Nima Gilani1, Artem Mikheev1, Inge Manuela Brinkmann2, Dibash Basukala1, Thomas Benkert3, Malika Kumbella1, James S. Babb1, Hersh Chandarana1, and Eric E. Sigmund1
1Department of Radiology, NYU Langone Health, New York, NY, United States, 2Siemens Medical Solutions USA Inc., New York, NY, United States, 3Siemens Healthcare GmbH, Erlangen, Germany
Synopsis
Keywords: Artifacts, Kidney, TOPUP, Field Inhomogeneity
Echo planar imaging is highly affected by field map inhomogeneity distortion artifact. Field map inhomogeneity has shown to be motion dependent in the kidneys. In the present work, we propose
an alternative method for correction of magnetic field inhomogeneity for renal
DWI in respiratory-resolved fashion. Specifically,
we collect a series of forward and reverse phase encoded b=0 images to sample kidney motion caused
by breathing, map the spatial and respiratory phase dependence of the
magnetic field inhomogeneity, and correct each image of free-breathing DWI series according to their respiratory phase.
Introduction
Echo planar imaging, a
common readout for diffusion weighted imaging (DWI), is sensitive to eddy currents, static magnetic field (B0) inhomogeneity,
and respiratory motion 1. Solutions include using bipolar gradients, acquiring reverse
and forward phase encodings for distortion correction 2,
and prospective triggering or retrospective registration.
Breathing causes craniocaudal kidney movements of 10-16 mm 3,4 and modulates field inhomogeneity (FI). Several
studies have demonstrated motion dependence of FI and corrected it retrospectively 5 or prospectively 6-9. Coll-Font et al. 6 performed correction of renal DWI distortion
using forward and reverse images of the same directional diffusion encoding acquired
with multiple echoes allowing for correction of each image at each breathing
phase separately. In the present work, we propose
an alternative method for correction of FI for renal
DWI in respiratory-resolved fashion. Specifically,
we employ 32 forward and 32 reversed b=0 images to sample kidney motion, and map the spatial and respiratory phase dependence of the
B0 inhomogeneity. These B0 maps are then applied to correct the DWI series according to their respiratory phase.Methods
In this HIPAA-compliant and
IRB-approved prospective study, 8 volunteers (6M, ages 28-51) provided written
informed consent and had abdominal imaging performed in a 3 T MRI system
(MAGNETOM Prisma; Siemens Healthcare, Erlangen, Germany) in supine position
with posterior spine array and anterior body array RF coils and chest leads for
ECG gating. Coronal oblique T2-weighted SSFSE (HASTE) images were collected for
anatomical reference. Sagittal phase-contrast (PC) MRI images through the left
renal artery were collected at multiple cardiac phases to estimate systolic and diastolic phases for kidney. With a
DWI research application sequence with dynamic field
correction, cardiac triggered oblique coronal DWI (TR/TE 2800/81 ms, matrix
192/192/1, resolution 2.2/2.2/5 mm) were collected at 10 b-values between 0-800
s/mm2 and 12 directions. To correct for FI, 32
right-to-left and 32 left-to-right phase-encoding b=0 images were acquired,
half before and half after the DWI acquisition to account for bulk
motion. Figure 1 is a pictorial
description of our method. The 64 b=0 images were
registered using mutual information (MI) metrics to one of the forward images in the
set taken as reference using the FireVoxel software, build 380, https://firevoxel.org/. The craniocaudal
translation of this rigid transformation was logged to create a limited number
of discrete motion bins, each containing at least five forward and reverse
images. Each bin was corrected using the TOPUP function of FSL software
(version, 6.0.1) and its FI map was outputted. Each image
of the DTI set was registered to the previous reference image, and distortion
corrected with the appropriate B0 map for its location.
