Yulia Shcherbakova1, Tijl A van der Velden1,2, and Peter R Seevinck1,2
1Imaging Division, UMC Utrecht, Utrecht, Netherlands, 2MRIguidance B.V., Utrecht, Netherlands
Synopsis
Keywords: Image Reconstruction, Skeletal, AI, acceleration, CS-SENSE, spine, sCT, bone
In
this work, we investigated the performance of an AI based algorithm - synthetic
CT generation - when subjected to compressed sensing-sensitivity encoding
(CS-SENSE) accelerated gradient echo images. We performed MR experiments in
five volunteers, using different
CS-SENSE acceleration factors for the MR acquisitions. Our results showed that
using CS-SENSE factors of 1.45 and 2 increased noise in the MR source images
but did not compromise the sCT reconstruction on visual inspection, which was
confirmed by quantitative metrics. However, CS-SENSE with a factor of 3 caused
artifacts in the sCT images which may affect the safety and diagnostic performance
of the product.
Introduction
In recent years, an increasing
number of Artificial Intelligence (AI) based image analysis algorithms for MR have
been developed and clinically introduced. In parallel, novel methods for
accelerated image acquisition are becoming clinically accepted, including
compressed sensing. For the eye, images using different acceleration techniques
may look similar, however, the choice of acceleration method can result in
differences on a detailed level. It is unclear if these virtually insignificant
differences in the MR images influence the performance of AI based algorithms.
One such acceleration technique is
the compressed sensing-sensitivity encoding (CS-SENSE) method (1), which
employs both coil sensitivities and sparsity for image acceleration. Since
becoming commercially available in 2018, it becomes more and more used in daily
clinical practice.
In this work, we investigate to what
extent a recently CE certified and FDA cleared sCT method is affected when
subjected to MR images accelerated with the CS-SENSE method. The sCT algorithm has
not been trained with CS-SENSE accelerated data.
We explore the influence of a
varying acceleration factor on the source images and on the sCT reconstructed
images both qualitatively and quantitatively. Methods
Five healthy volunteers were scanned
using a clinical 1.5T MR scanner (Philips Healthcare, Best, NL, software
release 5.7) using the built-in posterior coil. The study was approved by the local
IRB. RF-spoiled multi‐echo-gradient‐echo
MRI images of the lumbar spine were acquired sagittally with the following parameter
settings: FOV (AP,FH,RL) 220x278x100 mm3, acquisition voxel size
1x1x2 mm3, reconstruction voxel size 0.7x0.7x1 mm3, FA
10˚, TR/TE1/TE2 = 7ms/2.1 ms/4.2 ms, phase encode direction in feet-head with oversampling
in both F and H directions 90 mm, NSA 2. TE1 and TE2 were chosen almost
out-of-phase (OP) and in-phase (IP) for water-fat.
Four acquisitions were performed
with different acceleration settings: baseline scan with SENSE 1.2 (which is the
sCT product setting), and accelerated scans with CS-SENSE 1.45, CS-SENSE 2, and
CS-SENSE 3. All accelerated MR data were reconstructed with different denoising
level, which is a manual choice of the CS-SENSE product: No, Weak, Medium,
Strong.
Synthetic CTs were generated using
the CE certified product BoneMRI V1.5 (MRIguidance BV, Utrecht, The Netherlands).
Prior to comparison of images, registration (3) was performed for all
reconstructed sCT images to the sCT reconstructed from the baseline acquisition.
First, a visual inspection of the image
quality of both the source MRI images and the sCT reconstructions was
performed. Secondly, the performance of the sCT reconstruction algorithm was
quantitatively evaluated using two metrics:
- Structural similarity index for measuring
image quality (SSIM);
- Peak signal-to-noise ratio in dB (PSNR).
Both metrics were calculated for the entire sCT
imaging volume, after applying a body mask, with the baseline scan as a
reference.Results
Figure 1 shows the comparison of the
source OP and IP MR images and corresponding generated sCT images between the
baseline scan and accelerated with CS-SENSE 3 scans with different denoising
levels: no, weak, medium, and strong. Overall, more artifacts
and noise can be observed in the CS-SENSE accelerated source MR images, as well
as artifacts in the generated sCT images, especially at the level of L4-5 vertebrae where
SNR was already lower. The quality of the sCT generation improves with the use of a
stronger denoising level. Therefore for further analysis CS-SENSE
reconstruction with Strong denoising was used.
Baseline and accelerated source OP
and IP MR images and sCT images with different CS-SENSE factors and with
Strong denoising are shown in Figure 2. Increasing the acceleration factor
clearly introduces noise in the source images, which is less pronounced in the
sCT reconstructions. At CS-SENSE
factors of 1.45 and 2 the quality of the sCT reconstructions seems to be
minimally affected with virtually no loss of detail. CS-SENSE factor of 3 introduces
more artifacts and noise in the source MR images and loss of small details in
the sCT reconstructions: the vertebral edges are less sharp and less continuous
compared to the baseline images, as pinpointed by the red arrows.
Figure 3 shows an overview of the
generated sCT images for the five volunteers for different acceleration
factors. The overall quality of the sCT reconstructions for CS-SENSE 1.45 and 2
is virtually indistinguishable from the baseline scan, however with more
artifacts for CS-SENSE 3.
The scan times of all performed MRI
acquisitions are shown in Table 1.
Figure 4 shows the quantitative
results for the SSIM and the SNR calculated over five volunteers. PSNR values
are reduced for all CS-SENSE factors
with a similar amount, showing stable PSNR values for all acceleration factors,
whereas the SSIM decreases with increasing CS-SENSE factor. This reflects the
qualitative findings.Discussion and Conclusions
In this study, we have demonstrated that using
CS-SENSE up to an acceleration factor of 2 did not seem to compromise the
performance of a deep learning based sCT algorithm, based on qualitative and
quantitative assessment. At these acceleration levels, increased noise in
source MR images did not translate to noise in the sCT reconstructions. Note,
however, that a specific clinical question at hand will determine to what
extent the acquisition can be accelerated without degrading diagnostic
performance.Acknowledgements
This research was financially supported by the European Research Council (Grant no: 101020004).References
1. D. Liang
et.al. Accelerating SENSE Using Compressed Sensing. MRM 62 (2009)
2. M.C. Florkow
et.al. Deep learning–based MR‐to‐CT synthesis: The influence of
varying gradient
echo–based MR images as input channels. MRM 83 (2020)
3. R.J.A.
Kuiper et.al. CT to MR registration of complex deformations in the knee joint
through dual quaternion interpolation of rigid transforms. Phys Med Biol.
66 (2021)