Yang Wu1, Peng Sun2, Zhigang Wu2, Xiaoxiao Zhang2, Jing Zhang2, and Jiazheng Wang2
1Department of Medical Imaging, Wuhan Asia General Hosipital, Wuhan, China, 2Philips Healthcare, Beijing, China
Synopsis
Keywords: Myocardium, Cardiovascular
T2 mapping is a quantitative
technique for measuring T2 in tissues, which is gaining popularity in the MR
community, particularly for the detection of cardiac edema. There are several
representative T2 mapping methods, including GraSE and T2 preparation-based
schemes. However, there is still a lack of clinical experience with these T2
mapping methods. In this study, we tentatively compared the GraSE-based and T2
preparation-based methods for myocardium imaging.
Introduction
Cardiovascular
magnetic resonance (MR) imaging (cardiac MR) is emerging as an effective
noninvasive method for detecting myocardial edema1. Myocardial edema is
typically characterized by elevated free water content and prolonged T2 values
in the myocardium. The T2-weighted STIR method is currently the most commonly
used clinical setting for myocardial edema, but it may be limited by several
factors, such as phased-array coil-related signal inhomogeneity, motion-caused
signal intensityvariation, interference from the high signal of stagnant blood,
and et al.2. T2 mapping is a quantitative technique that can directly measure
T2 in tissues and may have the potential benefits of improved diagnostic
precision and robustness.
T2 mapping schemes include multi-echo gradient and spin-echo (GraSE) based
methods3 and T2 preparation-based methods4. There is still a lack of
experience with these clinical settings. To evaluate the performance of these
different T2 mapping methods, we compared multi-echo GraSE and T2
preparation-based T2 mapping of myocardium in three healthy volunteers.Materials and Methods
The
institutional IRB approved this study. In this study, three healthy volunteers
were recruited. Each subject was scanned using a 3.0T Elition system (Philips
Healthcare, the Netherlands) with a 32-channel torso and spine coil for
multi-echo GraSE and T2 preparation-based T2 mapping. The GraSE-based T2 mapping
was carried out in a multi-echo fashion, in which a few k-space lines with
different echo times (9 echoes) were acquired in each heartbeat and the entire
k-space was filled over several heartbeats. The GraSE sequence was scanned in
18 seconds with a TSE factor of 9 and an EPI factor of 7. The
T2 preparation-based method was designed in a completely different scheme,
employing balanced turbo field echo (bTFE) as an acquisition technique with TFE
factors of 92. The T2 preparation-based method was designed in a completely
different scheme, employing balanced turbo field echo (bTFE) as an acquisition
technique with TFE factors of 92. All of the k-space lines for a single image
with a specific T2 preparation time were acquired in three heartbeats, and the
T2 map was obtained by scanning several images with four T2 preparation times
(0, 23.3, 46.6, and 70ms) in 12 heartbeats. Table 1 lists the other parameters
of the two protocols. An experienced radiologist examined the acquired T2
mapping data (5 years). Each subject's myocardium was marked with three regions
of interest (ROIs). The three ROIs were used to calculate the mean and standard
deviation of myocardium T2 values. The Wilcoxon signed-rank test was used in
Python 3.8 to compare the multi-echo GraSE and T2 preparation-based myocardium
T2 mapping methods. To investigate the size and variation of the measured T2,
the mean and standard deviation of myocardium T2 values were tested. A p-value
less than 0.05 was deemed statistically significant.Results
Table
2 shows the measured T2 values of the myocardium using the multi-echo GraSE and
T2 preparation-based T2 mapping methods. The Wilcoxon signed-rank test revealed
that the mean T2 values calculated from the three ROIs (Figure 1) were not
statistically different between the two groups. In contrast, the standard
deviation of T2 values in the myocardium differed statistically (p<0.05) between
T2 mapping methods. The measurement variation was greater for the multi-echo
GraSE-based T2 mapping method than for the T2 preparation-based method.Discussion and Conclusion
In
this preliminary study, we discovered that the T2 preparation-based T2 mapping
method produced T2 values of the myocardium that were comparable to the
multi-echo GraSE-based counterpart, implying that both methods could be used to
detect myocardial edema. Further analysis revealed that the T2
preparation-based T2 mapping scheme was more stable than the multi-echo
GraSE-based T2 mapping method, with less variation in different regions of the
myocardium, which could be attributed to the GraSE readout's sensitivity to
susceptibility. However, the preliminary study's sample size is very small, and
further research with a larger sample size is required because T2 mapping is
gaining popularity.Acknowledgements
No other acknowledgement.References
1. McAlindon EJ, Pufulete M,
Harris JM, Lawton CB, Moon JC, Manghat N, Hamilton MC, Weale PJ,
Bucciarelli-Ducci C.Measurement of myocardium at risk with cardiovascular MR:
comparison of techniques for edema imaging. Radiology,2015; 275(1): 61-70.
2. Giri S, Chung YC, Merchant
A, Mihai G, Rajagopalan S, Raman SV, Simonetti OP. T2quantification for
improved detection of myocardial edema. Journal of cardiovascularmagnetic
resonance, 2009; 11: 56.
3. Sprinkart AM, Luetkens JA,
Traber F, Doerner J, Gieseke J, Schnackenburg B, Schmitz G,Thomas D, Homsi R,
Block W, Schild H, Naehle CP. Gradient Spin Echo (GraSE) imaging forfast
myocardial T2 mapping. Journal of cardiovascular magnetic resonance, 2015; 17:
12.
4. Zhang S, Le TT, Kabus S,
Su B, Hausenloy DJ, Cook SA, Chin CWL, Tan RS. Cardiac magneticresonance T1 and
extracellular volume mapping with motion correction and co-registrationbased on
fast elastic image registration. Magma, 2018; 31(1): 115-129.