Guangqi Li1, Xiaodong Ma2, Sisi Li1, Xinyu Ye1, Peter Börnert3,4, Xiaohong Joe Zhou5, and Hua Guo1
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China, 2Department of Radiology and Imaging Sciences, University of Utah, Salt Lake City, UT, United States, 3Radiology, C.J. Gorter Center for High-Field MRI, Leiden University Medical Center, Leiden, Netherlands, 4Philips Research, Hamburg, Germany, 5Center for MR Research and Departments of Radiology, Neurosurgery, and Biomedical Engineering, University of Illinois College of Medicine at Chicago, Chicago, IL, United States
Synopsis
Keywords: Data Acquisition, Diffusion Tensor Imaging
Different
multi-shot spiral sampling schemes have been developed for high-resolution DWI.
However, the performances of these sampling strategies such as variable-density
spiral (VDS), dual-density spiral (DDS) and uniform-density spiral (UDS) have
not been compared comprehensively. In this study, we compare multi-shot UDS,
VDS and DDS in brain DWI in terms of inter-shot phase error correction, overall
image quality and SNR performance. Both theoretical analysis and in-vivo
results demonstrate that UDS exhibits the best off-resonance performance among
the three spiral sampling patterns. Additionally,
UDS
achieves the highest SNR in diffusion imaging over the VDS and DDS
acquisitions.
Introduction
Diffusion-weighted imaging (DWI) has been
widely used in clinical diagnosis and neuroscience research. Various multi-shot
spiral samplings have been successfully developed to achieve high-resolution
DWI 1-4. For multi-shot DWI, one primary issue is to correct for
phase variations among different shots. The phase variations can be measured,
either by acquiring extra navigator signals or by computing this information
from imaging echoes. According to the ways of acquiring the phase information,
corresponding spiral sampling strategies can be divided into three main
categories, variable-density spiral (VDS) 5, dual-density spiral
(DDS) 6 and uniform-density spiral (UDS) 7. However, to
our knowledge, their performances in terms of image quality and efficiency have
not been fully compared. Moreover, their off-resonance and SNR performances are unclear and worthy of investigation. In
this study, we carried out a comprehensive comparison to investigate the
performances of UDS, VDS and DDS for multi-shot diffusion imaging.Methods
1. Spiral trajectories
design
To
set up a fair comparison, the number of interleaves and the readout durations
of these spiral samplings are kept the same. Under-sampling along the radial
direction 8 is adopted to increase the radial spacing of DDS and VDS
so that their readout durations can be reduced to the same length as UDS.
2. Data acquisition
All
experiments were performed on an Ingenia CX 3.0T scanner (Philips Healthcare,
Best, The Netherlands) using a 32-channel head coil. The gradient system was
operated at a maximum gradient strength of 31 mT/m and with a maximum slew rate
of 200 T/m/s.
Experiment 1, 4-shot
acquisition: FOV=210×210mm2, resolution=1.28×1.28mm2, matrix=164×164,
b value=1000 s/mm2, 12 diffusion directions, TE/TR=55/3000ms,
readout duration=26.0ms.
Experiment 2, 6-shot acquisition: FOV=210×210mm2,
resolution=0.99×0.99×4.0mm3, acquisition matrix=212×212, b
value=1000 s/mm2, 12 diffusion directions, TE/TR=55/3000ms, readout
duration=26.0ms.
In all experiments, SPIR technique was
used to suppress fat signals. In addition, low-resolution field maps acquired
using a multi-echo GRE sequence were used for deblurring. The 2D T2-weighted
TSE images and T2W-FLAIR images were acquired as anatomical references. The
resolution of the anatomical images matches the spiral DWI images of each scan.
3. Image reconstruction and processing
The diffusion images were off-line
reconstructed using the POCS-ICE algorithm 9, followed by
off-resonance correction. Color-coded FA maps were calculated using FSL toolbox 10.
SNR performance of the three spiral samplings was evaluated using a Monte
Carlo-based pseudo multiple replica method 11.Results and Discussion
1. Off-resonance performance
Figure
1 show the mean DWI with an in-plane resolution of 0.99×0.99mm2 acquired
by 6-shot UDS, VDS and DDS acquisitions, respectively. Six representative slices of the same
subject from Experiment 2 are shown. The T2W-FLAIR images are
shown in the bottom row as anatomical references. In general, the UDS-, VDS-
and DDS-based diffusion images all provide satisfactory anatomic integrity and
geometric fidelity in the regions where the B0 inhomogeneity is not so severe. Residual blurring artifacts (yellow dashed
circle) can be observed in the deblurred DDS images. There is a slight error
in the VDS images (yellow arrow head). UDS exhibits the lowest static B0
off-resonance artifacts. This in vivo results indicate that UDS has better
off-resonance performance than DDS and VDS. Thus UDS is suitable for
high-efficiency diffusion imaging with long spiral readouts because it is less
vulnerable to off-resonance effect.
