4081

Bilateral Orthogonality Generating Acquisitions Method for Homogeneous T2* Images Using Dual Channel Parallel Transmission at 7T
Celik Boga1 and Anke Henning1
1UT Southwestern Medical Center, Dallas, TX, United States

Synopsis

Keywords: High-Field MRI, RF Pulse Design & Fields

The T2*contrast specifically benefits from 7T human MRI systems and is frequently used in Neuroimaging. However, B1+ and B0 inhomogeneity at 7T limits the effectiveness of the T2*. A method originally proposed to minimize channel effects in the field of telecommunications, was adopted to a dual channel parallel transmission (pTx) human 7T MRI system. Final images are obtained by the unique combination of four small angle GRE acquisitons with adjusted flip angles for each channel. The proposed method is easy to implement, robust and provides homogeneous T2* images of the whole brain.

Introduction

The contrast specifically benefits from 7T human MRI systems and has been used as diagnostic/prognostic marker for Alzheimers Disease1, Multiple Sclerosis2,3 and Parkinsons Disease4. However, transmit field ($$$B_1^+$$$) and main static magnetic field ($$$B_0$$$) inhomogeneity ($$$Δ$$$) at 7T5,6 can severely degrade the image quality.
In this work, the Bilateral Orthogonality Generating Acquisitions method originally proposed to eliminate channel effects in the field of telecommunication7,8 was adopted to a dual channel parallel transmission (pTx) human 7T MRI system. It is demonstrated that homogeneous contrast images can be obtained without the requirement for prior mapping of and . In the first step, two spoiled gradient recalled echo (GRE) images with adjusted flip angles combined in order to generate an orthogonality between acquisitions have to be obtained, similar to the space time diversity method in wireless communications7,8. In addition, two single channel GRE acquisitions are utilized to calculate images without $$$B_1^+$$$ and $$$B_0$$$ inhomogeneity effects.

