Dmitrij Kravchenko1,2, Alexander Isaak1,2, Narine Mesropyan1,2, Claus Christian Pieper1, Daniel Kuetting1,2, Leon M. Bischoff1,2, Shuo Zhang3, Christoph Katemann3, Johannes M. Peeters4, Oliver Weber3, Ulrike Attenberger1, and Julian Luetkens1,2
1Diagnostic and interventional radiology, University Hospital Bonn, Bonn, Germany, 2Quantitative Imaging Laboratory Bonn, Bonn, Germany, 3Philips GmbH Market DACH, Hamburg, Germany, 4Philips MR Clinical Science, Best, Netherlands
Synopsis
Keywords: Machine Learning/Artificial Intelligence, Cardiovascular
AI assisted upscaling of low-resolution cine bSSFP images
yields comparable image quality to conventional images with no clinically
significant difference in volumetric data at a reduction of acquisition time by
a factor of 1.5 to 2.
Keywords: Artificial intelligence, acceleration, Superresolution,
cardiac MRI
Introduction
Cardiac magnetic resonance imaging (MRI) is a vital tool in
a number of pathologies such as acute myocarditis or myocardial infarction.
Unfortunately, long acquisition
times and confined spaces limit the use of cardiac MRI. Novel imaging techniques offer a solution to one of
these problems by reducing acquisition times using undersampling in combination
with artificial intelligence (AI) reconstruction. In this study we compared
standard ECG-triggered balanced steady state free precession (bSSFP) cine
images to ECG triggered low-resolution bSSFP images with AI upscaling, termed superresolution. Methods
Cardiac MR cine images were acquired in healthy volunteers
using standard ECG-triggered bSSFP techniques as well as ECG-triggered low-resolution
bSSFP with AI upscaling. Images were reconstructed using a vendor provided
prototype (Philips SmartSpeed Precise Image). This AI-based reconstruction
technique consists of a series of convolutional neural networks (CNNs):
Adaptive-CS-Net (1) allowed to
reconstruct images acquired with Compressed SENSE based variable density
undersampling patterns. CNN was applied during coil combination, removing the
noise and undersampling artifacts from the images in order to obtain good image
quality from accelerated acquisitions (2). Subsequently,
Precise Image Net, an AI-model, was applied to remove ringing artefacts and to
replace the traditional zero-filling strategy to increase the matrix size and
therewith the sharpness of the images; these type of networks are known as superresolution
networks (3, 4). The network was
trained on pairs of low- and high-resolution data with k-space crops to induce
ringing. Data consistency checks were implemented to match the resulting
k-space with the measured k-space data. The full reconstruction pipeline
generates images with improved SNR and sharpness, higher matrix size and
reduced ringing artefacts, and can be applied to all 2D cartesian acquisitions.
The level of noise reduction was variable with settings ranging from weak,
moderate, strong, to maximum. Left ventricular ejection fraction (LVEF), left
ventricular end diastolic volume (LVEDV), left ventricular end diastolic volume
index (LVEDVi), and interventricular septum thickness at diastole (IVSD) were
compared using the repeated measures analysis of variance (ANOVA) between six
groups: native resolution, low-resolution, and low-resolution with upscaling
and different levels of denoising. Acquisition times were compared using the
student’s paired t test. Apparent signal to noise ratios (aSNR) and apparent
contrast to noise ratios (aCNR) were calculated as previously described and
compared using repeated measures ANOVA (5). Subjective image
quality assessment was rated by two radiologists for all short-axis and
4-chamber views for all six data sets on a 5-point rating scale regarding three
image criteria: blood-pool to myocardium contrast, endocardial edge definition,
and artefacts, as previously described (6). An overall score
was determined by the equal weight average of all three criteria: 1 non-diagnostic,
5 excellent. Results
Data from 10 participants was acquired (30.3±3.1 years old,
8 male). Acquisition duration for superresolution acquisitions was
significantly shorter than standard acquisitions for 4-chamber views (6.20±1.42
vs 8.64±0.33 s, p=.0003) as well as for short-axis view (53.84±8.47 vs
93.83±10.04 s, p<.0001). Subjective image quality was not significantly
different between the groups for 4-chamber views or short-axis views (p=.685
and p=.073). Interclass correlation ranged from excellent for conventional
4-chamber views (0.966, CI 0.891-0.989) to good, for example low-resolution
4-chamber views (0.812, CI 0.403-0.941). No difference was found between the
groups regarding aSNR and aCNR (p=.42 and p=.42). Regarding volumetry, no
difference was observed between the groups regarding LVEF (p=.64) or IVSD
(p=.72). Significant differences in LVEDV were noted between the standard
sequence and superresolution with maximum denoising (171.1±32.7 ml vs
164.3±31.2, p=.039), superresolution weak vs superresolution maximum (167.7±31.7
vs 164.3±31.2, p=.0006), superresolution moderate vs superresolution strong
(167.8±31.6 vs 165.5±30.8, p=.025), and superresolution moderate vs superresolution
maximum (167.8±31.6 vs 164.3±31.2, p=.005). Discussion
AI assisted upscaling led to significantly faster
acquisition time averaging about 1.5 up to 2 times faster than conventional
techniques without affecting aSNR and aCNR, as a result of improved SNR by
lower resolution which can then be sacrificed for higher acceleration factors.
