Julian Rauch1,2, Frederik B. Laun3, Dominik Ludwig1, Maxim Zaitsev4, Mark E. Ladd1,2,5, Peter Bachert1,2, and Tristan A. Kuder1,2
1Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 3University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 4Medical Physics, Department of Radiology, Faculty of Medicine, Medical Center University of Freiburg, Freiburg, Germany, 5Faculty of Medicine, Heidelberg University, Heidelberg, Germany
Synopsis
Keywords: Artifacts, Artifacts
Intravoxel dephasing generated by Maxwell or
concomitant fields can cause image artifacts like signal voids or falsify the
quantification. In this study, an optimization scheme to reduce concomitant
field effects in diffusion sequences with single pairs of bipolar gradients on
each axis is presented. Oscillating gradients are added onto the original
gradient pulses with the aim of reducing the concomitant phase without
significant changes in the sequence properties. The proposed method is
evaluated in both measurements and simulations, and gives rise to a positive
effect on the signal for arbitrary diffusion wave vector pairs.
Introduction
Double diffusion encoding (DDE) schemes with two pairs
of bipolar gradients separated by a refocusing pulse are prone to severe
artifacts induced by so-called Maxwell or concomitant fields. They manifest themselves as additional non-linear phase contributions on transverse
magnetization1,2 that often corrupt measurements of diffusion metrics such as microscopic anisotropy3,4.
In this study, we demonstrate in both simulations and
measurements the positive effect of superimposing optimized oscillating
gradients onto the original waveforms in the above-mentioned DDE sequences to
reduce the concomitant phase without changing the sequence parameters
significantly.
In
particular, local gradient coils with asymmetric windings optimized to provide
very high amplitudes are expected to generate significant concomitant fields, while the
oscillations are easier to implement using such coils5,6 due to higher available slew
rates and higher peripheral nerve stimulation (PNS) thresholds.Methods
Concomitant fields to the lowest order are given by $$\mathrm{B_C(x,y,z,t)}=\frac{1}{2\mathrm{B_0}}(\mathrm{G_x^2z^2+G_y^2z^2+G_z^2\frac{x^2+y^2}{4}-G_xG_zxz-G_yG_zyz}),\quad[1]$$ where the gradients $$$\mathrm{G_x}$$$,
$$$\mathrm{G_y}$$$ and $$$\mathrm{G_z}$$$ alter the static magnetic field $$$\mathrm{\vec{B}_0=B_0\hat{e}_z}$$$ along the
physical scanner axes $$$\mathrm{x}$$$, $$$\mathrm{y}$$$ and $$$\mathrm{z}$$$,
respectively.
The concomitant phase is then given by
$$\mathrm{\phi_C(x,y,z)=\gamma\int{B_C(x,y,z,t)}\,dt,\quad[2]}$$ with $$$\gamma$$$
being the gyromagnetic ratio.
The absolute remaining signal in a voxel due to the
concomitant phase-induced signal loss can be numerically approximated using
$$\mathrm{S_C(x,y,z)=\left|\frac{\sum_i^{N} \sum_j^{M} \sum_k^{K}e^{i \phi_C(x_i,y_j,z_k)}}{N\cdot{M}\cdot{K}}\right|},\quad[3]$$ where $$$\mathrm{N,\,M}$$$ and $$$\mathrm{K}$$$
denote the number
of considered sub-voxels along each dimension, respectively.
The concomitant phase-reducing waveforms were designed
using the "patternsearch" algorithm in MATLAB (The Mathworks, Inc, MA).
In the optimization process, for each of the 60 pairs
of diffusion wave vectors published by Jespersen et al.4, oscillations were
added onto the original gradient pulses that apply a b-value of 2000 s/mm2
along every direction (Figure 1).
The sequences
were optimized such that the signal loss based on Equation 3 in a transversal slice
shifted by $$$\mathrm{{\Delta}z}=$$$$$$\,$$$-50$$$\,$$$mm from the isocenter with a voxel size of 3$$$\,$$$$$$\times$$$$$$\,$$$3$$$\,$$$$$$\times$$$$$$\,$$$3$$$\,$$$mm3 at $$$\mathrm{FOV=}$$$$$$\,$$$300$$$\,$$$mm$$$\times$$$300$$$\,$$$mm was minimal. $$$\mathrm{B_0}$$$ was set to 2.89 T.
These parameters were used in all further experiments in this study.
Respective
constraints kept a maximum gradient amplitude of 75 mT/m, the duration of the
original waveforms, and a maximum deviation of $$$\mathrm{q=\gamma\int{G}\,dt\leq}$$$$$$\,$$$50$$$\,$$$m-1 for each modified gradient pulse as well as a total
b-value deviation of 6% of the total weighting. A maximum slew rate of 188.6 mT/m/ms was allowed.
