Abad, et al., Calibration of concomitant field offsets using phase contrast MRI for asymmetric gradient coils. Magn Reson Med, doi:10.1002/mrm.29452 (2022)

[7] B.J. Wilm et al., Minimizing the echo time in diffusion imaging using spiral readouts and a head gradient system. Magn Reson Med 84:3117-3127 (2020)

Figures

Figure 1. Equations for the proposed, phase-based transverse magnetic field measurement.

Figure 2. Acquisition and processing workflow to experimentally determine the transverse magnetic field per unit current (bT BT/I) produced by a gradient coil.

Figure 3. Full workflow to determine the 3D vector field map of a Gx gradient coil. All imaging scans are performed in 2D, while the azimuthal (cosine) symmetry of the Gx coil permits computation of the magnetic field in 3D. Similar workflow applies to vector field mapping of Gy, Gz gradient coils.

Figure 4. As-designed vs measured vector magnetic field maps of asymmetric (x-axis) and symmetric (z-axis) coils of the head-insert gradient coil.

Figure 5. Envisioned concomitant field correction workflow based on measured vector magnetic fields and gradient current waveforms.

Proc. Intl. Soc. Mag. Reson. Med. 31 (2023)
4569
DOI: https://doi.org/10.58530/2023/4569