Yolanda Ohene1,2, Elizabeth Powell3, Samo Lasič4,5, Geoff J. M. Parker3,6, Laura M. Parkes1,2, and Ben R. Dickie2,7
1Division of Psychology, Communication and Human Neuroscience, University of Manchester, Manchester, United Kingdom, 2Geoffrey Jefferson Brain Research Centre, University of Manchester, Manchester, United Kingdom, 3UCL, London, United Kingdom, 4Danish Research Centre for Magnetic Resonance, Copenhagen, Denmark, 5Random Walk Imaging, Åkarp, Sweden, 6Bioxydyn Limited, Manchester, United Kingdom, 7Division of Informatics, University of Manchester, Manchester, United Kingdom
Synopsis
Keywords: Data Processing, Permeability
Filter
exchange imaging (FEXI) is a promising technique for measuring water exchange across
the blood-brain barrier (BBB). However, the application of FEXI for the rodent
brain requires thinner slices and therefore higher crusher gradients which lead
to a progressive underestimation of the apparent exchange rate (AXR). Here, we
implement a crusher-compensated exchange rate (CCXR) model which reduces the
bias induced by the crusher gradients and allows more accurate estimates of BBB
water exchange in the rat brain.
INTRODUCTION
Filter exchange imaging (FEXI) is a
promising technique to measure water exchange across the blood-brain barrier (BBB;
termed BBB-FEXI) in the human brain [1, 2]. The implementation of BBB-FEXI in the rodent brain would permit
controlled studies of disease using transgenic models. However, the markedly
smaller rodent brain requires thinner imaging slices. The minimum required magnitude
of the crusher gradients in the FEXI sequence increases for thinner slices (Figure
1). Previously, Lasič et al. demonstrated that high crusher gradient
magnitudes cause an underestimation of the apparent exchange rate (AXR) [3]. These effects are exacerbated when
using low filter b-values and biases are greater for low BBB exchange rates.
Therefore, there is a need to explore extended signal models which are able to
compensate for the effects of the crusher gradients, facilitating the use of
thinner imaging slices needed for accurate BBB water exchange estimates in the
rodent brain.METHODS
Simulations
To investigate the impact of crusher gradients on ADC'(tm)
and ADCeq(tm) at three slice thicknesses ∆z = 2.5, 4.0 and 10.0 mm, synthetic signals incorporating the effects of
crusher gradients were generated in Matlab R2021a using a two-compartment model
[3]. Input parameters: water exchange
rate, kin = 2.38 s-1, k = kin
+ kout = AXR = 2.5 s-1 [4], intravascular volume fraction fi
= 0.05, intravascular and extravascular diffusivities, Di =
6.5 x 10-3 mm2/s and De = 0.65 x 10-3
mm2/s respectively [5]. Signals were simulated with the
filter block switched on and switched off (bf = 250 and 0 s/mm2 respectively) at mixing times, tm = 0.025,
0.05, 0.1, 0.2 and 0.3 s. The detection block was simulated with eight readout
b-values, b = 0, 25, 54, 116, 250, 539, 1160, 2500 s/mm2.
In vivo acquisition & analysis
To confirm simulated effects in vivo
at ∆z = 2.5 mm and 4.0 mm, male F334 rats (n = 5) were scanned with a Bruker
Avance III console interfaced with an Agilent 7T 16-cm bore magnet using the BBB-FEXI
sequence. Imaging parameters were matched to simulations for bf, tm and b = 0, 250 s/mm2; TR = 5000 s; matrix size = 64 x 64; FOV = 32 x 32 mm2;
single slice; repetitions = 10, with a spin-echo EPI readout.
The signals were evaluated in Matlab
R2021a using the standard AXR model [6, 7] and the full signal model [3], which accounts for effects of the
crusher gradients, here coined the crusher-compensated exchange rate (CCXR)
model: S = S'(tm) exp(-((qd2D + K)td)) x exp(-((qm2D + K)tm)) x exp(-((qf2D + K)tf)) f,
where qf, qm and qd
are the dephasing parameters associated with the filter block (f), mixing block
(m) which incorporates the effect crusher gradients and detection block (d) (see
Figure 1), D, K and f are matrices with intra-/extravascular
diffusivities, forward/ backward exchange rates (kin /
kout) and intra-/extra signal fraction respectively.
