Giovanni Costa1, Margarethus Maarten Paulides1, and Irena Zivkovic1
1Department of Electrical Engineering, Technical University of Eindhoven, Eindhoven, Netherlands
Synopsis
Keywords: RF Arrays & Systems, High-Field MRI, Coupling, SCC coils, Faraday's law of induction, SAR reduction
Minimizing
coupling between coils constitutes a major challenge in the design of flexible
arrays for UHF MRI. Recently, Shielded-coaxial-cable coils (SCC) were shown to
be flexible and intrinsically highly decoupled: however, the decoupling
mechanism of SCC remains unclear. In this work, we use an high frequency Faraday-equivalent
model of the induced voltage to examine coupling between two SCCs in different
geometrical configurations. We show that lower coupling in SCC derives from
both lower magnitude of electric and magnetic fields as compared to a loop coil, and we discuss the implications
in the design of coils for MRI.
Introduction
Especially
in the design of flexible arrays, a major challenge is minimizing coupling
between coils. Shielded-Coaxial-Cables (SCC) provide flexibility, intrinsic
decoupling characteristics, and straight-forward fabrication1,2, but
their decoupling mechanism is still unclear. Understanding coupling of SCCs
could enable the design of novel and more efficient coil elements. In this
work, we derived an high frequency Faraday-equivalent model of the induced
voltage and examined coupling between two SCCs in different configurations.Methods
Using
the reciprocity theorem3 , the reaction concept4 , and
the full set of Maxwell equations, we derived an expression for the coupled
voltage of two coils:
$$ (1) \; \; \; V_{21} = \oint_S ( \textbf{E}_{2}\times \textbf{H}_{1} ) \cdot d\textbf{S} + j\omega \left( \int_V \varepsilon \textbf{E}_2 \cdot \textbf{E}_1 dV + \int_V \mu \textbf{H}_2 \cdot \textbf{H}_1 dV \right) $$
where $$$V_{21}$$$ is the open-circuit voltage induced at the port of coil 2 when coil 1 is
active–i.e. the coupled voltage - $$$\textbf{E}_{1}$$$ ($$$\textbf{H}_{1}$$$)
is the electric (magnetic) field radiated by coil 1, when coil 1 is excited with 1A of current and
coil 2 is open-circuited–and vice versa for $$$\textbf{E}_{2}$$$ and $$$\textbf{H}_{2}$$$-$$$V$$$ is a volume containing at least source 2,
and the surface integral is extended over the contour of $$$V$$$, $$$S$$$.
Expression
(1) generalizes Faraday’s law of induction, as it represents the voltage
induced at the ports of a coil at any desired frequency and for any
configuration. Faraday’s law of induction has been extensively used in MRI
literature to model coupling in the low-frequency approximation5,6.
In
expression (1), it is possible to recognize similar terms of electric coupling
and magnetic coupling as defined in Hong7,8 , or Awai9-11
. The additional term accounts for electric and magnetic coupling contributions
outside $$$V$$$. Note that the theory of Hong was used to
explain the rationale in the design of Self-Decoupled-Coils12, and
that electric coupling in Hong is defined the same as in Roemer5 .
We apply
expression (1) to the case of two-coil arrays, one of of SCCs and the other of loops,
in different configurations, at
300MHz. As input, we used the EM fields simulated by full-wave
simulation software (CST MWS, Darmstradt, Germany), and numerical Monte-Carlo
integration.
Coils were
simulated at 12mm distance from a square phantom εr=78, σ=0.6[S/m] (figure
1) using the frequency domain solver with adaptive tetrahedral mesh. SCCs
simulations were validated earlier by experiments13 , and we use the same settings for
the loop coil simulations. Coils were excited at the ports by 1A.
To perform integrals, we
used a spherical volume radius=400mm encompassing both sources and phantom. We
use point lists to export the $$$\textbf{E}_{2}\times \textbf{H}_{1}$$$ , $$$\textbf{E}_2 \cdot \textbf{E}_1$$$, $$$\textbf{H}_2 \cdot \textbf{H}_1$$$ fields at defined locations from CST into MATLAB,
in which they were integrated, and compared to the voltage $$$V_{21}$$$ by CST.Results
Figure 2
shows that in each configuration the induced voltage was negligible for the
SCCs as compared to the loops. The average relative difference between
expression (1) and $$$V_{21}$$$ by CST was -1.8% for the real part and 2.7% for the
imaginary part.