Methods

For a dual channel pTX system using the same block pulse with amplitude $$$\beta$$$ for both channels without any gradient trajectory the flip angle in a single pixel $$$i$$$ can be written as $$$ \alpha ^i = (s_1^i e^{j \theta_1} +s_2^i e^{j \theta_2} )c_{\Delta B_0}^i e^{-j\phi_{\Delta B_0}^i} \beta$$$6. $$$s_1$$$ and $$$s_2$$$ are the channel specific transmit sensitivities and $$$\theta_1$$$ and $$$\theta_2$$$ are the transmission phases of each channel. Amplitude and phase effects of $$$\Delta B_0$$$ are $$$c_{\Delta B_0}^i = \gamma m_0 \Delta t {sin(0.5 N_t \gamma \Delta B_0^i \Delta t )}/{sin(0.5 \gamma \Delta B_0^i \Delta t )}$$$ and $$$\phi_{\Delta B_0}^i = 0.5(N_t +1)\gamma\Delta B_0^i\Delta t$$$. Defining $$$q_{1,2}^i=s_{1,2}^ic_{\Delta B_0}^ie^{-j\phi_{\Delta B_0}^i}e^{\theta_{1,2}} $$$ the flip angle expression including both and inhomogeneity effects simplifies to $$$\alpha^i_{1,2}=q_{1,2}^i\beta$$$.
Using the small tip angle approximation6, spoiled GRE signal equation9 can be simplified to $$$S = \rho E_2^* \alpha $$$ , where$$$\rho$$$ is the proton density and $$$E_2^* = e^{-TE/T_2^*}$$$ . For a dual channel pTx system, spoiled GRE signal can be further written as $$$S = \rho E_2^* (q_1+q_2)\beta$$$.
For the suggested method the first two input images are defined as $$$S_1 = \rho E_{2,1}^* (q_1+q_2)\beta$$$ and $$$S_2 = \rho E_{2,1}^* (- q_1+q_2)\beta^*$$$. Since the transmit phase of the pulses is included in $$$\theta_{1,2}$$$, $$$\beta$$$ defines only the real valued amplitude of the block pulse resulting in $$$\beta = \beta^*$$$ , which then leads to $$$S_2 = \rho E_{2,1}^* (- q_1+q_2)\beta$$$ . The negative flip angle is implemented via the additional transmit phase of $$$\pi$$$ for the first channel.
The second two acquisitions utilize the same block pulses, but channels are used individually, resulting in $$$S_{1,2} = \rho E_{2,2}^*q_{1,2}\beta$$$ . By combining the four signals as in7, $$$C_1$$$ and $$$C_2$$$ can be defined as $$$C_1 = S_3^*S_1+S_4S_2^*$$$ and $$$C_2 = S_4^*S_1-S_3S_2^*$$$. In matrix format, combined signals can be written as follows.
$$ \left[ \begin{matrix}C_1\\C_2\end{matrix}\right] =\rho^2E_{2,1}^*E_{2,2}^*\left[\begin{matrix} q_1^*\beta&q_2\beta\\q_2^*\beta&-q_1\beta\end{matrix}\right]\left[ \begin{matrix}q_1& q_2\\q_2^*& - q_1^*\end{matrix}\right]\left[ \begin{matrix}\beta\\\beta\end{matrix}\right] = \rho^2E_{2,1}^*E_{2,2}^* \beta^2\left[\begin{matrix}|q_1|^2+|q_2|^2\\|q_1|^2+|q_2|^2\end{matrix}\right]$$ Orthogonality of $$$C_1$$$ and $$$C_2$$$ is only valid if there is no phase mismatch between transmit and receive channels. Since in MRI often different transmit and receive coils are used and complex transmit and receive sensitivities do generally not match. This assumption does not hold and results in phase errors. Due to a lack of orthogonality the obtained $$$T_2^*$$$ images from $$$C_1$$$ and $$$C_2$$$ show distortions. In order to eliminate effects of the phase errors, magnitudes of $$$C_1$$$ and $$$C_2$$$ have to be averaged before obtaining the final $$$T_2^*$$$ image which is then calculated as $$$I = \frac{0.5\sqrt{C_1^*C_1+C_2^*C_2}}{S_3^*S_3+S_4^*S_4}$$$ . This, also enables the use of a separate receive coil.
For the acquisition of the data, a 7T Philips Healthcare whole body human MRI system with two transmit channels is used along with the 32 channel receive Nova Medical head coil as receive coil. For two healthy volunteers, the four above described 3D spoiled GRE acquisitions were acquired each with voxel size of $$$1.5x1.5x1.5 mm$$$, $$$144x144x100$$$ acquisition matrix and $$$5^o$$$ flip angle. $$$TE_1/TR = 20/22 ms $$$ with 2 averages and $$$TE_2/TR = 1.28/4.2 ms$$$ with 5 averages are used for first two and last two acquisitions. The final $$$T_2^*$$$image has the effective echo time of $$$TE_{eff} = TE_1-TE_2$$$ and is free of effects of $$$B_1^+$$$ and $$$\Delta B_0$$$.

Results

Figure 1 illustrates the images that have to be acquired as input for the proposed method. Rows (d,h) of Figures 2-4 demonstrate that the proposed method provides whole brain $$$T_2^*$$$ images without the inhomogeneity effects of $$$B_1^+$$$ and $$$B_0$$$ and in contrast to the CP mode image in rows (a,e). Rows (b,f) and (c,g) in Figures 2-4, shows that $$$T_2^*$$$ images obtained solely from $$$C_1$$$ and $$$C_2$$$ are severely affected from the phase errors in all orientations but rows d and h in these figures shows that the final images obtained via amplitude combination of $$$C_1$$$ and $$$C_2$$$ is not affected. Figure 5 shows representative zoomed versions of the final $$$T_2^*$$$ images.

Discussion and Conclusion

This newly introduced method is easy to implement, robust and provides homogeneous $$$T_2^*$$$ images of the whole brain. Herein it is implemented for dual-channel pTx systems but an extension to 8 channel pTx systems is underway.

Acknowledgements

This work was funded by Cancer Prevention and Research Institute of Texas (CPRIT) RR180056 and was performed under the rubric of the Advanced Imaging Research Center, UT Southwestern Medical Center.

