Kai Ruppert1, Luis Loza1, Faraz Amzajerdian1, Hooman Hamedani1, Mostafa K Ismail1, Ryan J Baron1, Ian F Duncan1, Harrilla Profka1, Stephen Kadlecek1, and Rahim R Rizi1
1Radiology, University of Pennsylvania, Philadelphia, PA, United States
Synopsis
Keywords: Hyperpolarized MR (Gas), Pulse Sequence Design
Chemical
shift saturation recovery (CSSR) MR spectroscopy using hyperpolarized xenon-129
provides metrics of pulmonary physiology by saturating the xenon dissolved-phase
magnetization in the lung with a 90° RF pulse and measuring the subsequent
signal recovery via gas exchange. Our measurements in a rat demonstrate that
chemical shift inversion recovery (CSIR) spectroscopy, which replaces the
saturation with an inversion pulse, produces equivalent results to CSSR but
with greater robustness with respect to both low signal amplitudes at short
delay times and incomplete saturation.
Purpose
Chemical
shift saturation recovery (CSSR) MR spectroscopy using hyperpolarized xenon-129
(HXe) has been shown to yield important metrics of pulmonary function in humans
as well as various animal species1-9. This is accomplished by selectively
saturating the magnetization of xenon dissolved in the lung parenchyma at a
chemical shift of approximately 200 ppm with a narrow-bandwidth 90° RF pulse
while leaving the gas phase (GP) magnetization at 0 ppm largely unaffected. Regrowth
of the signal of the membrane (Mem) and red blood cell resonances is measured by
collecting spectra at a variable delay time after the saturation pulse and then
fitting to a theoretical gas exchange model. Especially at short delay times,
however, the dissolved-phase signals are small and potentially contaminated by residual
magnetization due to imperfect saturation, leading to significant relative
errors for these data points. In this work, we investigated whether the use of
chemical shift inversion recovery (CSIR) spectroscopy, which substitutes the
saturation with an inversion RF pulse, can increase measurement accuracy.Methods
A healthy male
Sprague-Dawley rat (340g) was anesthetized, intubated, and mechanically
ventilated using a home-built ventilator with either air (during non-imaging
periods) or a HXe gas mixture (during imaging; 21% O2, 79% N2/HXe) at a tidal
volume of 10 ml/kg and breathing rate of 43 breaths per minute. Imaging was
performed using a 3T horizontal-bore animal imaging system (Bruker Biospec); 1L
of enriched xenon gas was used for each experiment, polarized using a prototype
commercial optical pumping system (XeBox-E10, Xemed LLC, NH). Proton
T2-weighted fast spin-echo images were acquired for localization prior to
spectroscopy experiments. All studies were approved by the Institutional Animal
Care and Use Committee of the University of Pennsylvania.
Data was acquired using a
customized MR spectroscopy pulse sequence. During a 5-s end-inspiratory breath hold, the
dissolved xenon magnetization was either saturated or inverted with a 90° or
180° Gaussian RF pulse (0.913 ms pulse duration), respectively. Following a
variable delay time Δτ
ranging from 1.15 to ~100 ms, a 90° Gaussian RF excitation pulse (0.913 ms
pulse duration) was applied, and a free induction decay was acquired (50 kHz
receiver bandwidth, 31.56 ms sampling duration). The next saturation/inversion
pulse was applied after a fixed gas exchange time of 80 ms. This sequence was
repeated 20 times during the same breath hold for each Δτ. Following Fourier transform, spectra in steady
state were averaged and the areas underneath the gas phase (GP, 0 ppm),
membrane (Mem, 197 ppm), and red blood cell (RBC, 211 ppm) peaks were
numerically integrated. Septal wall thickness and capillary transit time were
calculated using the analytical uptake model of Patz et al.10, based
on the recovery of the averaged membrane signal following each
saturation/inversion pulse.Results and Discussion
Figure 1 depicts the Mem-GP ratios following
either saturation or inversion as a function of Δτ and the fitted Patz model function (solid lines). The
apparent alveolar septal wall thickness was 6.8 μm for both measurement types, with respective
capillary transit times of 0.34 s (CSSR) and 0.17 s (CSIR). However, given that
no data was acquired for delay times longer than 100 ms, these transit times
are not reliable. Although CSSR and CSIR produced identical values for the
septal wall thickness, the inversion measurements appear to be much smoother—and
not just at short delay times, as anticipated, but over the entire Δτ interval, confirming the potential for improved
measurement accuracy when using CSIR rather than CSSR spectroscopy. Additional
studies in a larger animal cohort and human subjects will be required to
validate our observations, but our measurements already indicate the
equivalence of the two acquisition techniques. A potential drawback of CSIR
spectroscopy is the challenge
of implementing it on a human MRI scanner due to the high power requirements of
RF inversion pulses with sufficiently broad bandwidth to invert all
dissolved-phase resonances.Conclusion
Our preliminary study indicates that the
proposed CSIR spectroscopy technique yields pulmonary function metrics
comparable to those obtained with the established CSSR spectroscopy technique
while also being less sensitive to incomplete saturation of the dissolved-phase
resonances and the low signal amplitudes at short delay times.Acknowledgements
Supported by NIH grant
R01 HL142258.References
[1] Ruppert K et
al. NMR of hyperpolarized 129Xe in the canine chest: spectral dynamics during a
breath-hold. NMR Biomed 2000;13:220-228.
[2] Butler JP et
al. Measuring surface-area-to-volume ratios in soft porous materials using
laser-polarized xenon interphase exchange nuclear magnetic resonance. J Phys
Condens Matter 2002;14:L297-L304.
[3] Qing et al.
Assessment of lung function in asthma and COPD using hyperpolarized 129Xe
chemical shift saturation recovery spectroscopy and dissolved-phase MRI. NMR in
Biomed 2014;27(12):1490-1501.
[4] Zhong et al.
Simultaneous assessment of both lung morphometry and gas exchange function
within a single breath‐hold by
hyperpolarized 129Xe MRI. NMR in Biomed 2017; 30(8).
[5] Doganay et al.
Quantification of regional early stage gas exchange changes using
hyperpolarized 129Xe MRI in a rat model of radiation-induced lung injury. Med
Phys 2016;43(5):2410-2420.
[6] Kern et al. Regional
investigation of lung function and microstructure 129Xe chemical
shift saturation recovery parameters by localized and dissolved‐phase
imaging: A reproducibility study. MRM 2019; 81(1):13-24.
[7] Ruppert et al.
Using Hyperpolarized Xenon-129 MRI to Quantify Early-Stage Lung Disease in
Smokers. Acad Radiol 2019; 26(3):355-366.
[8] Kern et al.
Mapping of regional lung microstructural parameters using 129 Xe dissolved‐phase
MRI in healthy volunteers hyperpolarized and patients with chronic obstructive
pulmonary disease. MRM 2018; 81(4):2360-2373.
[9] Zanette et al.
Physiological gas exchange mapping of hyperpolarized 129Xe using
spiral-IDEAL and MOXE in a model of regional radiation-induced lung injury. Med
Phys 2018;45(2):803-816.
[10] Patz et al.
Diffusion of hyperpolarized 129Xe in the lung: a simplified model of 129Xe
septal uptake and experimental results. New J Physics 2011;13:015009.