Guy Shlomo Baz1,2, Edna Furman-Haran 2,3, and Rita Shmidt1,2
1Department of brain sciences, Weizmann Institute, Rehovot, Israel, 2The Azrieli National Institute for Human Brain Imaging and Research, Weizmann Institute, Rehovot, Israel, 3Life Sciences Core Facilities, Weizmann Institute, Rehovot, Israel
Synopsis
Keywords: fMRI (task based), High-Field MRI
Ultra-high field MRI provides increased
sensitivity, which we aim to utilize for improving the temporal resolution in
functional studies. To investigate the achievable resolution at 7T MRI, a
dynamic phantom that can generate an fMRI-like time-series was used. A dataset
based on block-design with defined time shifts and a range of contrast-to-noise
values was used to characterize the effective temporal resolution. Estimated
temporal resolution was x1.7 times better for multi-echo compared to
single-echo EPI, estimated as 146ms for a scan with TR of 600ms. This study
offers a novel approach of optimizing protocols and new insights into fMRI
temporal resolution.
Introduction
Ultra-high field 7T
human MRI offers increased BOLD sensitivity as well as higher temporal signal-to-noise (tSNR) for functional
studies1. While many studies use the increased sensitivity to achieve
higher spatial resolution, in this study we aim to examine methods to increase
temporal resolution. Multi-echo EPI was demonstrated useful for better capturing
the BOLD signals, due to its increased tSNR (see Fig. 1) as well as the
capability to separate the BOLD and non-BOLD contributions2-4. Improved tSNR is also expected to contribute
to an increase in effective temporal resolution (ETR). To characterize the ETR of
an experimental protocol, it is required to generate a BOLD-like signal in a
controllable manner. This can be achieved using a dynamic phantom that can
simulate the BOLD signal changes inside the scanner. In previous studies, such
phantom has enabled a better definition of the temporal noise characteristics
by mimicking the BOLD time-series of a resting-state experiment5,6. In
the current study, the expected signal of a block-design paradigm was generated
using a dynamic phantom to characterize the ETR at 7T MRI and optimize it by
using multi-echo EPI. Methods
The dynamic phantom’s compartments
consist of a controller with 2 valves, delivering around 85psi air pressure to
a pneumatic motor attached to a cylindrical head filled with agarose gel and an
optical encoder (ALA scientific inc., New York, USA7; Fig. 2A). Each valve
rotates the head either clockwise or counterclockwise with controllable rotated
angles. The optical encoder supplies indication on the actual movement. By
rotating the agarose head, an increase or decrease of the signal at a certain
group of voxels can be achieved (Fig. 2B), where two compartments- “Active” (T2*
=48.5ms) and “Inactive” (T2*=44.5ms) define the
range of signal change. The target tSNR of the phantom experiments (Fig. 3A) was
set based on an in-vivo scan (see scan parameters in Fig.1) by adjusting the flip
angle. Scan parameters of the multi-echo EPI8,9: TR = 600 ms, TEs =
14.4, 43.6 & 72.9 ms, in-plane resolution 2.13x2.13 mm2, slice
thickness 3 mm.
The dynamic signal was designed
to mimic a BOLD signal in a block-design paradigm and consisted of 6 blocks of 12
seconds “on” followed by an 18 seconds “off” period. It was generated as a
convolution of a stimuli vector with the hemodynamic response function (using
SPM software package; Fig. 2D). Two-hundred TRs at the beginning of the scan
were added to estimate the tSNR without any “task”. During scanning session, 6
repetitions of 3 types of the mentioned signal were performed, each with a
different time shift of the stimuli onset: 0 ms, 100 ms & 600 ms. To
identify the voxels of interest that represent the dynamic signals, a standard
general linear model analysis was performed, and voxels with t-test above 10
were selected (Fig. 3B). For each scan, the time-series was averaged across
blocks (Fig. 3C). To estimate the time lag between the different scans,
cross-correlation between all scans to one reference scan was computed for each
voxel. The ETR, defined as the standard deviation of the estimated time shift across
voxels, was computed separately for each echo and for the combined echoes average.
