Yunkun Zhao1 and Xiaoliang Zhang1
1Biomedical Engineering, State University of New York at Buffalo, Buffalo, NY, United States
Synopsis
Keywords: High-Field MRI, RF Arrays & Systems
This study
investigates the effect of lumped inductors used in RF resonant circuits on the
magnetic field strength generated by the radiofrequency (RF) coils. The results
show that the lumped inductors can reduce B1 fields in samples.
Introduction
Radiofrequency coil
array plays an essential role in MR signal excitation and reception.
Multichannel MRI systems require an RF coil array to increase image quality in
terms of SNR and image speed. One key obstacle in RF coil array design is the
electromagnetically coupling between coil elements of the RF array. In order to
address this issue, lumped inductors are often used in the decoupling circuits
or in coil’s resonant circuit. In some occasions, the lumped inductors are also
used to tune the coil, either array or non-array, to the desired frequency [1].
Inductor is a device which stores magnetic fields. Due to the confined area of
a lumped inductor, the magnetic field stored in the lumped inductor may not be
“seen” by the imaging sample or subject, potentially resulting in reduced B1
fields in the sample and thus reducing the detection sensitivity. In this
study, we investigate how magnetic field strength will change with the
inductance of the lumped inductors used in the RF coil circuit.
Methods
The coil used in
this study is shown in Fig.1. We designed a 10×10 cm2 LC loop coil with 2
capacitors and a lumped inductor. A phantom (imaging sample) positioned 10mm
above the coil has a size of 50mm in height and75mm in width. Length of the phantom is defined as an
imaging area, which reflects the position of the imaging object in real
applications. The LC loop coil was built with 6.35mm width copper tape, and the
lumped inductor in the solenoid shape was built with 16 AWG copper wire. The
length of the lumped inductor is 10mm, and the outer radius of the inductor is
1.25mm. These dimensions remained the same throughout the experiment. The
inductance value changes by increasing or decreasing the turns of the solenoid inductor.
There are 7 cases been tested: 4 turns (9nH inductance), 5 turns (14nH
inductance), 8 turns (35nH inductance), 10 turns (55nH inductance), 16 turns (142nH
inductance), 20 turns (221nH inductance). No lumped inductor in the circuit is
defined as 0nH. The coils in every case were tuned to 298MHz (the Larmor
Frequency of proton 1H at 7T) using tuning capacitors. Numerical results of the
proposed designs are obtained using electromagnetic simulation software CST
Studio Suite (Dassault Systèmes, Paris, France). The performance of the coils
was evaluated by a 2D B1 field distribution plot and a 1D plot of magnetic
field strength inside the imaging area or sample. All the results were
normalized to the input power at the resonant frequency with 0.5Watt total
power from the power source. Results
Fig.2 shows the B1
field distribution at all different inductance values. Calculated field
distribution results show that B1 field around LC loop coil in the transverse
plane decreases with the increase in inductance of the lumped inductor. Fig.3 demonstrates
the 1D profile of the B1 fields in two different orientations. Both 1D plots in
Fig. 3 further indicate the B1 field reduction with increase in lumped
inductance. Fig.4 shows the B1 field distribution at different inductance
values in the imaging sample. The result indicates the same magnetic field
strength changing pattern in Fig.2, in which the field strength decreases and
shifts to the right side. Fig.5 shows the average magnetic field strength plot in
the imaging area, and the step size of the data point is 0.25mm. The result
reveals that when using a lumped inductor with 221nH inductance, the average
field strength decreased by 62% compared with a coil with no lumped inductor. Conclusion
In this study, we
investigate the effect of the lumped inductor on the magnetic field strength of
an RF coil. Results show that the use of lumped inductor in resonant circuit
would negatively impact the B1 field in the sample, ultimately degrading the MR
imaging sensitivity.Acknowledgements
This work is supported in part by the NIH under a BRP grant
U01 EB023829 and by State University of New York (SUNY) under SUNY Empire
Innovation Professorship Award.References
1. Yan X, Gore J C, Grissom W A. Self-decoupled
radiofrequency coils for magnetic resonance imaging[J]. Nature communications,
2018, 9(1): 1-12.