Zhigang Wu1, Yajing Zhang2, Guillaume Gilbert3, Wengu Su4, Yan Zhao5, and Jiazheng Wang6
1Philips Healthcare, Shenzhen, Ltd., Shenzhen, China, 2Philips Health Technology, Suzhou, China, 3MR Clinical Science, Philips Healthcare, Mississauga, ON, Canada, 4BU MR Application, Philips Health Technology, Suzhou, China, 5BU MR R&D, Philips Health Technology, Suzhou, China, 6Philips Healthcare, Beijing, China
Synopsis
Keywords: Pulse Sequence Design, Diffusion/other diffusion imaging techniques, Diffusion, Motion compensated diffusion gradients, Reduced FOV imaging, Multi-shot DWI
Reduced
FOV imaging (rFOV) and multi-shot DWI both are very useful techniques to improve
spatial resolution for detection of pancreatic lesion. However, respiratory and
cardiovascular motion will introduce severe artifacts and ADC bias. Motion compensated
diffusion gradients (MOCO) could be used to improve the image quality. In this
study, a new sequence named MOCO-rFOV IRIS was developed, which combines the
advantages of rFOV, MOCO and image reconstruction using image-space sampling
function based multi-shot DWI (IRIS). Results from in vivo data
demonstrated that the proposed method could be used to realize motion robust and
high resolution DWI for pancreas
Introduction
Diffusion imaging has showed great
potential for the detection, staging, and treatment monitoring of pancreatic
cancer. However, conventional diffusion imaging suffers from several challenges
including low spatial resolution, distortions caused by B0 inhomogeneity, and
signal loss caused by respiratory and cardiovascular motion1. Recently, several advanced techniques have been
developed to solve these challenges. Reduced FOV imaging (rFOV)2,3 has been used to reduce distortion and
increase resolution for pancreas. Multi-shot diffusion such as multiplexed
sensitivity-encoding (MUSE)4 and image-space sampling
function based multi-shot DWI (IRIS)5,could
also be used to reduce the distortion for brain and other organ free from
physiological motions. However, severe artifacts and signal voids can be
introduced for both single -and multi-shot diffusion due to phase errors and
signal loss from bulk motion, physiological motion (breathing, intestinal
peristalsis, cardiac motion etc.), which have been investigated recently for
brain6 and liver7. Ruiqi et al. showed that cardiovascular motion could introduce
artifacts and ADC bias in pancreas DWI, and it could be addressed by motion-compensated
diffusion gradients8.
In this work, we developed a motion robust, reduced distortion and high resolution
DWI by combining rFOV, IRIS and motion-compensated gradients.Methods
Pulse sequence: We combined reduced FOV imaging with
IRIS 9, called rFOV IRIS. Fig. 1A shows the conventional rFOV IRIS
scheme which acquires data using normal pulsed gradients without motion
compensation. Fig. 1B shows our proposed scheme, it uses second order motion -diffusion
gradients to reduce the phase errors for rFOV IRIS, it’s named as MOCO-rFOV
IRIS, means the combination of motion-compensated diffusion gradients (MOCO),
rFOV and IRIS.
To evaluate performance of MOCO-rFOV IRIS, conventional DWI (ssDWI) and rFOV
without motion compensated diffusion gradients were used as reference which is
based on single shot EPI acquisition. Several other rFOV IRIS diffusion schemes
were compared with ssDWI, which were rFOV IRIS with conventional pulsed
gradients without motion compensation (M0-rFOV IRIS), rFOV IRIS with first-order
motion-diffusion gradients (M1-rFOV IRIS), and our proposal MOCO-rFOV IRIS using second-order
motion-compensated diffusion gradients.
All scanning was performed on a Philips 3.0T Elition system (Philips
Healthcare, Suzhou, China) , 24-ch torso & spine coil was used. The study
was approved by the local IRB. All Results
Compared to ssDWI and conventional
rFOV, Fig. 2 shows that M0-rFOV IRIS will show severe artifacts which are
introduced by phase errors between shots due to the motion from respiratory and
cardiovascular, which is too large and can’t be corrected by the navigator echoes.