The contrast-to-noise (CNR) ratio was defined here as the maximal signal change
(see Fig. 3C) divided by the average standard deviation of the signal. Then, all
scans’ data were divided into 15 bins according to the signals’ CNR and ETR was
computed for each bin.Results
Averaged estimated time
shift for the 100ms-shifted signals was 63 ± 133 ms
and for the 600 ms-shifted signals 627 ± 143 ms
(figure 3D). The best ETR as a function of echo time corresponded to TE=43 ms,
which is proximate to the T2* of the phantom (as
expected). The estimated ETR of the combined-TEs signal was 146 ms, which is x1.7
smaller compared to the optimal single echo result (Fig. 4A). Lastly, ETR as a
function of CNR shows negative correlation (better CNR results in improved ETR;
r = -0.9, p<0.01; Fig. 4B).Discussion
In our study we suggest
a new way to investigate the effective temporal resolution of the fMRI
signal using the standard deviation of the estimated time-shifts of BOLD
mimicking signals generated using a dynamic phantom. Employing the method on
multi-echo EPI showed that signal of combined echoes provides x1.7 times better
ETR compared to single-echo. Important to note that the ETR can be better than the
repetition time (TR), since we can use the combined strength of the measured
data-points. In the scan parameters that were used here, the ETR of a combined multi-echo
dataset was 146 ms, while the TR was 600 ms. Finally, our results demonstrate
how both signal change and noise, represented in the CNR, dictate the actual
ETR. Conclusion
Our study offers both a
novel approach to examine and optimize fMRI protocols and an insight into fMRI
temporal effectivity. In the experimental setup examined here, multi-echo
EPI at 7T provided temporal resolution of <150 ms. However, further analysis
is required, incorporating additional factors that exist in a real experiment and
affect the ETR such as the physiological noise and the “resting-state”
activity. Acknowledgements
We are grateful to the Weizmann
Institute’s MRI technician team - E. Tegareh and N. Oshri - for assistance with
the dynamic phantom handling and scanning, to Dr. Amir Seginer for discussions
on the data analysis, Ghil Jona for assistance with cartridge preparation and
to ALA scientific inc. team for their extensive support. We also acknowledge the receipt of the pulse
sequences from the Center for Magnetic Resonance Research (CMRR), University of
Minnesota, USA. Dr. E.
Furman-Haran holds the Calin and Elaine Rovinescu Research Fellow Chair for
Brain Research.References
[1] Viessmann,
O., & Polimeni, J. R. (2021). High-resolution fMRI at 7 Tesla: challenges,
promises and recent developments for individual-focused fMRI studies. Current
Opinion in Behavioral Sciences, 40, 96-104. [2] Gonzalez-Castillo, J., Panwar, P., Buchanan, L.
C., Caballero-Gaudes, C., Handwerker, D. A., Jangraw, D. C., ... &
Bandettini, P. A. (2016). Evaluation of multi-echo ICA denoising for task based
fMRI studies: Block designs, rapid event-related designs, and cardiac-gated
fMRI. Neuroimage, 141, 452-468. [3] Kundu, P., Voon, V., Balchandani, P.,
Lombardo, M. V., Poser, B. A., & Bandettini, P. A. (2017). Multi-echo fMRI:
a review of applications in fMRI denoising and analysis of BOLD signals. Neuroimage, 154,
59-80. [4] Bhavsar, S., Zvyagintsev, M., &
Mathiak, K. (2014). BOLD sensitivity and SNR characteristics of parallel
imaging-accelerated single-shot multi-echo EPI for fMRI. Neuroimage, 84,
65-75. [5] Kumar, R., Tan, L., Kriegstein, A., Lithen, A., Polimeni, J. R.,
Mujica-Parodi, L. R., & Strey, H. H. (2021). Ground-truth “resting-state”
signal provides data-driven estimation and correction for scanner distortion of
fMRI time-series dynamics. NeuroImage, 227, 117584.
[6] DeDora, D. J., Nedic, S., Katti, P., Arnab,
S., Wald, L. L., Takahashi, A., ... & Mujica-Parodi, L. R. (2016). Signal
fluctuation sensitivity: An improved metric for optimizing detection of
resting-state fMRI networks. Frontiers in Neuroscience, 10,
180. [7] https://alascience.com.
[8] Moeller, S., Yacoub, E., Olman, C.A., Auerbach, E., Strupp, J., Harel, N.,
& Ugurbil, K. (2009). Multiband multislice
GE-EPI at 7 Tesla, with 16-fold acceleration using partial parallel
imaging with application to high spatial and temporal whole-brain fMRI. Magn.
Reson. Med,63,1144.[9] Feinberg, D.A., Moeller, S., Smith, S.M., Auerbach, E.,
Ramanna, S., Glasser, M.F., Miller, K.L., Ugurbil, K., & Yacoub, E. (2010).
Multiplexed echo planar imaging for sub-second whole brain fMRI and fast
diffusion imaging. Plos One, 5, e15710