Seonghwan Yee1
1Radiology, Massachusetts General Hospital, Boston, MA, United States
Synopsis
Keywords: Heart, Relaxometry, MOLLI
The modified Look-Locker imaging (MOLLI) technique, widely
used in cardiac T1 mapping, requires specification of an acquisition timing
scheme. For better quantification of wide range of T1 values, different acquisition
schemes have been used for the prior- and post-contrast conditions. However, the
strategy of having two different acquisition schemes might add complexity to
the consistent management of clinical protocols, especially in a clinical
environment with multiple MRI systems. Hence, a phantom experiment was
performed to examine the necessity of this strategy. The results here suggest
that having a separate acquisition scheme for the post-contrast condition would
not be necessary.
Introduction
When cardiac T1 mapping is performed with the MOLLI
technique (1), different acquisition timing schemes are applied separately for
the non-contrast native (or long T1) and the post-contrast enhanced (or short
T1) conditions. It is mainly because of the two reasons. First, the significantly
shortened post-contrast T1 values are thought to be better measured by a more rapid
acquisition timing scheme in the early dynamic recovery phase of the inverted
magnetization. Second, when compared to the tissue with relatively long T1 in
the non-contrast native condition, the short T1 tissue would sufficiently recover
within the shorter time before the next inversion pulse. For example, the
native (long) acquisition scheme could be the 5s of initial acquisition after
the first inversion, followed by the 3s of rest and 3s of the last acquisition
after the second inversion, which could be noted as 5s(3s)3s. With the similar
notation, the enhanced (short) scheme could be 4s(1s)3s(1s)2s. However, in a
clinical environment with multiple MRI systems, the distinction and application
of two separate acquisition timing schemes, namely the long and the short schemes,
may unnecessarily add complexity to the prompt—and consistent—optimization of the
cardiac T1 mapping protocols and propagation of them across multiple systems. Hence,
a phantom experiment was performed to compare the two different acquisition schemes
for a wide range of T1 values.Methods
The MRI phantom: A T1 phantom was built using 12
vials (each of 50 mL capacity and 4.8 cm2 cross-sectional area)
containing distilled water mixed with different concentrations of gadolinium
contrast agent (gadobenate dimeglumine). The prepared concentration ranged from
0.03 mg/mL to 0.7 mg/mL to make the T1 of each vial in the range approximately
from 200 to 2000 ms.
MRI scan: MRI scans were performed in a 1.5 T MRI
system (Ingenia, Philips, Netherlands). The T1 phantom was first scanned with
the IR-based sequence (TI= 50 to 2800 ms, TR=6 s, TE= 8 ms) to determine the T1
value of each vial, which was then taken as a reference T1 at the time of the
experiment. Then, the MOLLI technique was applied with the scanner-provided simulation
of the ECG trigger. The simulated ECG signal was generated at a fixed heart rate
(HR) of 100 bpm and also at different HRs ranging from 40 to 160 bpm in the
steps of 20 bpm. For each HR, both long and short acquisition schemes were
applied, and they were 5s(3s)3s and 4s(1s)3s(1s)2s, respectively. All the analyses
for determining T1 values were performed using the MATLAB-based in-house
program, by which the signal change from each vial was sampled and fitted to
the 3-parameter exponential T1 modeling equation (2). Results
The T1 values of the phantom, ranging from 216 to 2017 ms in
the 12 vials, are shown as a map in Fig. 1. The signal changes detected for all
the vials by the MOLLI technique with the simulated HR of 100 bpm are shown in
Fig. 2, where the curves for the long and the short acquisition schemes are
separated into (a) and (b). Shown in Fig. 3 are the MOLLI-determined T1 values compared
to the reference T1 values, for the HR of 100 bpm, with the long and the short acquisition
schemes. As expected, the short acquisition scheme failed to accurately detect
the long T1 values. In this case, the deviation started from the T1 value of approximately
1000 ms, and starting from the T1 value of approximately 1500 ms, the error becomes
10% or more. However, the long acquisition scheme did not show any clear
deviation from the reference values even for the short T1 values less than 400
ms, and the errors over the whole tested T1 range was consistently less than
2%. The comparisons made with a separate set of measurements at different HRs
are also shown in Fig. 4. For all different HRs, the short (enhanced) acquisition
scheme, as expected, deviated from the reference in the higher T1 values, and
even in the short T1 value range (< 400 ms), it did not show a clear
advantage compared to the long (native) acquisition scheme. Discussion
The short (or enhanced) scheme is designed to provide the
better accuracy in measuring the shortened T1 values in the heart after
contrast injection. However, the results here suggest that the accuracy of the short
scheme is in fact not better than the long (native) scheme. In addition, the strategy
of having two different acquisition schemes, to the author’s experience as a
clinical MRI physicist, have added complexity to the consistent management of multiple
clinical protocols and across multiple MRI systems. In this regard, the possible
single acquisition scheme strategy may help reduce the complexity. Another suggestion
from the results here is that the acquisition scheme does not need to be
tailored by the post-contrast time and the single fixed acquisition scheme
could be used even with a very late post-contrast time.Conclusion
In the
cardiac T1 mapping utilizing the MOLLI technique, the single acquisition timing
scheme, namely the long (native) scheme, can be used for both prior- or
post-contrast conditions without sacrificing the measurement accuracy. Acknowledgements
No acknowledgement found.References
1. Messroghli DR, et al., Modified
Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the
heart, Magn Reson Med, 52:141-146 (2004)
2. Taylor AJ, et
al., T1 mapping: basic techniques and clinical applications, JACC:
Cardiovascular Imaging, 9(1):67-81 (2016)