Celik Boga1, Ole Geldschlager2, and Anke Henning1
1UT Southwestern Medical Center, Dallas, TX, United States, 2Max Planck Institute for Biological Cybernetics, Tuebingen, Germany
Synopsis
Keywords: RF Pulse Design & Fields, High-Field MRI
The Universal
Pulse (UP) concept to parallel transmission radiofrequency (RF) pulse design
was introduced to avoid time consuming B
1+ and
ΔB
0 data acquisition before each clinical scan. Introduced large flip angle design methods utilize kT-point trajectories and small tip angle approximation. In this work, new quasi damping scheme for Newton's methods method is introduced for large flip angle universal pulse design using numerical solution of the Bloch equation. Introduced quasi-damping scheme can be utilized in
large flip angle Universal Pulse design for all flip angles including 180
o pulses, without
the loss of performance.
Introduction
The Universal
Pulse (UP) concept to parallel transmission radiofrequency (RF) pulse design
was introduced by Gras1 in order to avoid time consuming
and
data
acquisition before each clinical scan. Previously, large flip angle (
) universal pulses have been introduced, however
these methods combine kT-point trajectories with the small tip angle (STA) framework2-4. In contrast, large flip angle pulses are typically designed using the
numerical solution of the Bloch equation via Euler-Rodrigues rotation formula5,6.
In this work, solving the Bloch equations directly is combined with the Newton’s
method with a novel quasi-damping scheme for the design of high performance large
flip angle UPs. Methods
Numerical solution of the Bloch equation via Euler-Rodrigues rotation formula6 is
used herein for designing large flip angle universal parallel transmission pulses
for a given gradient trajectory instead of the small tip angle approximation7.
All relaxation effects are ignored for the pulses designed in this work. The
calculated magnetization for all 18 volunteers is denoted as $$$M$$$
and the desired
magnetization as $$$M_{des}$$$. Initially, the magnetization $$$M$$$ is assumed to be in z- direction. The desired magnetization for excitation pulses is $$$M_{des}=\left[\begin{matrix}0&sin(\alpha_{des})&cos(\alpha_{des}) \end{matrix}\right]^T$$$ for a rotation around the x-axis with angle $$$\alpha_{des}$$$. For the design of UPs the cost function $$$||M-M_{des}||_2^2$$$ was minimized using the Newton’s method. Defining pulses in vector form as $$$p$$$
and pulse value
updates at each iteration as $$$\Delta p$$$, the pulse update can be calculated as $$$\Delta p_k=J^{-1}(M_{des}-M)$$$ for iteration $$$k$$$. $$$J$$$
indicates
the Jacobian matrix, which is calculated by the finite difference method, and $$$M_k$$$ denotes
the solution of Bloch equation using $$$p_k$$$
. Then, the pulse for the next iteration is calculated
as $$$p_{k+1}=p_k+\Delta p_k$$$. Due to ill-conditioning of $$$J$$$
, Moore-Penrose pseudo
inverse with
tolerance $$$10^{-3}$$$ is used for calculating
$$$\Delta p$$$ at each iteration.
Instead of classical damping where pulse updates are
scaled with the step size $$$\gamma<1$$$, such that $$$p_{k+1}=p_k+\gamma\Delta p_k$$$
, damping herein is realized by increasing $$$\alpha_{des}$$$ step wise as
iterations progress. This method will be denoted as quasi-damping with $$$N$$$
steps, where $$$N$$$ denotes the
total number of increment steps for $$$\alpha_{des}$$$.
$$$N=1$$$ indicates that no damping is used. In order to
design $$$180^o$$$ pulses at least
2 steps for $$$\alpha_{des}$$$ are required since
$$$J$$$ has all zero rows for magnetization in z
direction and $$$M_{des}$$$
has
all zero elements for magnetization in x and y direction. This also prevents the
use of classic damped Newton’s method for
pulse design
regardless of the $$$\gamma$$$
.