While volumetric analysis showed nearly identical results for LVEF and IVSD,
LVEDV demonstrated a tendency to decrease going from conventional sequences and
low-resolution sequences with maximum AI denoising. One possible reason for
this may be progressive denoising at the edge of high contrast areas, such as
the blood pool and myocardium, as the denoising algorithm increases, leading to
discrete changes in perceived volume. On the other hand, results might be
biased due to the small participant group and this phenomenon might disappear
as more data sets are analyzed. Clinically, the differences were not relevant
with the largest mean difference being 6.7 ml (conventional vs maximum AI) and
about 3 ml for the other differences. Although low-resolution images without
upscaling did appear unsharp compared to their native resolution and upscaled
counterparts, no statistical differences were noted for any groups regarding
subjective image quality. This may be due to the fact that low-resolution
acquisitions without upscaling scored relatively high for the artefacts
category, thus inflating the aggregate subjective image quality score while the
other two categories, contrast and edge definition, scored less. Conclusion
Low-resolution AI assisted upscaling of cardiac cine MRI
sequences leads to a significant reduction in acquisition times without a
significant difference in volumetric results or subjective image quality in
healthy volunteer participants. Acknowledgements
No acknowledgement found.References
1. Pezzotti N, Weerdt E de, Yousefi S, Elmahdy MS, van Gemert J, Schülke C et al. Adaptive-CS-Net: FastMRI with Adaptive Intelligence; 13-Dec-19. Available from: URL: https://arxiv.org/pdf/1912.12259.
2. Peeters JM, Chung H, Valvano G, Yakisikli D, van Gemert J, de Weerdt E and van de Ven K. Philips SmartSpeed. No compromise Image quality and speed at your fingertips. Philips SmartSpeed white paper.
3. Dong C, Loy CC, He K, Tang X. Image Super-Resolution Using Deep Convolutional Networks; 31-Dec-14. Available from: URL: https://arxiv.org/pdf/1501.00092.
4. Li Y, Sixou B, Peyrin F. A Review of the Deep Learning Methods for Medical Images Super Resolution Problems. IRBM 2021; 42(2):120–33. Available from: URL: https://www.sciencedirect.com/science/article/pii/S1959031820301408.
5. Bischoff LM, Katemann C, Isaak A, Mesropyan N, Wichtmann B, Kravchenko D et al. T2 Turbo Spin Echo With Compressed Sensing and Propeller Acquisition (Sampling k-Space by Utilizing Rotating Blades) for Fast and Motion Robust Prostate MRI: Comparison With Conventional Acquisition. Investigative Radiology 2022; 00004424-990000000-00049(September 1, 2022):10.1097/RLI.0000000000000923. Available from: URL: https://journals.lww.com/investigativeradiology/Fulltext/9900/T2_Turbo_Spin_Echo_With_Compressed_Sensing_and.49.aspx.
6. Pednekar AS, Wang H, Flamm S, Cheong BY, Muthupillai R. Two-center clinical validation and quantitative assessment of respiratory triggered retrospectively cardiac gated balanced-SSFP cine cardiovascular magnetic resonance imaging in adults. J Cardiovasc Magn Reson 2018; 20(1):44. Available from: URL: https://jcmr-online.biomedcentral.com/articles/10.1186/s12968-018-0467-6.