References

1. Schwarz ST, Mougin O, Xing Y, et al. Parkinson's disease related signal change in the nigrosomes 1–5 and the substantia nigra using T2* weighted 7T MRI. NeuroImage: Clinical. 2018;19:683-689. doi:10.1016/j.nicl.2018.05.027

2. Tallantyre EC, Brookes MJ, Dixon JE, Morgan PS, Evangelou N, Morris PG. Demonstrating the perivascular distribution of MS lesions in vivo with 7-Tesla MRI. Neurology. 2008;70(22):2076-2078. doi:10.1212/01.wnl.0000313377.49555.2e

3. Tallantyre EC, Dixon JE, Donaldson I, et al. Ultra-high-field imaging distinguishes MS lesions from asymptomatic white matter lesions. Neurology. 2011;76(6):534-539. doi:10.1212/wnl.0b013e31820b7630

4. van Rooden S, Doan NT, Versluis MJ, et al. 7T t2∗-weighted magnetic resonance imaging reveals cortical phase differences between early- and late-onset alzheimer's disease. Neurobiology of Aging. 2015;36(1):20-26. doi:10.1016/j.neurobiolaging.2014.07.006

5. Saekho S, Boada FE, Noll DC, Stenger VA. Small tip angle three-dimensional tailored radiofrequency slab-select pulse for reducedb1 inhomogeneity at 3 T. Magnetic Resonance in Medicine. 2005;53(2):479-484. doi:10.1002/mrm.20358

6. Grissom W, Yip C-yu, Zhang Z, Stenger VA, Fessler JA, Noll DC. Spatial domain method for the design of RF pulses in multicoil parallel excitation. Magnetic Resonance in Medicine. 2006;56(3):620-629. doi:10.1002/mrm.20978

7. Alamouti SM. A simple transmit diversity technique for wireless communications. IEEE Journal on Selected Areas in Communications. 1998;16(8):1451-1458. doi:10.1109/49.730453

8. Goldsmith A. Wireless Communications. Cambridge University Press; 2013.

9. Yarnykh VL. Optimal radiofrequency and gradient spoiling for improved accuracy of T1 and B1 measurements using fast steady-state techniques. Magnetic Resonance in Medicine. 2010;63(6):1610-1626. doi:10.1002/mrm.22394

Figures

Input images for the proposed method: (a) GRE image acquired in CP mode (S1), (b) GRE image acquired with channel 1 having an additional transmit phase of π (S2), (c) GRE image of single channel acquisition for channel 1 (S3) and (d) GRE image of single channel acquisition for channel 2 (S4). All images are magnitude images of spoiled GRE acquisitions using a 7T dual channel TX MRI system and show transversal slice 50 from volunteer 1.

Validation of proposed method in transversal direction: (a,e) CP mode image; (b,f) intermediate combined magnitude image for channel 1 (C1); (c,g) intermediate combined magnitude image for channel 2 (C2); (d,h) final corrected homogeneous images using magnitude combination of C1 and C2. All images are magnitude images of spoiled GRE acquisitions using a 7T dual channel TX MRI system and show transversal slices 30, 40, 50, 60 and 70 from volunteer 1 (a-d) and volunteer 2 (e-h).

Validation of proposed method in transversal direction: (a,e) CP mode image; (b,f) intermediate combined magnitude image for channel 1 (C1); (c,g) intermediate combined magnitude image for channel 2 (C2); (d,h) final corrected homogeneous images using magnitude combination of C1 and C2. All images are magnitude images of spoiled GRE acquisitions using a 7T dual channel TX MRI system and show coronal slices 44, 56, 72, 86 and 100 from volunteer 1 (a-d) and volunteer 2 (e-h).

Validation of proposed method in transversal direction: (a,e) CP mode image; (b,f) intermediate combined magnitude image for channel 1 (C1); (c,g) intermediate combined magnitude image for channel 2 (C2); (d,h) final corrected homogeneous images using magnitude combination of C1 and C2. All images are magnitude images of spoiled GRE acquisitions using a 7T dual channel TX MRI system and show sagittal slices 44, 56, 72, 86 and 100 from volunteer 1 (a-d) and volunteer 2 (e-h).

Final homogeneous T2* weighted images of transverse slice 60 for Volunteer 1 (a) and 40 for Volunteer 2 (b). Iron rich tissue and blood vessels appear dark, grey matter and white matter in different shades of grey and cerebrospinal fluid appears white.

Proc. Intl. Soc. Mag. Reson. Med. 31 (2023)
4081
DOI: https://doi.org/10.58530/2023/4081