Bilguun Nurzed1, Dennis Hieronymi1, Thomas Wilhelm Eigentler1, and Thoralf Niendorf1,2,3
1Max-Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany, 2Experimental and Clinical Research Center (ECRC), a joint cooperation between the Charité Medical Faculty and the Max-Delbrück Center for Molecular Medicine, Berlin, Germany, 3MRI.TOOLS, Berlin, Germany
Synopsis
Keywords: RF Arrays & Systems, Bioeffects & Magnetic Fields
Investigation of the power correlation
matrix at 21.0 T is vital to elucidate the electrodynamics for human body and cardiac MR as well as the RF antenna performance. Recognizing this opportunity, this abstract assesses the power
losses for multi-channel RF transceiver arrays for human body and cardiac MR using 32-channel
pTx/Rx self-grounded bow-tie building block (SGBT). Losses were determined
inside the tissue, lossy material of the antenna, lumped elements of the tuning
and matching network and due to coupling and radiation.
Introduction
Pushing the boundaries of UHF-MR and unlocking the potential
of extreme field MR (EF, B0≥14.0T) [1] motivates research into the electrodynamics.
Loss analysis will be a major key to understand the electrodynamic constraints
and is essential for developing RF coils tailored for body and cardiac
EF-MR. Assessing where the power at EF dissipates will help us to minimize the
undesired losses outside the tissue as well as the comparison of different
array designs. Recognizing this opportunity this work elucidates the loss mechanism
at high spin excitation frequency (f=900MHz) using a 32-channel RF array
based upon self-grounded bow-tie building blocks (SGBT) [2].Methods
To investigate the loss
mechanism as a function of the magnetic field strength (B0) five 32-channel
SGBT arrays were modeled on the human voxel model Duke [3] (Figure
1a). With increasing B0 the antenna
dimensions were adapted to the 1H resonance
frequency at 7.0T (f=297.2 MHz) to 21.0T (f=900 MHz). The
shortened antennas were modeled at the identical center positions at each B0
resulting in an increased nearest neighbor distance (d). Furthermore, at
21.0T three high-density RF arrays with 32 (d=0mm) and 96 elements (d=0mm;10mm)
were modeled (Figure 2). For each SGBT a lossy parallel capacitor and a lossy
serial inductor were used for tuning and matching. EMF simulations were performed in CST
Microwave Studio 2020. The electrical material parameters
of the antennas and the tissue parameters provided by the IT‘IS Foundation [4] were adapted to the corresponding resonance frequency. Using a
framework for the calculation of the power correlation
matrices [5] the loss terms for the RF arrays were evaluated in Matlab 2019b. Using the power
correlation matrix and the single channel B1+ amplitude
the intrinsic (only sample losses) and realistic (sample,
coil, and coupling losses) transmit efficiency (TXE) were
evaluated in the entire 3D heart [6]. The sum of the magnitude of the superposed B1+
yields theoretical upper bounds for the RF arrays regarding TXE.Results
A worst-case
reflection of Sii < −15.6 dB was observed for all setups. At
21.0T the RF array revealed a coupling of Sij < −24.3 dB
(d=32.2mm) due to the increased nearest neighbor distance and Sij < −11.6 dB
(d= 0.0mm)
for the high-density RF arrays (Table 1). At 7.0T 77.8%
of the power dissipates into Duke, 11.8% in coupling, and 9.0% in the antenna (Figure 3). At
21.0 T with d= 32.2mm, 78.8% of the power dissipates into Duke, 0.2%
in coupling, and 18.5% in the antenna. For the 96-channel high-density array
(d=10mm) revealed 66.8% of the power dissipates into Duke, 4.0% in coupling, and
26.8% in the antenna. Losses in lumped elements and radiation accounted less
than 1.6% for all setups for each field strength. The inaccuracy of the power
correlation matrix was below 1.3%. The superposed B1+
maps (Figure 1b) revealed for the
7.0T setup a mean (min) TXE (realistic) of 0.57 (0.24) μT/√W
and for the high-density array (d=10mm) at 21.0T 0.64 (0.13) μT/√W
(Figure 2).Discussion
Highest undesired losses were found in the
antenna and coupling. The higher field strengths revealed increased losses in
the antenna. The distance between the antennas plays a vital role in the loss
mechanism where a closer setup causes more interaction between the antennas
resulting in additional losses and vice versa. The shortened antennas enabled
an improved channel count of 96 elements where the enhanced distance (0mm ->
10mm) revealed lower losses in the antenna and coupling resulting in a higher power transfer into the body. The narrow field of view (FOV) of the shortened antennas and the enlarged
distance (d=32.2mm) between the antennas towards 21.0T caused less interference of
the individual EMFs. This and the higher losses at 21.0T resulted in a
lower TXE compared to the 7.0T baseline setup with an identical channel
count (Figure 1b). The closer setup of the 32-channel array showed a better
overlap of the EMFs which resulted in a higher mean TXE compared with distanced
antennas. Yet, with 32-channel the minimum TXE could not be addressed which is
one of the main challenges of cardiac MR. The 96-channel setup with a higher
degree of freedom for B1+ shaping showed higher minimum
TXE compared to the 32-channel setup. The high-density RF array with enhanced
channel count and distance is beneficial to achieve the highest TXE.Conclusion
This simulation study provides a technical
foundation for the design and implementation of RF arrays tailored for body and
cardiac MRI at 21.0T. The spatial arrangement of an RF array needs to ensure low
coupling as well as an overlap of the individual EMFs to facilitate successful
control and shaping of the transmission field. For the feasibility of body and
cardiac MRI at 21.0T low loss RF antenna concepts with low mutual
coupling are required in order to achieve the highest transmit efficiency.Acknowledgements
This project has received funding from the
European Research Council (ERC) under the European Union's Horizon 2020
research and innovation program under grant agreement No 743077 (ThermalMR).References
1. Niendorf T,
Barth M, Kober F, Trattnig S (2016) From ultrahigh to extreme field magnetic
resonance: where physics, biology and medicine meet. Magnetic Resonance
Materials in Physics, Biology and Medicine 29 (3):309-311.
2.
Eigentler TW, Kuehne A, Boehmert L, Dietrich S, Els A, Waiczies H, Niendorf T
(2021) 32-Channel self-grounded bow-tie transceiver array for cardiac MR at
7.0T. Magn Reson Med 86 (5):2862-2879.
3.
Christ A, Kainz W, Hahn EG, Honegger K, Zefferer M, Neufeld E, Rascher W, Janka
R, Bautz W, Chen J, Kiefer B, Schmitt P, Hollenbach HP, Shen J, Oberle M,
Szczerba D, Kam A, Guag JW, Kuster N (2010) The Virtual Family - Development of
surface-based anatomical models of two adults and two children for dosimetric
simulations. Physics in Medicine and Biology 55.
4.
Foundation II (2018) Tissue Properties Database V4.0. doi:10.13099/VIP21000-04-0.
5.
Kuehne A, Goluch S, Waxmann P, Seifert F, Ittermann B, Moser E, Laistler E
(2015) Power balance and loss mechanism analysis in RF transmit coil arrays.
Magn Reson Med 74:1165-1176.
6. Georgakis IP, Polimeridis AG,
Lattanzi R (2020) A formalism to investigate the optimal transmit efficiency in
radiofrequency shimming. NMR Biomed 33:1-18.