Marc Dubois1, Amira Trabelsi1,2,3, Tania Vergara Gomez1,2,3, Pierre Jomin3, Djamel Berrahou1, David Bendahan2,4, Frank Kober2,4, Virginie Callot2,4, Stefan Enoch3, and Redha Abdeddaim3
1Multiwave Imaging, Marseille, France, 2Aix Marseille Univ, CNRS, CRMBM, Marseille, France, 3Aix Marseille Univ, CNRS, Centrale Marseille, Institut Fresnel, Marseille, France, 4APHM, Hopital Universitaire Timone, CEMEREM, Marseille, France
Synopsis
Keywords: New Devices, Spinal Cord, Metamaterial, Metasurface
Spine imaging is usually conducted at 3T given the need of high spatial
resolution. Unfortunately, the lumbar and thoracic spine regions of interest
will lie in areas of weak transmit efficiency. Recent studies showed that high
permittivity ceramic blocks can improve image quality. Here we present a new
approach based on ultra-thin, lightweight, flexible metasurface pads yielding
improved efficiency and homogeneity of the B
1+ field along the spine in
clinical 3T MRI settings. Our design offers seamless integration to regular
equipment leading to more efficient and safer spine imaging protocols.
Introduction
3T MR scanners are popular in a clinical context, offering higher SNR
compared to the widely deployed 1.5T. Higher field strength leads to shorter
scanning times or higher resolution images especially sought for spine imaging
[1]. However, when imaging large cross-sectional areas such as the abdomen or
the thorax, inhomogeneities of the radiofrequency field transmitted by the body
coil generates a skewed flip angle distribution affecting image contrast [2].
High dielectric constant pads have been used as an efficient solution to
mitigate radiofrequency inhomogeneities as presented in a recent review article
[4]. Progress in MRI metamaterials has gained significant traction allowing
more degrees of freedom in their design compared to traditional dielectric pads
[5-7]. Focusing on 3T spine imaging, Koolstra et al. demonstrated that the use lead zirconate
titanate blocks with relative permittivity reaching over a thousand can simultaneously
improve image quality and lower power deposition [3]. However, the production
of such materials is very limited, even in terms of dimensions. Moreover, the
seven blocks used in the study were 3-cm thick, weighed more than a kilogram
each and had to be placed carefully on the patient table. Here, we present an
alternative approach based on ultra-thin, lightweight, and flexible metasurface
pads. Our design offers a seamless integration into standard MRI equipment, as
well as a robust and cost-effective solution leading to more efficient and
safer 3T spine imaging protocols.Materials and Methods
The proposed metasurface works under the resonance hybridization
principle [6,8]. A series of parallel conductors were printed on a thin
polyimide substrate (200mm) and sealed in a polymer-coated fabric
pouch. The conductors are inductively coupled to the body birdcage coil and
redistribute the B1+ field [6]. Figure 1
presents photographs of the WearMe metasurface. A single pad was placed on top
of the patient table beneath the volunteer. Experiments were performed on a
single volunteer (Male, 193cm, BMI 29 kg/m2) in a 3T Vida MRI
scanner (Siemens Healthineers, Erlangen, Germany). Twelve channels of the
integrated table coils were activated during the experiments.
Flip Angle (FA) maps were acquired at the thoraco-lumbar levels with a
turbo-flash (TFL) MRI sequence [9] in sagittal orientations with a 650 V
reference voltage, TR/TE= 5000/1.8ms, FA= 8°, FOV= 250x400 mm,
matrix= 80x128 and 13 slices of 15-mm thickness. Raw images were used to
delineate the position of the spinal cord and to report the region of interest
in the FA maps. T1-weighted and T2-weighted images were also acquired with
recalibrated reference voltage based on the FA maps (T1-w 2D-TSE: TR/TE=
641/9.6ms, ETL= 3, Acceleration factor= 3, FOV= 238x400 mm,
matrix= 210x352, 30 slices of 5-mm thickness; T2-w 2D-TSE HASTE (80%
partial Fourier) TR/TE= 2000/97ms, ETL= 109, FOV= 325x400 mm,
matrix= 208x256, 30 slices of 5-mm thickness.Results
FA maps and profiles are shown in Figure 2. Along the spinal cord, a 36%
increase in FA was observed with the WearMe configuration for a given reference
voltage (Fig 2b). The homogeneity of the FA distribution was evaluated along
the cord using the relative standard deviation (RSD)(Fig.2c). We observed an
RSD of 17% in the reference case and a reduction to 9% with WearMe. FA maps
were used to normalize the reference voltage to an 80° target angle in the
lumbar spine: 747V for the reference and 548V with WearMe. Consequently, the
input power was reduced by 47% in the presence of WearMe. SAR simulations (data
not shown) were performed before the in vivo studies and showed that
global and max 10g local SAR were marginally affected by the presence of
WearMe. Furthermore, our results have shown that using WearMe, a similar FA can
be obtained with half the input power thereby limiting power absorbed. High-resolution T2w and T1w images obtained
with and without WearMe are reported in Figure 3 and 4. Each acquisition was
performed with the specific reference voltage given above. A more homogeneous
contrast was observed along the whole field of view, especially going toward
the thoracic spine.Discussion
Our experimental results indicate that the initial FA inhomogeneity
along the spine was reduced by 50% in the presence of WearMe leading to a 9%
RSD along a 350mm profile following the spinal cord. High-resolution images in
Figures 3 and 4 confirmed that the improved FA homogeneity resulted in a better
contrast along the spinal cord with a single acquisition which could be helpful
for an improved detection of signal abnormalities across such a large field of
view. This can also be of interest for a scan time reduction as the imaging
protocol may require multi-stage MRI acquisitions to cover all spine sections.
The performance of the WearMe pad is currently assessed in a larger cohort of
volunteers and in different MRI scanners with various birdcage coil designs.Conclusion
WearMe, the ultra-thin, lightweight, flexible metasurface pad, has been
shown to dramatically improve transmit efficiency and contrast homogeneity over
a 40cm-long field of view in the case of 3T spine MRI. Altogether, our design
is seamlessly integrated with standard MRI equipment and offers a robust and
cost-effective solution to improve and accelerate 3T spine MRI protocols.Acknowledgements
The project leading to this publication has received funding from the
Excellence Initiative of Aix-Marseille University - A*MIDEX, a french
"Investissements d'Avenir" programme.References
[1] Vertinsky,
AT, et al. Neuroimaging Clin N Am 2007; 17:117–136
[2] Dietrich, O, et al. Eur J Radiol 2008;
65:29–35.
3
[3] Koolstra,
K, et al., Magnetic Resonance in Medicine 2018; 79: 1192-1199
[4] Webb, AG, et al. Magnetic Resonance
Materials in Physics, Biology and Medicine 2022; 1-20
[5]
Dubois, M, et al.
Physical Review X 2018; 8.3: 031083
[6] Vergara, T, et al., Proc. Intl. Soc. Mag. Reson. Med. 2021;
29.1405
[7] Vorobyev, V, et al.
Magnetic Resonance in Medicine 2022; 87.1:496-508
[8] Jouvaud, C, et
al. Applied Physics Letters 2016; 108(2):023503.13
[9] Chung, S, et al. Magnetic Resonance in Medicine 2010;64(2):439–446