William Mathieu1, Charbel Matta1, Milica Popović1, and Reza Farivar2
1Electrical and Computer Engineering, McGill University, Montreal, QC, Canada, 2Ophthalmology, McGill University, Montreal, QC, Canada
Synopsis
Keywords: RF Arrays & Systems, RF Arrays & Systems
We present a size-adaptable flexible-plate 128-channel receive coil
design with initial results. Our approach allows for increased element density as
elements can be placed closer to the surface of the scalp. Arrays are built on
semi-flexible plates with boundaries roughly following cranial sutures. The aim
is to accommodate at least 99% of head shapes and sizes. Our 18-channel
“occipital” plate array (1 of 8 plates) was compared to the posterior portion
(40-channels) of the Siemens 64 Head/Neck coil using SNR maps. Results show
that our design improves average and max SNR by 1.9- and 7.4-fold,
respectively.
Introduction
Image quality in MRI is heavily dependent on the proximity of receptive
elements to their target volume. The long-standing design paradigm in receive-only
coils is a phased-array which necessitates a rigid body design to ensure maximum
stability to support the highest possible element count1-4. This
one-size-fits-all approach may improve the signal-to-noise ratio (SNR) for a
head which happens to perfectly fit the rigid shell, however this is seldom the
case. The advantages that a high channel count coil provides are squandered if
the shape of the coil does not optimally fit a given head. Our design aims to
decrease element diameters to the extent that SNR is maximized. However, in
doing so, the receptive fields of elements decrease, an effect that can be offset
by maximizing filling factor, which we achieve through flexible, movable
plates. On these plates generic phased arrays are constructed, and between
plates we employ bending elements on ultra-flexible film. Our goal is to construct
a coil which comfortably fits any adult head shape and size to the 99th
percentile, and in doing so eliminating any SNR-gap between individuals. Here
we present the design of the complete 128-channel coil, and the performance of
a single “occipital” plate, compared to a commercial Siemens head coil.Methods
An average human head model was used as the initial shape from which
flexible plates were partitioned. Plate boundaries roughly follow the locations
of the cranial sutures. Boundaries were then adjusted for ideal coverage of the
brain and optimal element tessellation; an early iteration is shown in Figure
1a. The final configuration is shown in Figure 1b, which features 8 plates (eye
plate not seen). Plates were 3D printed in flexible thermoplastic polyurethane (TPU)
material.
To fulfill the requirement of 128 channels, with the
consideration of ensuring adequate signal at cortical depths, element diameters
were set at 36-mm. Each plate is attached to a non-magnetic pneumatic piston, which
moves the plate into position, normal to the surface of the head, see Figure 1c
and d. To fill the gaps between plates, bending elements constructed on thin
film will be employed, outlined in previous work5. The pistons, plates, and
circuitry are supported by an outer shell, seen in Figure 1e, which plugs into the
patient table of a standard 3T clinical Siemens Prisma scanner (Siemens Healthineers,
Erlangen, Germany).
A standard element receive-only circuit was employed,
shown in Figure 2(a). Elements were tuned, matched, and preamps were decoupled using
standard methods6 with a 2-port vector network
analyzer (Copper Mountain Technologies, Indianapolis, USA). Given the limited
surface area of the plates, all electrical components, excluding capacitors CT
and C1, were placed on vertical printed circuit boards (PCBs), normal
to the surface of the plates, two such boards are shown in Figure 2c. Each
board supports two element circuits, one on each side.
To secure elements to the plates while allowing them
to move with different head shapes and sizes, loop wires were passed through the
plates, turning the plates into a substrate for the
coil circuits. Four such elements constructed in this way are shown in Figure 2b
and c, corresponding to the outside and inside surfaces of a plate, respectively.
MRI experiments were conducted using the Siemens
Prisma. The occipital portion of our coil, see Figure 3, was tested against the
posterior section of a Siemens 64 Head/Neck coil (40-channels). A standard 1.9-L
Siemens bottle phantom was used. An integrated calculation environment (ICE), IceMGHCoilArrayReconUtil, for RF coil quality evaluation was provided by the Martinos
Center for Biomedical Imaging (MGH Department of Radiology, Harvard Medical
School)7. The environment outputs optimally combined SNR maps and channel noise
covariance matrices. The SNRMap sequence had the following scanner parameters: TE = 6
ms, TR = 30 ms, FOV = 383 mm × 383 mm, slice-thickness = 5 mm, orientation =
transversal, BW = 395 Hz/Pixel, averages = 4, and FA = 10o.Results and Discussion
Figure 4 shows the SNR performance comparison between the two tested
coils. All SNR maps were windowed to 150 units. For both the average and
maximum values the occipital array outperforms the Siemens 64 Head/Neck. As
reported in Figure 4, average and max SNR values, were 12 and 135 for the
Siemens coil, and 23 and 998 for our occipital coil. Despite the higher element
count, the Siemens coil is limited by its inability to conform to the shape of
the bottle phantom. Channel noise correlation matrices were also collected, see Figure 5, showing
higher overall noise in the occipital array. Average channel noise correlations were nnn for the siemens coil
and mmm for the occipital
array.Conclusion
Channel count has been limited by the inability to place small elements
close enough to their sample. The
results presented here show that our approach, using semi-flexible plates,
allows for smaller element loop diameters leading to higher SNR without
sacrificing sample coverage.Acknowledgements
I'd like to thank my supervisors Reza Farivar and Milica Popović and my colleague Charbel Matta. I'd also like to thank Gerald Moran at Siemens Healthineers for his continual support of the project. This project was financially supported by Mitacs in partnership with Siemens Healthineers and MEDTEQ+.
References
- Farivar, R., Grigorov, F., van der
Kouwe, A.J., Wald, L.L. & Keil, B. Dense, shape‐optimized posterior
32‐channel coil for submillimeter functional imaging of visual cortex at 3T. Magnetic resonance in medicine 76, 321-328 (2016).
- Ghotra, A., et al. A size‐adaptive 32‐channel array coil for awake infant
neuroimaging at 3 Tesla MRI. Magnetic
Resonance in Medicine 86,
1773-1785 (2021).
- Keil, B., et al. A 64‐channel 3T array coil for accelerated brain MRI. Magnetic resonance in medicine 70, 248-258 (2013).
- Wiggins, G.C., et al. 96‐Channel receive‐only head coil for 3 Tesla: design
optimization and evaluation. Magnetic
Resonance in Medicine: An Official Journal of the International Society for
Magnetic Resonance in Medicine 62,
754-762 (2009).
- Mathieu, W., Popović, M. &
Farivar, R. Towards Brain MRI Adaptable to Head Size: Bowing RF Coil Phased
Arrays. IEEE Journal of Electromagnetics,
RF and Microwaves in Medicine and Biology (2021).
- Keil, B. & Wald, L.L. Massively
parallel MRI detector arrays. Journal of
magnetic resonance 229, 75-89
(2013).
- Polimeni, J.R.
& Wald, L.L. Documentation for icemghcoilarrayreconutil software. (Harvard Medical School, 2009).