Petra J van Houdt1, Abel Bregman1, Kari Tanderup2, Robert Hudej3, Marko Zaletelj3, Barbara Ĺ egedin3,4, Eva E Schaake1, Ellen M Kerkhof5, Laura A Velema5, Remi A Nout6, and Uulke A van der Heide1
1Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, Netherlands, 2Clinical Medicine, Aarhus University Hospital, Aarhus, Denmark, 3Radiotherapy, Institute of Oncology Ljubljana, Ljubljana, Slovenia, 4Faculty of Medicine, University of Ljubljana, Ljubljana, Slovenia, 5Radiotherapy, Leiden University Medical Center, Leiden, Netherlands, 6Radiotherapy, Erasmus MC Cancer Institute, University Medical Center Rotterdam, Rotterdam, Netherlands
Synopsis
Keywords: Cancer, Quantitative Imaging, cervical cancer
The reproducibility of quantitative MRI parameters in a
multicenter setting is not trivial. In this study we evaluated the in-vivo
multicenter consistency of quantitative MRI parameters in cervical cancer
patients. DWI, T2 mapping and T1 mapping were available from 66 patients from 3
institutes. ADC values were measured consistently across institutes despite
differences in acquisition protocols. For T2 and T1 values differences between
institutes were observed, which need to be accounted in future evaluation of
the quantitative MRI parameters in relation to treatment outcome.
Introduction
Quantitative
MRI parameters derived from for example diffusion weighted imaging (DWI) and
dynamic contrast-enhanced (DCE) MRI have shown potential to predict treatment
outcome in cervical cancer1,2. The next step in the validation of
such biomarkers is to test them in a larger multicenter setting. However, the
reproducibility of quantitative MRI parameters in a multicenter setting is not
trivial. Phantom QA was performed before the start of the IQ-EMBRACE trial to
minimize the differences between institutional acquisition protocols3.
The aim of this study was to evaluate the in-vivo multicenter consistency of
quantitative MRI parameters in cervical cancer patients.Methods
93
patients were included in IQ-EMBRACE in six institutes (clinicaltrials.gov NCT03210428). The study was approved by the institutional research boards and all patients sigend informed consent for participation in the study. For reproducibility assessment the data from
the institutes with more than ten patients were included (n = 85, 3
institutes). A pre-treatment MRI exam was performed including a transversal
T2-weighted image, diffusion weighted imaging, T2 mapping, T1 mapping with variable
flip angle approach (Table 1). An ADC map was generated with a log-linear fit
using b-values above 200 s/mm2. T2 maps were generated at the scanner with a
mono-exponential fit for institute 2 and 3. For institute 1 these were not
available and generated offline with a log-linear fit to the signal intensities
of all echo times. For comparison, additional T2 maps were generated for
institute 2 and 3 with the same software and the same number of echoes. T1 maps
were derived using linear-least squares implementation. For the measurements
done at 3T, the T1 map was calculated with and without B1 correction. Tumors
were delineated in each center on the T2-weighted MRI according to the EMBRACE
guidelines. Median values for the tumor were determined per patient and
compared on institutional-level with a non-parametric Kruskal-Wallis test (p
< 0.05).Results
From the 85 patients, 5 patients were excluded due to
technical reasons (wrong protocol or issue with data import), whereas for 14
patients the tumor delineations were not available yet. This resulted in a
total of 66 data sets analysed (n = 36 for institute 1, n = 15 for institute 2,
n = 15 for institute 3). Table 2 shows the patient characteristics per
institute. Furthermore, in 2 cases a T2 map could not be calculated due to
incomplete data, and in 3 cases the ADC map could not be used due to
geometrical distortions.
Example images for each institute are shown in Fig. 1.
Median ADC values between institutes were not significantly different (p=0.19,
Fig. 2). The median T2 values of institute 1 were significantly higher compared
to the T2 values of institute 2 and 3 (p < 0.001). The T2 values fitted with
a mono-exponential fit were similar to the T2 values estimated with a log-linear
fit. The median T1 values were significantly lower in institute 1 compared to
the T1 values without B1 correction of institute 2 and 3 (p < 0.001). The T1
values decreased with B1 correction for both institutes.Discussion
ADC values were measured consistently across institutes
despite differences in acquisition protocols. The differences in T2 values may
be attributed to the different acquisition methods used, where in institute 1 a
series of single echo spin echo acquisitions was acquired, whereas in institute
2 and 3 multi-echo spin echo acquisitions were used. Higher T2 values were also
observed with the phantom measurements for institute 1, although the difference
was not as large as in the in-vivo data (see Fig. 3a). For T1, a difference in
T1 values was observed with previous phantom measurements as well (see Fig.
3b). In addition, the in-vivo differences can also be explained by the
difference in field strength, as T1 values are typically 30% higher at 3 T
compared to 1.5 T.
In-vivo reproducibility assessment requires the same subject
to be scanned on different systems, which is logistically challenging. In this
study we tried to evaluate multicenter consistency by comparing the median
values of tumor tissue between institutes. However, tumors differ across
patients. An alternative might be to investigate a representative healthy
reference region which is more comparable between patients, for example the
healthy myometrium or a muscle. However, the menstrual cycle of a patient can
influence the quantitative values in the myometrium4. Muscles have
lower signal intensity on T2-weighted images and DWI compared to tumor tissue,
resulting in a lower signal-to-noise ratio. Therefore, the differences found
with these reference structures might not be representative for differences
between institutes in the range of values of tumor tissue.Conclusion
In this study, the consistency of quantitative MRI
parameters in a multi-center setting was shown. ADC values were measured consistently
across institutes, whereas differences were found between institutes for T2 and
T1 mapping. These differences need to be accounted in future evaluation of the
quantitative MRI parameters in relation to treatment outcome.Acknowledgements
This work was funded by a research grant from
Varian Medical Systems.References
1.
Halle C, Andersen E, Lando M, Aarnes EK, Hasvold G, Holden M,
Syljuåsen RG, Sundfør K, Kristensen GB, Holm R, Malinen E, Lyng H.
Hypoxia-induced gene expression in chemoradioresistant cervical cancer revealed
by dynamic contrast-enhanced MRI. Cancer Res. 2012 Oct 15;72(20):5285-95. doi:
10.1158/0008-5472.
2.
Skipar K, Hompland T, Lund KV, Løndalen A, Malinen E,
Kristensen GB, Lindemann K, Nakken ES, Bruheim K, Lyng H. Risk of recurrence
after chemoradiotherapy identified by multimodal MRI and 18F-FDG-PET/CT in
locally advanced cervical cancer. Radiother Oncol. 2022 Sep 14;176:17-24. doi:
10.1016/j.radonc.2022.09.002.
3.
van Houdt PJ, Kallehauge JF, Tanderup K, Nout R, Zaletelj M,
Tadic T, van Kesteren ZJ, van den Berg CAT, Georg D, Côté JC, Levesque IR,
Swamidas J, Malinen E, Telliskivi S, Brynolfsson P, Mahmood F, van der Heide
UA; EMBRACE Collaborative Group. Phantom-based quality assurance for
multicenter quantitative MRI in locally advanced cervical cancer. Radiother
Oncol. 2020 Dec;153:114-121. doi: 10.1016/j.radonc.2020.09.013.
4.
Kido A, Kataoka
M, Koyama T, Yamamoto A, Saga T, Togashi K. Changes in apparent diffusion
coefficients in the normal uterus during different phases of the menstrual
cycle. Br J Radiol. 2010 Jun;83(990):524-8. doi: 10.1259/bjr/11056533. PMID:
20505034; PMCID: PMC3473588.