The intrasession
repeatability was evaluated in F334 rats (n = 15) for both AXR and CCXR models at
∆z = 4 mm. The first five repetitions were used for test and the second
five repetitions for retest. The coefficient
of variation (CoV) is given by sw/ µ x 100%.RESULTS
The simulations show that as slice
thickness decreases, recovery of ADC’(tm) becomes
progressively attenuated leading to underestimation of AXR (2.19 s-1,
1.16 s-1 to 0.28 s-1 for ∆z = 10.0 mm, 4.0 mm to
2.5 mm respectively), which is also reflected in the in vivo data where AXR = 1.24 ± 0.07 s-1 and 0.49 ± 0.06
s-1 at ∆z = 4.0 mm and 2.5 mm respectively (Figure 2). The CCXR
model was able to recover the ground truth kin value of
2.38 s-1 for the signals simulated at both slice thicknesses at 4.0
mm and 2.5 mm, as expected (Figure 3a-b). In the in vivo protocol, the CCXR
model gave mean kin values of 3.00 s-1 and 3.23 s-1
for ∆z = 4.0 mm and 2.5 mm respectively (Figure 3c-d). The mean
water exchange values (± standard deviation) for the repeatability study were AXR = 0.97 ± 0.32 s-1 and kin
= 3.11 ± 1.08 s-1 (Figure 4). Both models demonstrated
reasonable repeatability with CoV of 33%
and 35% for the AXR and CCXR models respectively.DISCUSSION & CONCLUSION
Simulated and in vivo data show progressive
underestimation of BBB water exchange measurements with decreasing slice
thickness when using the AXR model. The CCXR model removes the biases induced
by the crusher gradients, and provides
measurements of water exchange in the rodent brain that are more consistent
with the literature estimates (~2.5 s-1 [4]). The CCXR model enables thinner slices
in both the human and the rodent brain, which could be valuable for probing
specific small brain regions particularly affected in diseases, such as the
hippocampus in Alzheimer’s disease.Acknowledgements
This
work is funded by the EPSRC: grant code EP/S031510/1. SL has
received funding from the European Research Council (ERC) under the European
Union’s Horizon 2020 research and innovation programme
(grant agreement No 804746).References
[1] Bai, R., et al., Feasibility of filter-exchange imaging
(FEXI) in measuring different exchange processes in human brain.
NeuroImage, 2020. 219: p. 117039.
[2] Powell, E.O., Yolanda; Battiston,
Marco; Parkes, Laura M.; Parker, Geoff J. M. , Voxel-wise compartmental modelling of blood-brain barrier water
exchnage measurements using FEXI, in 31st
Annual Meeting of International Society for Magnetic Resonance in Medicine.
2022: London, UK.
[3] Lasič, S., et al., Effects of imaging gradients in sequences
with varying longitudinal storage time—Case of diffusion exchange imaging.
Magnetic Resonance in Medicine, 2018. 79(4):
p. 2228-2235.
[4] Dickie, B.R., G.J.M. Parker, and
L.M. Parkes, Measuring water exchange
across the blood-brain barrier using MRI. Progress in Nuclear Magnetic
Resonance Spectroscopy, 2020. 116:
p. 19-39.
[5] Scott, L.A., et al., Characterisation of microvessel blood
velocity and segment length in the brain using multi-diffusion-time
diffusion-weighted MRI. Journal of Cerebral Blood Flow & Metabolism,
2020. 41(8): p. 1939-1953.
[6] Lasič, S., et al., Apparent exchange rate mapping with
diffusion MRI. Magnetic Resonance in Medicine, 2011. 66(2): p. 356-365.
[7] Nilsson, M., et
al., Noninvasive mapping of water
diffusional exchange in the human brain using filter-exchange imaging. Magn
Reson Med, 2013. 69(6): p. 1573-81.