XueHua Peng1, LiFei Ma2, Rui Zhang2, and XiaoMeng Wu2
1Radiology Department, WuHan Children’s Hosptial, WuHan, China, 2philips, shanghai, China
Synopsis
Keywords: Parkinson's Disease, Quantitative Susceptibility mapping, Rapid QSM measurement, EPI acquisition, Neuro system
Quantitative susceptibility mapping(QSM) can be used for
quantitatively measurement of neuro system magnetic susceptibility information
caused by various diseases, such as Alzheimer's disease (AD), Parkinson's
disease (PD) and neuro vascular diseases. In this study, a rapid QSM acquisition
based on EPI sequence was proposed to improve clinical feasibility of QSM. The
results demonstrated the rapid QSM were very close to those obtained by standard
QSM.
INTRODUCTION
QSM can be a sensitive method to visualize the Iron
metabolism disorders which played a role in the neuronal apoptosis of
neurodegenerative diseases, such as Alzheimer's disease (AD), Parkinson's
disease (PD), etc. [1]. It also can monitor the patients with hemorrhagic or ischemic
stroke, cerebrovascular malformation similar with SWI [2]. This study intends
to propose a rapid QSM measurement based on EPI sequence to improve the
clinical applicability of QSM quantitative examination.METOHDS
Study sample: Fifteen subjects (Nine males. Age
range from 24 to 48 years old) took part into this study with approval by local
institutional review board, a written consent form was obtained from each
subject.
MR imaging: MR imaging was acquired on a 3.0 T MR
scanner (Ingenia, Philips Hearthcare). The standard QSM sequence adopted 10-echo
fast gradient echo sequence, TR 41.0ms, TE 2.8ms, slice thickness 1.4mm with
0mm gap, FOV 240 × 240 mm2, scanning time 4:53 min; EPI-QSM sequence
adopted EPI sequence, TR 3000ms, TE 35ms, slice thickness 2mm with 0mm gap, FOV
240 × 240 mm2 ,two dummy phases and 5 dynamic phases were collected,
scanning time 21 seconds.
DATA analysis: The EPI-QSM data were arithmetically
averaged across phases. Two groups of data did spatial registration to align space
position. Then, the QSM quantities calculated by STI-SUITE (505 Cory Hall, MC #
1770, Berkeley, CA) after phase unwrapping, background field removal, magnetic
susceptibility inversion.RESULTS
EPI-QSM and standard QSM sequences obtained similar parameter images (see Figure 1).
Twenty-one regions of interest (ROI)
were selected within whole brain showing in Figure 2.
The whole brain QSM results listed in
Table 1. Statistics showed QSM quantities measured by two sequences were
different(t=-2.538, P=0.012)and yet significantly
correlated(R2=0.615, P=0.00) as shown in Figure 3.DISCUSSION
The rapid EPI-QSM significantly shortens
scanning time completing acquisition within 21 seconds, with which could
improve the patient's comfort and success rate of scanning.
The rapid QSM value emerge good
consistency with the standard QSM within whole cerebral cortex. It is worth
noting that EPI-QSM was easy to produce deformation in inferior frontal and
temporal gyrus, but QSM quantification are not sensitive to geometric
deformation [Li J et. al MRM 2012]. Therefore, in
practical application, combining with anatomical images, QSM can accurately
locate the lesions during diagnosis.CONCLUSION
Rapid EPI-QSM acquisition shorten the scanning
time of QSM quantitative collection. QSM quantification is significantly
consistent with the value of standard QSM sequence, so it can be used for
quantitative measurement of magnetic susceptibility in various diseases in the neuro
system.Acknowledgements
noneReferences
1.Ke
Y, Ming Qian Z. Iron misregulation in the brain: a primary cause of
neurodegenerative disorders[J]. Lancet
Neurol, 2003, 2(4): 246-53.
2.Schneider
TM, Nagel AM, Zorn M, et al. Quantitative susceptibility
mapping and (23) Na imaging-based in vitro characterization of blood clotting
kinetics[J]. NMR Biomed, 2018, 31(6): e3926.
3.Li J, Chang S, Liu T, Wang Q, et
al. Reducing the object
orientation dependence of susceptibility effects in gradient echo MRI through
quantitative susceptibility mapping. MAGNET RESON MED. 2012,11-01;68(5):1563-9.