Maxim Terekhov1, David Lohr1, Christoph Aigner2, Sebastian Dietrich2, Sebastian Schmitter2, and Laura M. Schreiber1
1Chair of Molecular and Cellular Imaging, Comprehensive Heart Failure Center, University Hospital Würzburg, Wuerzburg, Germany, 2Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Berlin, Germany
Synopsis
Keywords: Parallel Transmit & Multiband, Cardiovascular
Parallel
transmit (pTX) technology is an emerging tool for improving of the B
1+-field
homogeneity in cardiac MRI (cMRI) at the ultra-high magnetic field. In this
work, we propose a methodology to enhance the characteristics of the shaped B
1+
field using a tailored cost function for the optimization procedure computing
complex transmit vectors of magnitudes and phases for driving TX array. Using the cost functions (CF) based
on the local B
1 gradients with fine-tuning by weighting coefficients
allows for considerable improvement of static pTX B
1-shimming
quality in compared to traditional CF using the coefficient-of-variation (CoV)
of B
1.
Introduction
During
the last decades parallel transmit (pTX) technology became an emerging tool for
improving the B1+-field homogeneity in cardiac MRI (cMRI)
at the ultra-high magnetic field. Manipulating the driving voltages of transmitting
(Tx) array elements allows for shaping the spatial distribution of B1+
with desirable spatial properties within a region of interest (ROI). In this work, we propose a methodology to
improve the characteristics of the shaped B1+ field using tailored cost functions (CF) for the optimization procedure by computing
complex transmit voltage vectors for the TX array. Preliminary results[1] in phantoms
using in-house developed arrays demonstrate that using a CF based on the B1
gradients allows taking into account the properties of the local surface
Tx-arrays that by its geometry generates B1 with intrinsic heterogeneity in
specific directions (most often anterior-posterior). As a result a considerable
improvement of static pTX B1-shimming quality can be achieved in
comparison to traditional CF using the coefficient-of-variation (CoV) of B1. In this work, the dataset of B1+
obtained by a commercial cardiac array (MRI Tools, Berlin, Germany) acquired from 36
subjects[2] has been
used to explore further this hypothesis. Methods
The
combined magnetic field of the array is given $$$ B_1(r) = \sum_{k=1}^{N} C_k \cdot b_{1k}(r) $$$ where b1k(r) are
complex absolute or relative B1-maps measured for each Tx-channel.
The B1-shimming of an array is performed by control of the Tx-vector {C1..CN} to achieve the targeted spatial homogeneity of
combined field B1(r). Finding
out a vector is formulated as an optimization problem for the specific cost
function (CF) including characteristics of the targeted B1 in the
region-of-interest. The efficient usage of the RF-power with the specific
Tx-vector is controlled via the “transmit efficiency factor”. The summary of the
optimization problem and cost functions is shown in Fig 1a. In the first step,
we explore the efficiency of B1 optimization using CF based on the coefficient-of-variation
(CoV) of B1 (“statistical-based” CF, further SCF) with
extension factors aimed improving homogeneity and managing destructive
interference. In the second step, we probed the CFs based on the spatial
gradients of targeted B1 with adjustable weighting coefficients (“gradient-driven”
CF, further GCF). In
order to ensure a fair comparison of results using different CF with regard to
the used RF-power the equal boundary conditions for the real and imaginary part of the vector were used in the
optimization solver (equivalent to the limitation of
RF-power-per-channel). Additionally, all resulted vectors were normalized as before evaluating the statistical metrics of
combined B1.
As
a data source of B1 maps, the dataset acquired in the context of [3] and [4] was used. This
includes B1-maps acquired using RF-array with 8 dipole Tx-elements and 24 loops
composing 8Tx/32Rx architecture for pTX application for cMRI at 7T. The dataset
includes B1 maps of 8 individual channels and 3D masks of optimization
ROIs. The computation was done using an in-house
developed Matlab toolbox [5]. Results
Figure
2 demonstrates the results of using SCF (CF1stat.. CF4stat) to optimize default B1 maps in
the example subject (#4 in the dataset). For the reference CF1stat
(including only CoV), the improvement of
homogeneity (CoV decreased by factors 2 to 3) is observed (Figure 2b). Using the basic function CF1stat removes destructive
interferences only partially. Using tailored SCF with extension factors allows
for significantly better managing the destructive interferences which are
manifested both visually (Figure 2a, arrows marks) and by an increase of the
minimal B1 value up to factor
5 (Figure 2c).
Figure
3 demonstrates the results of B1 optimization using SCF and
GCF respectively in the example subject. Using fine-tuning of the
weighting coefficient for GCF allows for achieving further improvement in
managing destructive interference compared to both the reference and extended
SCF.
Figure
4 shows default and optimized B1 maps in 12 example subjects. The slice
with the maximal area of the mask is demonstrated. One can observe that GCF-optimization
provides additional smoothness of the B1-maps
(examples are labeled) in comparison to SCF-optimization. This could be
seen in the Figure 5a demonstrating examples of vertical profiles in the same
slice. In addition to that, using GCF provides an additional gain of both
minimal and mean value of B1 compared to SCF-optimization (Figure 5b). Finally,
Figure 5c shows that using GCF is up to 40% less computationally expensive compared to CF
based on CoV ( even without extension factors)Discussion
Our
results show that CFs extending the standard CoV-based function provide options
for efficient suppression of destructive interferences of B1 using static
pTX-shimming. This introduces, however, additional computational costs in the
optimization. Using CF based on weighted spatial metrics allows taking into
account the geometry of surface Tx-arrays for UHF cMRI which introduces
intrinsic gradients in a shaped B1. Using fine-tuning of weighting coefficients
allows us to better address both destructive interferences and intrinsic
spatial inhomogeneities of B1 in the heart at UHF. Conclusion
It
was demonstrated that using specifically tailored cost functions allows for
better addressing the destructive interferences of B1 in the CMRI at 7T using
static pTX shimming. The same approach
can be transferred further for dynamic pTX optimization tasks using both tailored
and universal RF pulses. Acknowledgements
L.M Schreiber receives research support by Siemens
Healthineers. The position of D. Lohr is partially funded by this research
support. References
1. Terekhov, M., Elabyad, I,,
Schreiber, L. Optimal Cost Functions for
the Static pTX B1+-Shimming in Ultra-High Magnetic Field cardiac MRI in ISMRM Workshop "Ultra-High Field
MR". 2022. Lisbon, Portugal.
2. Aigner,
C.S., et al., Calibration-free pTx of the
human heart at 7T via 3D universal pulses. Magn Reson Med, 2022. 87(1): p. 70-84.
3. Aigner,
C.S., S. Dietrich, and S. Schmitter, Respiration
induced B 1 + changes and their impact on universal and tailored 3D kT-point
parallel transmission pulses for 7T cardiac imaging. Magn Reson Med, 2022. 87(6): p. 2862-2871.
4. Dietrich,
S., et al., 3D Free-breathing
multichannel absolute B 1 + Mapping in the human body at 7T. Magn Reson
Med, 2020.
5. Terekhov, M., I.A. Elabyad, and L.M.
Schreiber, Global optimization of default
phases for parallel transmit coils for ultra-high-field cardiac MRI. PLoS
One, 2021. 16(8): p. e0255341.