Lukas Hingerl1, Bernhard Strasser1, Gilbert Hangel1,2, Stanislav Motyka1, Fabian Niess1, Eva Niess1, Alexandra Lipka1, Dario Goranovic1, Philipp Lazen1, Stephan Gruber1, Siegfried Trattnig1,3, and Wolfgang Bogner1
1Department of Biomedical Imaging and Image-guided Therapy, Radiology and Nuclear Medicine, Medical University of Vienna, HFMR Centre, Vienna, Austria, 2Department of Neurosurgery, Medical University of Vienna, Vienna, Austria, 3Institute for Clinical Molecular MRI in Musculoskeletal System, Karl Landsteiner Society, Vienna, Austria
Synopsis
Keywords: Spectroscopy, Spectroscopy, Lipid Fat Suppression Removal
We present an elegant and easy to implement method for MRI and CSI steady-state sequences to remove or suppress
lipids or other components with short transverse relaxation times by neither introducing additional
pulses nor hardware and by just altering the excitation pulse phase.
Introduction
Lipids
are one of the largest sources for corrupt data in Magnetic Resonance
Spectroscopic Imaging (MRSI) since they contaminate the signal and
decrease the quality of findings1. Lipids also resonate spectrally on
a similar chemical shift as the important tumor biomarker and
metabolite lactate and hamper therefore lactate imaging in the
diseased tissue since their signal magnitude can excel the lactate
signal by several factors. Spatially, as well as spectrally, lipid
signals can leak on completely wrong positions or resonances,
impeding therefor the proper quantification of other metabolites.
Also, in the field of MRI off-resonances make the separation of water
and fat difficult since the individual signals mix up2. Different
methods to prevent, suppress or remove lipid signals exist like
Fat-Sat; the Dixon method, mathematical algorithms (L1/L2-regularization); lipid-inversion pulses (STIR); or
subtraction-based spectral-selective methods (MEGA-editing) or
hardware-based lipid-crushing via external coils.
The purpose of
this work is to present a novel method to remove/suppress lipids for
steady-state sequences by neither introducing additional pulses nor
hardware.Methods
Radio-frequency (RF) spoiling3 used in GRE-MRI is a
widely used tool to suppress the transverse signal of components with
long T2 by quadratically incrementing
the excitation pulse phase from TR to TR (e.g. phase increment $$$\theta$$$=117° - see Figure
1,top). ‘RF-phase-based’ or ‘Partial RF
spoiling’ methods4,5 for T2-mapping exploited the strong T2-dependence – which forms the basis
for the proposed method here – as well as the resilience concerning
different T1s. An interesting signal behavior at 3T (Figure
1,bottom) can be observed in Bloch simulations as well as via the
analytic solution6 if TRs ~<300ms but >~100ms: Lipids (short
T2s) remain invariant for all incremented phases but long T2
components not. By complex subtraction of two scans with different
phase increments lipids and fat can be nulled. The optimal values for $$$\theta_1$$$ and $$$\theta_2$$$=-$$$\theta_1$$$ was selected by the simulated signal maximum of
the compound of interest.
We
measured three sequences on a Siemens 3T Prisma: A spoiled, unbalanced GRE
with 196x196 matrix, FOV 200x200 mm2, slice thickness 10 mm, TR=160ms, FA=65°, 2 averages with $$$\theta$$$=+24°, $$$\theta$$$=-24° and $$$\theta$$$=0° (instead of traditional RF spoiling) and both, a spoiled,
unbalanced un-localized FID and a FID-MRSI sequence (TE=1ms, TRs=230ms,
FA=80°, $$$\theta$$$=+40°, $$$\theta$$$=-40°, $$$\theta$$$=0° and 500 averages for FID and 20 for FID-MRSI). Both featured a 3-pulse water
suppression module and a short signal readout of 128 spectral points
with 1000Hz spectral bandwidth. The matrix size of the FID-MRSI sequence was 8x8, across a FOV=200x200mm2, slice thickness 10mm.