The corrected and uncorrected images
were registered to their corresponding HASTE image, and their MI metrics compared. Finally, a
line profile analysis evaluated the variability of the apical region over the
full DTI acquisition with and without motion-resolved distortion correction. The dependence of field map
inhomogeneity on the kidney side (i.e. right and left), region (i.e. lower,
middle, upper, and apical), and breathing phase was statistically analyzed
using mixed model analysis of variance. Tests for
significant variation with each factor in this model were quantified by an F
statistic and p-value. Mixed model analysis of variance was used to
assess the effect of each DWI, side and method (N (No correction), O (one bin
correction), and M (multi-bin correction)) on MI. MI metrics from different methods (N,O,M)
were compared via t-statistics of pairwise comparisons for all acquired images. For the line profile analysis, all line
profile standard deviations for a given subject, kidney, and correction method
(O or M) were averaged and compared with a paired sample t-test between
correction methods. Results
After exclusion of one subject due to low image quality, seven subjects contributed to
FI derivations for the lower, middle and upper regions and five subjects
contributed to FI derivations for the apical layer, registration comparison and line
profile analysis.
Figure 2 shows motion-dependent FI maps of one subject. FI was found to vary significantly (p<0.001) with each factor considered (Region : F = 294.36; Side: F = 61.32; Phase: F = 4.5). Figures 3, 4, and 5 show the group dependence of field inhomogeneity on kidney side, region, and breathing phase, respectively.
MI was found to vary significantly (p<0.001)
with consideration of factors kidney side: F = 374.9 and correction method:
F=
574.4. MI was significantly lower for N than for each of the corrected methods on each side and over
both sides combined (p<0.001). Additionally, MI was significantly lower for O than M in the left kidney (p<0.001), whereas, MI difference for the two methods was not significant in the right kidney (p=0.86). In the line profile analysis, multi-binning resulted in significantly
better registration (p=0.006).Discussion
Improvements in morphologic accuracy can be obtained with
motion-resolved distortion correction in the kidney. The presented method utilizes 64
reverse and forward images acquired
in 230 seconds, which may be beneficial for minimizing scan time. The
results of this study also indicate significant dependence of field
inhomogeneity on kidney side, region, and breathing phase. The information in this
spatiotemporal mapping might facilitate prospective B0 shimming on 7T scanners, similar to earlier works 5.Acknowledgements
Funding support is acknowledged from the National Institute of Health (NIH).References
1. Pierpaoli,
C., Artifacts in diffusion MRI.
Diffusion MRI: theory, methods and applications, 2010: p. 303-318.
2. Smith, S.M., et al., Advances in functional and structural MR
image analysis and implementation as FSL. Neuroimage, 2004. 23 Suppl 1: p. S208-19.
3. Hallman, J.L., et al., A four-dimensional computed tomography
analysis of multiorgan abdominal motion. Int J Radiat Oncol Biol Phys,
2012. 83(1): p. 435-41.
4. Bussels, B., et al., Respiration-induced movement of the upper
abdominal organs: a pitfall for the three-dimensional conformal radiation
treatment of pancreatic cancer. Radiother Oncol, 2003. 68(1): p. 69-74.
5. Vannesjo, S.J., et al., Spatiotemporal characterization of
breathing-induced B0 field fluctuations in the cervical spinal cord at 7T.
Neuroimage, 2018. 167: p. 191-202.
6. Coll-Font, J., et al., Retrospective Distortion and Motion
Correction for Free-Breathing DW-MRI of the Kidneys Using Dual-Echo EPI and
Slice-to-Volume Registration. J Magn Reson Imaging, 2021. 53(5): p. 1432-1443.
7. Afacan, O., et al. Dual-echo blip reversed EPI acquisition
enables distortion correction in the presence of motion in diffusion-weighted
MRI. in Int Soc Magn Reson Med.
2019.
8. Afacan, O., et al., Simultaneous Motion and Distortion
Correction Using Dual-Echo Diffusion-Weighted MRI. J Neuroimaging, 2020. 30(3): p. 276-285.
9. Gallichan, D., et al., Reducing distortions in diffusion-weighted
echo planar imaging with a dual-echo blip-reversed sequence. Magn Reson
Med, 2010. 64(2): p. 382-90.