Yishi Wang1, Zhen Zhang2, Rui Wang2, Xiuzheng Yue1, Fang Wang2, Rongrong Zhu2, and Ruoshui Ha2
1Philips Healthcare, Beijing, China, 2Medical Imaging Center, People's Hospital of Ningxia Hui Autonomous Region, Yinchuan, China
Synopsis
Keywords: Pulse Sequence Design, Diffusion/other diffusion imaging techniques, Cardiac Diffusion, Black Blood, Motion Compensation
Second-order motion-compensated diffusion imaging is a robust solution for cardiac diffusion imaging but prone to bright blood signals due to motion compensation itself. In this work, we incorporated black blood module into second-order motion-compensated sequence to achieve cardiac diffusion imaging with blood signal suppressed.
Introduction
As an emerging motion compensation
technique, a second-order motion-compensated spin echo (M2-SE) sequence can be
used to reduce the signal loss for cardiac diffusion imaging [1]. However, the
signal of the flowing blood is also compensated to induce a bright blood signal
in the acquired diffusion images. Double inversion recovery (DIR) is a common
scheme to obtain black blood images of the heart such as in T2 weighted images.
This study aimed to combine M2-SE and DIR to achieve robust black blood cardiac
diffusion imaging.Methods
The combination of DIR and M2-SE was implemented
on the Philips MR platform. The sequence was tested on two healthy volunteers
on two different 3T scanners (Ingenia CX and Ingenia Elition X, Best, the
Netherlands). The sequence diagram was shown in Figure 1. All diffusion
weighted images were acquired in the short-axis plane using two schemes on each
scanner. One used M2-SE combined with DIR and the other used M2-SE only. The
sequence was performed with an ECG trigger and breath hold. The inversion delay
of DIR was calculated automatically according to the heart rate and the repetition
time was set to 2 heartbeats. The ECG trigger delay was set to mid-systole
according to a previous study. The detailed scan parameters are shown in Table
1. Results
Figure 2 showed the images of the first
volunteer. There was a hyperintensity blood signal in the b=0 image using M2-SE
only, while the signal contamination from the blood pool was well suppressed in
the images of all three b values when DIR was combined with M2-SE. Figure 3
showed the images of b=400 s/mm2 of the second volunteer. There was
an obvious residual blood pool signal in the image from M2-SE while the blood
signal was well suppressed in M2-SE with DIR.Discussion and conclusion
M2-SE is an effective method for cardiac
diffusion imaging as an alternative to stimulated echo method. However, the
motion-compensated diffusion encoding gradient also compensated the blood flow
signal which may hamper the observation of real changes in the myocardium. We
incorporated DIR for blood signal suppression into M2-SE to achieve black blood
cardiac DWI and showed its feasibility on two different 3T scanners. Further
study should be performed to evaluate its clinical value with a larger cohort. Acknowledgements
No acknowledgement found.References
1. Stoeck, et.al. Second-order motion-compensated spin echo diffusion tensor imaging of the human heart. MRM, 2016.