The GRE
sequence was used for imaging of the T2-array of a NIST7 phantom
together with a turbo-spin-echo sequence for determining the T2
values and the FID/FID-MRSI sequences where used on a home-built
spectroscopy phantom (water, N-acetylaspartat, creatine, choline-compounds,
glutamate, glutamine, myo-inositol, lactate) with corn oil-soaked textiles wrapped around to
mimic lipid layers.Results
Figure
1, top shows pulse-acquire Bloch-simulation results of the first time
point of long- and short-T2 components for traditional RF spoiling
with short TRs and rather large FAs. By increasing the TR
and the FA, it can be observed that at certain pulse phase increments
short-T2
components can be cancelled by subtraction,Figure 1, bottom. Figure
2 shows simulated spectra before and after suppression, noised as
well as noiseless. GRE imaging results and T2-maps of the T2-array of
a NIST phantom are displayed in Figure 3 which demonstrates the
excellent suppression (factor ~100 for shortest T2s) results of the method. The dashed line
indicates the position of the line profile in Figure 4,left. Figure 4,
right, shows how the signal changes for different T2s. Figure 5 shows MR
spectroscopy results of the un-localized and the localized scan and
the excellent recovery of lipid free spectra (~90% suppression efficiency) with good baselines,
however at the cost of significantly reduced SNR since only ~5-10% of the original signal remains (given realistic T2 values).Discussion & Conclusion
Although
the method has yet not been successfully tested in vivo, its
advantages are numerous: 1) No additional radio frequency pulses are
needed, i.e. no increased energy disposition and no pulse
imperfections; 2) the phase of the excitation pulse is more
accurate to set than the flip angle of any other potentially needed
pulses; 3) the suppression is spectrally and spatially global, i.e.,
not dependent on the resonance band; 4) no 'co-suppressed' signals as
in MEGA-editing, Fat-Sat or regularization algorithms
since T2s of lipids are distinctively small; 5) the method is related to RF spoiling and
easy to adapt and the post processing is straightforward.
The
main drawback of the method is the inefficiency of the obtained SNR. The second drawback is the motion sensitivity of the method since
phase changes in the observed signal are of the same order of
magnitude as the desired phase modulations.
We
presented an elegant and easy to implement method to remove/suppress
short T2 components such as lipids by merely altering the excitation
pulse phase of a steady-state sequence, without the need of
additional pulses or hardware. Additional ways to improve the SNR
inefficiency are however required to capitalize from the numerous
advantages of this potential tool for in vivo applications. For non-localized whole-brain MRS the method could be already relevant8.Acknowledgements
Austrian Science Fund (FWF): Projekt number TAI-676References
1. Tkáč I, Deelchand D, Dreher W, et al. Water and lipid suppression
techniques for advanced 1 H MRS and MRSI of the human brain: Experts' consensus
recommendations.
NMR Biomed. 2021;34(5):e4459. doi:10.1002/nbm.4459
2. Bley TA, Wieben O, François CJ, Brittain JH, Reeder SB. Fat and water
magnetic resonance imaging. J Magn Reson Imaging. 2010;31(1):4-18.
doi:10.1002/jmri.21895
3. Zur, Y., Wood, M.L. and Neuringer, L.J., Spoiling of transverse
magnetization in steady-state sequences. Magn. Reson. Med. 1991, 21: 251-263.
https://doi.org/10.1002/mrm.1910210210
4. Wang X, Hernando D, Reeder SB. Phase-based T2 mapping with gradient echo imaging [published correction appears in Magn Reson Med. 2022 Aug;88(2):1015]. Magn Reson Med. 2020;84(2):609-619. doi:10.1002/mrm.28138
5. Bieri O, Scheffler K, Welsch GH, Trattnig S,
Mamisch TC, Ganter C. Quantitative mapping of T2 using partial spoiling. Magn
Reson Med. 2011;66(2):410-418. doi:10.1002/mrm.22807
6. Tamada D, Field AS, Reeder SB. Simultaneous T1 -weighted and T2 -weighted 3D MRI using RF
phase-modulated gradient echo imaging. Magn Reson Med. 2022;87(4):1758-1770.
doi:10.1002/mrm.29077
7. https://www.nist.gov/publications/magnetic-resonance-imaging-biomarker-calibration-service-proton-spin-relaxation-times
8. Davitz MS, Wu WE, Soher BJ, et al. Quantifying global-brain metabolite
level changes with whole-head proton MR spectroscopy at 3T. Magn Reson Imaging. 2017;35:15-19. doi:10.1016/j.mri.2016.08.012