The measurements were carried out on a SIEMENS
Prisma 3T imaging system using a phantom with an aqueous
solution of polyvinylpyrrolidone (PVP) K30 43% (w/w) with $$$\mathrm{ADC\approx}$$$$$$\,$$$0.5$$$\,$$$μm2/ms.
100 repetitions of the
optimized and original waveforms for each diffusion wave vector pair were
acquired using an EPI readout ($$$\mathrm{TE}=$$$$$$\,$$$146$$$\,$$$ms, $$$\mathrm{TR}=$$$$$$\,$$$4000$$$\,$$$ms, $$$\mathrm{rBW}=$$$$$$\,$$$2000$$$\,$$$Hz/pix) with an 18-channel body coil. PNS
monitoring was disabled.
The SNR in every pixel in a
ROI (Figure 2) was calculated for each sequence and corrected for the specific
diffusion-induced signal loss. The relative signal gain for every diffusion
wave vector pair was obtained by pixelwise division of the SNR values for the
optimized and the original waveforms.
For the simulations, the remaining absolute signals
for each sequence were determined using Equation 3 with $$$\mathrm{N=M=K=}$$$$$$\,$$$101 and evaluated analogously to the measurements.Results
A comparison of the simulated and measured relative SNR
gains in the considered ROI shows a high agreement for an example wave vector
pair (Figure 2).
The calculation of the relative SNR gain for the
measurements (black in Figure 3) reveals a positive measurable effect of the
proposed optimization method of up to 40% for all considered pairs of diffusion
directions for the parameters used, and thereby confirms the positive prediction
of the simulations (red).
While the simulations predict a higher gain for the first
diffusion wave vector pairs, measurements and simulations coincide relatively
well for the other pairs. For these following pairs, the measurements tend to
exhibit a larger gain than the simulations.Discussion and Conclusion
As demonstrated, the addition of oscillations onto the
original gradients of arbitrary pairs of encoding wave vectors enables
concomitant field reduction while allowing for simultaneous diffusion weighting
without changing the sequence characteristics significantly.
The proposed optimization offers additional degrees of freedom that allow for
adequate shifting of the oscillation lobes to also reduce the two cross terms
in Equation 1. We
constrained the optimization in our study to avoid time extension of the optimized waveforms, and limited
the maximum gradient amplitude to 75 mT/m, which makes the method theoretically
feasible on commercially available gradient systems. The current limitation due to PNS is expected to be less relevant for
local gradient coils5,6 than
for whole-body systems.
The strong oscillations applied in the optimized
waveforms for the first wave vector pairs might have caused vibrations in the
phantom that led to the relatively large deviations between the measurements
and the simulations (Figure 2). The large number of repetitions to accurately determine
the SNR led to a total acquisition time of around 14 hours, such that
temperature changes might have impacted the results.
Limitations regarding the simulations include the number of sub-voxels used to determine the remaining
signal. Furthermore, the actual slice profile in the
measurements might be different from the assumed uniform profile in the
simulations.
In summary, the proposed method leads to signal gains
by reducing the concomitant phase without additional RF refocusing pulses.Acknowledgements
No acknowledgement found.References
1. Bernstein MA et al. Concomitant gradient terms in
phase contrast MR: analysis and correction. Magn Reson Med. 1998;39(2):300-8
2. Zhou XJ, Tan SG, Bernstein MA. Artifacts Induced by
Concomitant Magnetic Field in Fast Spin Echo Imaging. Magn Reson Med. 1998;40(4):582-91
3. Lawrenz M, Finsterbusch J. Double-wave-vector diffusion-weighted imaging
reveals microscopic diffusion anisotropy in the living human brain. Magn Reson
Med. 2013;69(4):1072-82
4. Jespersen SN et al. Orientationally
invariant metrics of apparent compartment eccentricity from double pulsed field
gradient diffusion experiments. NMR Biomed. 2013;26(12):1647-62
5. Weiger M, Overweg J, Rösler MB,
Froidevaux R, Hennel F, Wilm BJ, Penn A, Sturzenegger U, Schuth W, Mathlener M,
Borgo M, Börnert P, Leussler C, Luechinger R, Dietrich BE, Reber J, Brunner DO,
Schmid T, Vionnet L, Pruessmann KP. A high-performance gradient insert for
rapid and short-T2 imaging at full duty cycle. Magn Reson Med.
2018;79(6):3256-3266
6. Jia F, Littin S, Amrein P, Yu H, Magill AW, Kuder TA, Bickelhaupt S, Laun
F, Ladd ME, Zaitsev M. Design of a high-performance non-linear gradient coil
for diffusion weighted MRI of the breast, J. Magn. Reson. 2021;331:107052