The use of M1-rFOV IRIS with first-order motion compensated DWI reduces the
artifacts, but it still shows some residual artifacts. Our proposal MOCO-rFOV
IRIS with second-order motion-compensated diffusion gradients improves image quality dramatically with little
residual artifacts. MOCO-rFOV IRIS with
second-order motion shows smaller distortion and better spatial resolution than
other schemes.
To evaluate the performance of MOCO-rFOV IRIS further, Fig. 3 and Table 2 also
compare the ADC values from different schemes for kidney and pancreas over 5
continuous slices; it shows that ADC values by MOCO-rFOV IRIS have better
consistency between slices and smaller standard variance than observed for
other schemes, and it also provides ADC
values that are comparable to those from ssDWI and rFOV. The other schemes show
some abnormal values, which may arise from
respiratory or cardiovascular motion.For example, M0-rFOV IRIS shows a severe
bias in ADC value compared to ssDWI and to other schemes.Discussion
In this study, MOCO-rFOV IRIS was
implemented from the combination of rFOV, IRIS and motion-compensated diffusion
gradients. That approach was used to reduce the phase errors between shots,
minimize motion related artifacts and could improve image quality dramatically. Preliminary
results showed promising image quality with reduced ADC bias on kidney and
pancreas and less overall distortion. That
technique could be used for the
assessment of metastases, for which high spatial resolution, high image quality
and motion robustness is required. Further studies are required to test its performance in clinical settings,
and it could be also combined with navigator free multi-shot DWI acquisition in the future.Conclusions
We investigated the feasibility of MOCO-rFOV IRIS
which combines rFOV, IRIS and motion-compensated diffusion gradients for the
first time. Results from in vivo data demonstrated that MOCO-rFOV IRIS shows
better performance than other schemes for pancreas imaging. This technique
holds potential for clinical applications that require motion robustness, high resolution and low
distortion, such as pancreas, kidney etc.Acknowledgements
No.References
1.Diego H, et al. Quantitative
diffusion MRI of the abdomen and pelvis. Med Phys. 2022;49(4):2774-2793.
2.
Masahiro T, et al Reduced
Field-of-View Diffusion-Weighted Magnetic Resonance Imaging of the Pancreas
With Tilted Excitation Plane: A Preliminary Study. J Magn Reson Imaging.
2021;54(3):715-720.
3. Hyungjin K, et al. Reduced
Field-of-View Diffusion-Weighted Magnetic Resonance Imaging of the Pancreas:
Comparison with Conventional Single-Shot Echo-Planar Imaging. Korean J Radiol.
2015;16(6):1216-25.
4.
Nan-Kuei C, et al. A robust
multi-shot scan strategy for high-resolution diffusion weighted MRI enabled by
multiplexed sensitivity-encoding (MUSE). Neuroimage. 2013 May 15;72:41-7.
5. Ha-Kyu J, et al. High-resolution
human diffusion tensor imaging using 2-D navigated multishot SENSE EPI at 7T.
Magn Reson Med. 2013;69(3):793-802.
6. Eric SM, et al. Multi-shot
diffusion MRI of the human brain with motion-compensated oscillating gradients.
Proc. Intl. Soc. Mag. Reson. Med. 29 (2021) 1324.
7.
Yuxin Z, et al. Motion‐robust and blood‐suppressed M1‐optimized diffusion MR imaging of the
liver. Magn Reson Med. 2019 Jul;82(1):302-311.
8. Ruiqi G, et al. Characterization
and correction of cardiovascular motion artifacts in diffusion-weighted imaging
of the pancreas. Magn Reson Med. 2021;86(4):1956-1969.
9.
Wenxing F, et al.
High-Resolution DWI using reduced FOV multi-shot EPI with IRIS reconstruction.
Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)3052.