An existing dataset with single channel $$$B_1^+$$$ and $$$\Delta B_0$$$ maps for 18 volunteers presented in Geldschlaegber et al8 is utilized for universal large flip angle pulse design while testing the introduced quasi-damping scheme. In this work, 2D excitation UPs are designed with a fixed gradient trajectory shown in Figure 1. UPs designed using 1-5 steps with 3 iterations at each step for $$$90^o$$$ pulse and 2-6 steps with 3 iterations at each step for $$$180^o$$$ pulse are compared against UPs designed with the classically damped Newton method and STA UPs. Normalized root mean squared error (NRMSE) calculated from the desired versus the magnetizations generated by UPs is used as the measure of pulse performance.Results
Figures 2 and 3 show that $$$90^o$$$ and $$$180^o$$$ UPs designed via the proposed Newton method with quasi-damping result in highly uniform excitation profiles for all 18 subjects, which is not the case for STA based UPs. NRMSE plots shown in Figure 4(a,b), indicate that the proposed quasi-damping scheme for Newton’s method increases the performance of UPs as the number of steps increase. UPs based on the non-damped Newton’s method $$$(N=1)$$$ have substantially higher NRMSEs than any pulse design using the quasi-damped Newton method (Figure 4(a)). However, the decrease in NRMSE gets smaller as the number of steps increase. Similarly, $$$180^o$$$ pulses with 5 and 6 steps perform clearly better than pulses designed using a lower number of steps, indicating that performance of UPs can be increased via quasi-damping independently of the desired flip angle. Figure 4(c) shows that the proposed quasi-damping method results in the same average NRMSE as the damped Newton’s method, with step size $$$\gamma = 0.5$$$ , in 15 total iterations for the $$$90^o$$$ pulses. On the other hand, Figure 4(d) shows that the damped Newton’s method cannot be used for $$$180^o$$$ pulses whereas the quasi-damped Newton’s method works well for $$$180^o$$$ UP design.Discussion and Conclusion
In conclusion, the newly introduced quasi-damping
scheme can be utilized in large flip angle Universal Pulse design for all flip
angles including $$$180^o$$$ pulses, without
the loss of performance.Acknowledgements
This work was funded by Cancer Prevention and Research Institute of Texas (CPRIT) RR180056 and was performed under the rubric of the Advanced Imaging Research Center, UT Southwestern Medical Center.References
1.Gras V, Vignaud A,
Amadon A, Bihan D, Boulant N. Universal pulses: A new concept for
calibration‐free parallel transmission. Magnetic Resonance in Medicine.
2016;77(2):635-643. doi:10.1002/mrm.26148
2. Gras V, Mauconduit F, Vignaud A, et al.
Design of universal parallel-transmit refocusing Kt-point pulses and
application to 3D T2-weighted imaging at 7t. Magnetic Resonance in Medicine.
2017;80(1):53-65. doi:10.1002/mrm.27001
3. Pracht E., Gras V.,
Boulant N. and Stöcker T. “Whole Brain FLAIR Imaging at 7T Employing Universal
Pulses” In Proceedings of the 26th Annual Meeting of ISMRM, Paris,
France, 2018
4. Gras V, Pracht ED, Mauconduit F, Le Bihan
D, Stöcker T, Boulant N. Robust nonadiabatic T2 preparation using universal
parallel-transmit Kt-point pulses for 3D Flair imaging at 7 T. Magnetic
Resonance in Medicine. 2019;81(5):3202-3208. doi:10.1002/mrm.27645
5. Rund A, Aigner CS, Kunisch K, Stollberger
R. Magnetic resonance RF pulse design by optimal control with physical
constraints. IEEE Transactions on Medical Imaging. 2018;37(2):461-472.
doi:10.1109/tmi.2017.2758391
6. Luo T, Noll DC, Fessler JA, Nielsen J-F.
Joint design of RF and gradient waveforms via auto-differentiation for 3D
tailored excitation in MRI. IEEE Transactions on Medical Imaging.
2021;40(12):3305-3314. doi:10.1109/tmi.2021.3083104
7. Saekho S, Boada FE, Noll DC, Stenger VA.
Small tip angle three-dimensional tailored radiofrequency slab-select pulse for
reducedb1 inhomogeneity at 3 T. Magnetic Resonance in Medicine.
2005;53(2):479-484. doi:10.1002/mrm.20358
8. Geldschläger
O, Bosch D, Henning A. OTUP workflow: Target specific optimization of the
transmit k ‐space trajectory for flexible universal parallel transmit RF
Pulse Design. NMR in Biomedicine. 2022;35(8). doi:10.1002/nbm.4728