Belinda Ding1,2, Sydney Nicole Williams1, Minghao Zhang3, Jürgen Herrler4, Patrick Liebig4, Iulius Dragonu5, Radhouene Neji5, Christopher T. Rodgers3, and David A Porter1
1Imaging Centre of Excellence, University of Glasgow, Glasgow, Scotland, 2Siemens Healthcare Ltd., Glasgow, United Kingdom, 3Wolfson Brain Imaging Centre, University of Cambridge, Cambridge, United Kingdom, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Siemens Healthcare Ltd., London, United Kingdom
Synopsis
Keywords: Parallel Transmit & Multiband, Parallel Transmit & Multiband, phase schedule
Multiband(MB) imaging can
greatly reduce scan time, but MB pulses often encounter peak amplitude
constraints. An optimal phase schedule has been published for traditional
single-transmit (sTx) RF pulses. However, after an initial evaluation, we found
that it only has marginal benefits for parallel-transmit (pTx) pulses. We
compared four common optimisation algorithms (fmincon, fminsearch, simmulated
annealing, global search) to determine the most suitable algorithm for choosing
the best offset phases for a pTx pulse. Overall, fmincon efficiently selected
the optimal phase schedule of a pTx pulse and lowered peak amplitude by an
average of 15% when MB-factor=3.
Introduction
Simultaneous multislice(SMS) or multiband(MB)
imaging allows several slices to be imaged simultaneously and dramatically
reduces scan time. The simplest way to obtain MB excitation is to sum multiple
RF pulses with different frequency modulations:
$$RF_{MB}(t)=RF_{SB}(t)\sum_{k=1}^{N} e^{i(γGtx_k+\phi_k)}$$
where $$$RF_{MB}(t)$$$=MB-RF
waveform, $$$RF_{SB}(t)$$$=single-band
waveform, N=MB-factor, G=slice-select gradient amplide, and $$$x_k$$$ and are the spatial location and phase offset for
kth slice.
If $$$\phi_k$$$ is identical for all slices and slices are in
phase at t=0, the MB pulse’s peak amplitude scales with N. As such, MB-pulses
are prone to exceed the scanner peak amplitude limit. Peak amplitudes can be
reduced by allowing the phase of each slice to vary, and an optimal phase
schedule published by Wong is widely used for this[1].
Parallel RF transmission (pTx) plays a vital
role in reducing B1+ inhomogeneity associated with
ultra-high-field (UHF, B0 ≥ 7 T) imaging. In pTx, the complex-valued relationship
between transmit channels varies spatially with static pTx or B1+ shimming, offering additional degrees of
freedom to homogenise B1+[2].
With these additional degrees of freedom, the
solution space of optimal phases to reduce peak RF amplitude increases beyond
those provided by Wong[1]. This is because the phase difference between pTx
channels often varies along the slice dimension producing summation patterns
for SMS that might be ill-suited compared to the case of single transmission
(sTx) with uniform summation across channels (Figure 1).
This abstract aims to evaluate the
effectiveness of using the optimal single-transmit (sTx) phase schedule in pTx
pulses and compares different algorithms in choosing the best offset phases for
a pTx pulse. Initially, slice-selective B1+ shimming is
investigated.Methods
Both sTx and pTx MB-RF were created by summing
single-band sinc pulses using the CP and slice-specific B1-shimmed complex
coefficients (pTx), with no phase offset and the sTx optimal phase schedule[1]:
$$RF_{MB,j}(t)=RF_{SB}(t)\sum_{k=1}^{N}w_{i,j}e^{i(γGtx_k+\phi_k)}, j =1,2,...8$$
where the RF shim vector w has a unit norm and $$$w_{i, j}=\frac{1}{\sqrt{8}}e^{i(2\pi/8)(j-1)}$$$ in CP mode for an 8Tx system.