Figure
2 shows the color-coded FA maps obtained by the 6-shot UDS, VDS and DDS
samplings from Experiment 2. Seven slices
are shown. Multi-shot UDS, VDS and DDS diffusion imaging can provide correct
DTI metrics. However, it is obvious that the cFA results of DDS DW images are a little noisy compared to other two counterparts.
2.
Inter-shot phase error correction
Figure
3 shows the evolutions of estimated phases of 4 shots during the POCS-ICE
iterative reconstruction for one set of the in vivo data from
Experiment 1. The phase errors of each shot are estimated from the central
k-space data, which is densely sampled for VDS and DDS, but under-sampled for
UDS. The estimated phases for VDS and DDS after one iteration was very close to
the final iteratively updated inter-shot phase. In comparison, more number of
iterations are required to reach a stable estimation of shot-to-shot phase
errors for UDS than for the other two spirals.
3.
SNR performances
Figure 4 shows the SNR maps of b=1000 s/mm2
diffusion images with in-plane resolution of 0.99mm2 acquired by the
three spiral acquisitions from Experiment 2. The SNR maps from six representative slices are shown. The
corresponding mean SNR value across the whole brain is marked in the upper left
corner of the image. UDS shows the best SNR performance among the three spiral
samplings in the diffusion images with the same TE.Conclusion
This study performed a comprehensive comparison of
UDS, VDS and DDS acquisitions for multi-shot diffusion imaging. The results
demonstrate that UDS provides superior
off-resonance performance and SNR performance over VDS and DDS samplings.Acknowledgements
No
acknowledgement found.References
1 Liu C, Bammer R, Kim Dh, Moseley ME.
Self-navigated interleaved spiral (SNAILS): application to high-resolution
diffusion tensor imaging. Magn Reson Med 2004;52:1388–1396.
2 Liu C, Moseley ME, Bammer R.
Simultaneous phase correction and SENSE reconstruction for navigated multi-shot
DWI with non-cartesian k-space sampling. Magn Reson Med 2005;54:1412–1422.
3 Truong TK, Chen NK, Song AW. Inherent
correction of motion-induced phase errors in multishot spiral
diffusion-weighted imaging. Magn Reson Med. Oct 2012;68(4):1255-61.
4 Truong TK, Guidon A. High-resolution
multishot spiral diffusion tensor imaging with inherent correction of
motion-induced phase errors. Magn Reson Med. Feb 2014;71(2):790-6.
5. Kim DH, Adalsteinsson E, Spielman DM.
Simple analytic variable density spiral design. Magn Reson Med. Jul
2003;50(1):214-9.
6. Lin W, Börnert P, Huang F, Duensing GR,
Reykowski A. Generalized GRAPPA operators for wider spiral bands: rapid
self-calibrated parallel reconstruction for variable density spiral MRI. Magn
Reson Med. Oct 2011;66(4):1067-78
7. Glover GH. Simple analytic spiral
K-space algorithm. Magn Reson Med. Aug 1999;42(2):412-5.
8. Li G, Ye X, Shao X, et al. Four-shot Navigator-free Spiral Acquisition Strategy
for High-resolution Diffusion Imaging. In: Proceedings of the 29th Annual
Meeting of ISMRM. 2021. p1328.
9. Guo H, Ma X, Zhang Z, Zhang B, Yuan C,
Huang F. POCS-enhanced inherent correction of motion-induced phase errors
(POCS-ICE) for high-resolution multishot diffusion MRI. Magn Reson Med. Jan
2016;75(1):169-80.
10. Jenkinson M, Beckmann CF, Behrens TE,
Woolrich MW, Smith SM. FSL. Neuroimage 2012;62(2):782-790.
11 Robson PM, Grant AK, Madhuranthakam AJ,
Lattanzi R, Sodickson DK, McKenzie CA. Comprehensive quantification of
signal-to-noise ratio and g-factor for image-based and k-space-based parallel
imaging reconstructions. Magnetic resonance in medicine 2008;60(4):895-907.