Vitaliy Atamaniuk1, Jun Chen 2, Kevin Glaser2, Andrii Pozaruk1, Marzanna Obrzut3, Bogdan Obrzut4, Wojciech Domka5, Krzysztof Gutkowski6, Richard L Ehman2, and Marian Cholewa1
1Department of Biophysics, Institute of Physics, College of Natural Sciences, University of Rzeszow, Rzeszow, Poland, 2Department of Radiology, Mayo Clinic, Rochester, MN, United States, 3Center for Diagnostic Medical Sonography, Rzeszow, Poland, 4Department of Obstetrics and Gynecology, Institute of Medical Sciences, Medical College, University of Rzeszow, Rzeszow, Poland, 5Department of Otorhinolaryngology, Institute of Medical Sciences, Medical College, University of Rzeszow, Rzeszow, Poland, 6Institute of Medical Sciences, Medical College, University of Rzeszow, Rzeszow, Poland
Synopsis
Keywords: New Devices, Elastography
A
specialized driver was developed to enable magnetic resonance elastography of
the parotid and other salivary glands using high shear wave frequencies to
provide improved spatial resolution and optimized ergonomics. The performance
of this new device was tested in volunteer studies. Preliminary results showed
that suitable shear wave illumination of the parotid gland is possible at
frequencies as high as 120 Hz and that high-frequency shear wave MR
elastography of the parotid gland is feasible.
Introduction
Tumors of
the salivary glands are most often first detected by noticing a palpable mass. This
provides motivation to explore the potential of MR elastography for characterizing
benign and malignant salivary gland tumors and to address the current limitations
of conventional imaging techniques used to evaluate these lesions.
Parotid
glands are the most common site for the presentation of salivary gland tumors,
accounting for approximately 86% of cases [1].
Previous studies have demonstrated the feasibility and promise of
applying MRE technology to assess the mechanical properties of parotid tissues
[2,3]. The goal of this work described
here was to further advance MRE technology for this application by designing
and testing an ergonomic driver device capable of generating usable shear wave
fields in the glands at higher frequencies than the 60 Hz and 80 Hz values used
in previous works. Higher shear wave frequencies provide shorter shear wavelengths
and improve stiffness measurement accuracy in small lesions.Methods
The design
of the novel passive shear wave driver developed in this project was guided by
considerations of patient comfort, ergonomics, convenience of use, and ability
to localize the applied vibration to the anatomy of interest. The final design of the driver, fabricated
from polyvinyl chloride tubing, is illustrated in figure 1. Volunteer subject
testing of the performance of the novel driver was conducted using 3D vector
MRE acquisition at vibration frequencies of 60, 90, and 120 Hz on a 1.5T
whole-body Signa HDxt scanner (GE Healthcare, Milwaukee, WI, USA). The imaging
parameters were as follows: field of view = 24 cm, repetition time = 3117.6 ms,
echo time = 40.2 ms, spacing between slices = 3 mm, image matrix = 80 × 80,
slice thickness = 3 mm, flip angle = 90°, time steps = 4. A full 3D direct
inversion algorithm was used to process the MRE data. Results
Testing
demonstrated that the novel driver provided excellent shear wave illumination
in the parotid gland at all tested frequencies, including 120 Hz, as
demonstrated in figure 2. The calculated mean stiffness +/- SD in kPa of the
parotid glands was 1.28 +/- 0.28 at 60 Hz, 2.31 +/- 0.36 at 90 Hz, and 3.39 +/-
0.51 at 120 Hz (Fig. 3).Discussion
In order
to obtain reliable measurements of the stiffness of small lesions it is
necessary for the spatial resolution of the acquisition to be sufficient to
account for the spatial footprint of the inversion algorithm and for the
displacement-induced phase shift across that necessarily small spatial
footprint to be sufficiently high. The latter requirement is helped if the
shear wavelength is short as possible across the footprint, which provides motivation
for using higher shear wave frequencies.
A further consideration is that shear wave attenuation in tissue
increases with frequency. Designing the
driver to be placed as close as possible to the region of interest addresses
this factor.Conclusion
MR
elastography at shear wave frequencies as high as 120 Hz is feasible with the
use of the specialized shear wave driver developed in this project. The results provide the basis for planning
studies aimed at assessing the potential role of MRE in characterizing benign
and malignant parotid neoplasms.Acknowledgements
No acknowledgement found.References
1. Alvi S, Chudek D, Limaiem F. Parotid
Cancer. [Updated 2022 Aug 8]. In: StatPearls [Internet]. Treasure Island (FL):
StatPearls Publishing; 2022 Jan. Available from: https://www.ncbi.nlm.nih.gov/books/NBK538340/
2. Yeung D, Bhatia K, Lee Y, King A,
Garteiser P, Sinkus R, Ahuja A. MR elastography of the head and neck: Driver
design and initial results. Magnetic Resonance Imaging, Volume 31, Issue 4,
2013 624-629, ISSN 0730-725X, https://doi.org/10.1016/j.mri.2012.09.008.
3. Elsholtz FHJ, Reiter R, Marticorena Garcia
SR, Braun J, Sack I, Hamm B, Schaafs LA. Multifrequency magnetic resonance
elastography-based tomoelastography of the parotid glands-feasibility and
reference values. Dentomaxillofac Radiol. 2022 Jan 1;51(1):20210337. doi:
10.1259/dmfr.20210337. Epub 2021 Sep 24. PMID: 34558305; PMCID: PMC8693321.