Qing FU1, Jie Liu1, Ding-xi Liu1, Ting Yin2, Peng Sun3, Zi-qiao Lei1, and Fan Yang1
1Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China, 2MR Collaborations, Siemens Healthineers Ltd., Chengdu, China, 3Philips Healthcare, Beijing, China
Synopsis
Keywords: Breast, Breast, MRI; Dixon; fat suppression; 3T
For patients with breast lesion resection, conventional contrast enhanced T1WI images with SPAIR provides poor fat suppression in the post-mastectomy breasts side due to the uneven skin surface, inhomogeneous tissues environment, and the frequency-selective feature of SPAIR scheme. Our study aimed to investigate the performance of fat suppression with the Dixon method in single side post-mastectomy patients. Subjective and objective evaluation suggested the Dixon method is superior to conventional SPAIR fat suppression in depicting the bilateral breast. Dixon method provided better image uniformity and higher fat suppression efficiency, and showed significant advantages in delineating the anatomical structures, with better axillary and lesion visibilities, especially in the completely removed breast side.
Introduction/Purpose:
Contrast-enhanced T1WI (CE-T1WI) is
widely used in breast MRI with high sensitivity in breast lesion detection, and
high-quality fat suppression is essential for delineating of the enhanced lesions
[1-2]. Conventionally, frequency-selective
fat suppression with spectral attenuated inversion recovery (SPAIR) is
routinely applied in clinical practice, but it suffers
from inhomogeneous fat suppression, especially in the post-mastectomy
breast side. This is related to uneven skin
surface with contracted scars, field inhomogeneity in the broad contact
area between the skin surface and surrounding air, SPAIR relies on
frequency-selective excitation with lipid protons. Thus, poor fat suppression with
SPAIR in post-mastectomy breast side makes it difficult for radiologists to
give a definitive evaluation. Dixon method could achieve better fat suppression
by reconstruction of in- and out-phase images which are less sensitive to B0
inhomogeneities [3]. Therefore, current study aims to investigate image quality and diagnostic performance of breast CE-T1WI with Dixon in
post-mastectomy patients compared with SPAIR method.Materials and methods
A total of 47 patients underwent mastectomy and post-operate
breast CE-MRI were enrolled in this study. Both CE-T1WI with SPAIR and Dixon
were scanned using a 3T MR scanner
(Magnetom Skyra, Siemens Healthcare, Germany) with an 18-channel
breast coil. Acquisition time of the two CE-T1WI with different fat-suppression
methods were adjusted to be as similar as possible with the following
parameters: SPAIR method: TR/TE=4.04/1.52ms, matrix size= 403x448, voxel
size=0.8x0.8x0.8mm, TA=2min13s. Dixon method: TR/TE=5.52/2.46, 3.69ms, matrix
size= 374x416, voxel size=0.87x0.87x0.90mm, TA=2min16s.
Subjective evaluation was done by two
independent experienced radiologists using a 5-point scale in consensus [4], with
regards to fat-suppression efficiency and image uniformity, the visibility of
anatomical structures and axillary nodes, lesion conspicuity, and severity of
artifacts in bilateral breast sides (Table 1). For objective evaluation, CNR between enhanced lesions,
ectopectoralis, and fat were calculated. The edge sharpness of the skin surface
was measured in ImageJ software, which was defined as the mean slope of the
signal intensity changes from post-operate skin surface to the subcutaneous
fat. The bigger the value represents a greater edge sharpness. The uniformity
in post-operative breast side was measured using 3D Slicer software and the
Slicer-Radiomics extension (uniformity defined by IBSI as intensity
histogram uniformity). The related ROIs were drawn to cover the left /right
top quarter post-operative breast region and axillary region after the Dixon
and SPAIR images were resampled to the same resolution and co-registered. The
smaller the uniformity value represents the better fat suppression with
good tissue/fat contrast. The uniformity
values of the Dixon and SPAIR images were recorded for further comparison.
Paired t-test and Wilcoxon rank sum test were used for statistical analyses.
Results:
Subjective evaluation
As shown in Table 2, the Dixon method provided
superior fat than the SPAIR method with significantly better subjective scores
in regards to signal uniformity and degree of fat suppression (p<0.05) in a
wide FOV range, including both in the post-operative & non-operative breasts
and bilateral axillary areas (Fig 1-3). As for the anatomical structures’ definition, Dixon performed significantly better than
SPAIR in the post-operative breast side with p<0.001, while there was no
statistical difference for the non-operative breast side with p=0.699.
Although artifacts caused by cardiac impulse in
SPAIR (1.66±0.87) were significantly
less prominent than in Dixon (1.98±0.61) in the
post-operative breast side, and in Dixon (1.36±0.49)
was significantly better than in SPAIR (1.43±0.54) in
the non-operative breast side with p <0.05, the scores were ranged from
score 1 to 3, which did not deteriorate diagnostic confidence. As for axillary
visibility, Dixon outperformed SPAIR in evaluation for both post- and
non-operative breast sides with p<0.001 (Table 2).
For assessing the lesion conspicuity, Dixon
performed similarly with SPAIR in the non-operative breast side with p=0.083 (Table
2), As for the post-operative breast side, there were 6 patients with
abnormalities or enhancing lesions, statistical analysis was not performed due
to the small sample size, the subjective scores for SPAIR and Dixon were listed
as follows: 1,2,4,2,2,2 and 1,1,1,1,1,1, respectively. The incompletely and
partially unsuppressed signals in the post-operative regions severely hindered
the accurate depiction of anatomical structures and lesions (Fig 1-2),
thus indicating the superior performance of the Dixon method.
Objective evaluation
As shown in Table 2, CNRs between fat, ectopectoralis, and lesions were significantly
higher in images obtained by Dixon than by SPAIR (p<0.001), indicating
better image contrast for evaluation. The edge sharpness between the skin surface and subcutaneous fat
were significantly higher in Dixon than in SPAIR method (p<0.001). The signal
uniformity of the post-operative breast side was seen with better fat suppression in Dixon than in SPAIR with p<0.001.
Additional findings
In 7 patients, the separation of water-only and
fat-only images was incorrectly reconstructed, so nipples at one side or both
two sides demonstrated fat-only signal in water-only images, and water-only
signal in fat-only images, which would not interfere with the image interpretations
of Dixon images.
Conclusions:
Dixon is superior to SPAIR
with better image homogeneity of for fat suppression, and
higher CNR for lesion depiction and image quality than SPAIR in bilateral
breasts and axillary areas, especially in the postoperative breast through
qualitative and quantitative assessment. Dixon is more advised to replace SPAIR
in breast CE-T1WI for post-mastectomy patients.Acknowledgements
No acknowledgement found.References
1.Kuhl C. The current
status of breast MR imaging. Part I. Choice of technique, image interpretation,
diagnostic accuracy, and transfer to
clinical practice. Radiology. 2007;244:356–78.
2.Niitsu M, Tohno E, Itai
Y. Fat suppression strategies in enhanced MR imaging of the breast: comparison
of SPIR and water excitation sequences. J Magn Reson Imaging. 2003;18:310–4
3.Glover GH (1991)
Multipoint Dixon technique for water and fat proton and susceptibility imaging.
J Magn Reson Imaging 1:521–530
4. Kalovidouri, A.,
Firmenich, N., Delattre, B.M.A. et al. Fat suppression techniques for
breast MRI: Dixon versus spectral fat saturation for 3D T1-weighted at
3 T. Radiol med. 2017;122, 731–742.