Figure 3
shows that the value of the term $$$\oint_S ( \textbf{E}_{2}\times \textbf{H}_{1} ) \cdot d\textbf{S}$$$ was negligible as compared to the electric and
magnetic coupling terms. The SCCs showed much less electric and magnetic
coupling contributions compared to the loops.
Figure 4 shows that the magnitude
of the E and H fields were significantly lower in the SCC case, with comparable
value of the magnitude of B1+/√W
Figure 5 shows the magnitude of the E, H, B1+/√W field in a central
cross section, in the case of a single capacitive-segmented loop, and in the
SCC case. Plots were generated using 1A of current at the ports. The magnitude of the E and H fields was still
lower in the SCC case, with comparable magnitude of B1+/√W.Discussion
Both
electric and magnetic coupling were lower in SCCs. From expression (1) we
deduce that lower electric (magnetic) coupling can be either related to
orthogonality between two radiation profiles, or lower average amplitude of the
fields E1(H1) and/or E2(H2).
To discriminate between these two effects, we
examined the plot of the E and H fields in a central cross-section of the
phantom. We note that in every configuration the E and H fields -which
determine SAR in the tissue- were lower
in the SCC case, with comparable magnitude of the B1+/√W field -which
determine SNR in the image.
It is well known that capacitors block the radiation of the E field.
Therefore, we compared the E, H and B1+/√W fields
of a SCC with the E, H and B1+/√W field of a capacitive-segmented loop. To make
the two plots comparable, we considered the fields radiated for 1A of current
at the ports. We note that the E and H field were lower in the SCC case.Conclusions
In this work, we discuss coupling in SCCs
and further implications in the design of MRI coils. We demonstrate that lower
coupling in SCCs derives from lower magnitude of the E and H fields in the
calculation volume, with comparable magnitude of the B1+/√W field. Further,
we show that a shield is more effective than using capacitors to block
the propagation of the E and H fields.Acknowledgements
No acknowledgement found.References
1. T.
Ruytenberg, A. Webb, and I. Zivkovic, “A flexible five-channel
shielded-coaxial-cable (SCC) transceive neck coil for high-resolution carotid
imaging at 7T”, Magn Reson Med. 2020; 84:1672–1677
2. T.
Ruytenberg, A. Webb, and I. Zivkovic, “Shielded-coaxial-cable Coils as Receive
and Transceive Array Elements for 7T Human MRI.” Magn Reson Med (2019)
83(3):1135–46.
3. R. F.
Harrington, “Time-Harmonic Electromagnetic Fields.” Piscataway, NJ, USA: IEEE
Press, 2001
4. V. H.
Rumsey, “The Reaction Concept in Electromagnetic Theory”, Phys. Reo., sere 2,
vol. 94, no. 6, pp. 1483-1491, 1954
5. P.B.
Roemer et al. , “The NMR phased array”, in Magnetic Resonance in Medicine, vol.
16, pp.192-225, 1990
6. Constantinides, C D et al. “A phased array coil for human
cardiac imaging.”, in Magnetic resonance in medicine vol. 34, n. 1, pp. 92-98,
1995
7. J.-S.Hong , “Couplings of asynchronously tuned coupled microwave resonators”,
IEE Proc. Microw, Antenna and Propag., Vol. 147. No. 5, Ucloher 2000
8. J.-S.
Hong, M. J. Lancaster, “Microstrip Filters for RF/Microwave Applications”, John
Wiley & Sons, Inc. , 2001
9. I. Awai
and Y. Zhang, “Maxwell equations coupled mode theory and coupling coefficient”
, Proc. of Asia Pacific Microwave conference 2014, 1223-1225
10. I.
Awai and Y. Zhang, “Separation of Coupling Coefficient Between Resonators into
Magnetic and Electric Components Toward Its Application to BPF Development” ,
2008 China-Japan Joint Microwave Conference, 61-65
11. I.
Awai and Y. Zhang, “Coupling Coefficient of Resonators—An Intuitive Way of Its
Understanding”, Electronics and Communications in Japan, Part 2, Vol. 90, No.
9, 2007
12. Xinqiang Yan, John C. Gore, William A. Grissom, “Self-decoupled radiofrequency
coils for magnetic resonance imaging”, Nature Communications, 9:3481, pp. 1-12,
2018
13. G. Costa, , S. Güler ,
P. Baltus, M. Paulides, I. Zivkovic, “Investigation of correlation between
surface current and coupling using conventional coils and shielded coaxial
cable coils operating at 7T”, Proc. Int. Soc. For Mag. Reason. Med., 2022