Using a factorial study design, MB-pulses were
created with these parameters: 2 shim coefficients types (CP/pTx) x 5 pulse
durations (2,4,6,8,10ms) x 5 time-bandwidth products (2,3,4,5,6) x with/without
Hanning filter x 5 different MB factors (2,3,4,5,6) x no phase offset/sTx
optimal phase schedule[1]. For pTx pulses, slice-specific shim coefficients
were calculated from per-channel B1+ maps obtained from 6 healthy volunteers
(saturation-recovery turbo-FLASH, 34 axial slices, 3mm slice (30% gap)).
Optimal multiband offset phases were
calculated for one Hanning-filtered sinc pulse (duration=4ms, TBW=4) using four
optimisation methods (fmincon, fminsearch, simulated annealing, global search)
for MB-factors=3,4,5,6 in all volunteers. All optimisations were performed in
MATLAB (R2022a). Optimal sTx phases were used as starting points, and the
solutions were bounded between 0 and 2 when applicable. Paired Student’s t-tests were
conducted between optimisation methods, and the Holm-Bonferroni method was used
to correct for multiple comparisons [3]. An exhaustive grid search was also
performed for MB-factor=3. Each optimisation method was repeated 50 times with
random starting values in one MB pulse to assess each method’s stability.
The feasibility of a universal solution was
also investigated by comparing the similarity between slice-specific shim
vectors and optimal phases between volunteers.Results and Discussion
Peak B1+ across all
channels evaluating the optimal sTx phases in pTx MB pulses are presented in Figure
2. Overall, using optimal sTx phases results in
lower peak B1+ than using no phase offsets. However, the
reductions are slight compared to sTx pulses.
Figure 3 compares the solutions from each
optimisation method against the minimum peak amplitude obtained through the
grid search method in an MB3 pulse. Figures 4A-C present results from higher
MB-factors.
All four optimisation methods provided drastic
reductions in peak B1+ (15%-40%), and although the differences
between the methods are statistically significant, the actual differences are very
small and unlikely to have any practical impact to SAR management and RF pulse
design.
Global optimisation methods (simulated
annealing, global search) outperformed non-linear optimisation methods
(fmincon, fminsearch) but their long optimisation times make them unfeasible
for inline pulse design (Figure 4B). Overall, fmincon was better and more
reliable than fminsearch, which is likely because fmincon uses a gradient-based
approach. However, the optimisation time of fmincon increases drastically with
the MB factor. Thus, simulated annealing may be more reliable and equally
time-efficient for high MB factors.
For slice-by-slice static pTx, the database of
6 subjects showed low degrees of similarity amongst cross-subject shims,
discouraging the pursuit of universal slice-specific shims (Figure 5a). The
optimised phases also varied greatly between subjects and shims (Figure 5b).
Future work will require decreasing computation time
to implement this subject-specific phase optimisation approach for pTx SMS in
practice (e.g. by including gradient information in the optimiser). This is
particularly true for higher MB factors or with an extension of this study to
dynamic pTx (e.g., slice-selective spokes). Additionally, this work could be
explored at the point of MB pulse design, where introducing a looser peak RF
amplitude constraint before phase optimisation could improve pTx
performance. Conclusion
Overall, we have shown that using sTx optimal
phase schedule has only marginal benefits when applied to a pTx MB pulse and
shim-specific phase optimisation should be performed to reduce peak RF
amplitude.Acknowledgements
MZ
is supported by the Medical Research Council (grant number MR N013433-1) and
the Cambridge Trust.References
[1] Wong, E. Optimized Phase Schedules for
Minimising Peak RF Power in Simultaneous Multi-Slice RF Excitation Pulses. in
Proc. 20th Annu. Meet. ISMRM vol. 20 (2012).
[2] Padormo, F., Beqiri, A., Hajnal, J. V. & Malik, S. J.
Parallel transmission for ultrahigh-field imaging. NMR Biomed. 29, 1145–1161
(2016).
[3] Holm, S. A Simple Sequentially
Rejective Multiple Test Procedure. Scandinavian Journal of Statistics 6, 65–70
(1979).