Ke Zhang1, Simon M.F. Triphan1, Oliver Sedlaczek1,2, Christian Ziener2, Hans-Ulrich Kauczor1, Heinz-Peter Schlemmer2, and Felix T. Kurz2
1Department of Diagnostic and Interventional Radiology, Heidelberg University Hospital, Heidelberg, Germany, 2Department of Radiology, German Cancer Research Center, Heidelberg, Germany
Synopsis
Keywords: Pulse Sequence Design, Arterial spin labelling
Liver perfusion is an important physiological parameter in health and
disease (1,2). In the measurement of liver perfusion using arterial
spin labelling (ASL), respiratory motion is a major challenge. In this study,
respiratory motion information is acquired from a projection signal and used to
adjust the position of the excited slice in real time. The feasibility of
free-breathing multi-slice liver perfusion imaging using spin-echo EPI based
pseudo-continuous ASL (pCASL) with navigator-based slice tracking method is
investigated.
PURPOSE
To apply the navigator-based slice tracking method to prospectively
compensate the respiratory motion for liver pseudo-continuous arterial spin
labeling (pCASL) using spin-echo EPI (SE-EPI) acquisition.METHODS
Measurements
were performed using an 18-channel body and spine receive RF coil on a 1.5T
scanner (Aera, Siemens Healthineers AG, Erlangen, Germany). A single
gradient-echo slice selection and projection readout at the location of the
diaphragm along the inferior-superior (IS) direction was acquired as a
navigator. Navigator acquisition and fat suppression were inserted before each
transverse imaging slice of the readouts of 2D spin-echo EPI based pCASL (NAV-
pCASL) sequence (Fig. 1), with 50 measurements including 32 training navigators
at the beginning of measurements obtained in 5.8 minutes. Sequence parameters
were as follows: TE =12 ms, FOV=400×240 mm3, partial Fourier = 6/8,
in-plane iPAT factor=3, matrix size=120×72×8, resolution=3.3×3.3×8 mm3,
slice gap = 4 mm, labeling duration = 1650 ms, postlabeling delay (PLD) =2000
ms, TR = 7000 ms, TI for SPAIR (SPectral Attenuated Inversion Recovery) fat
suppression=90 ms, FOVnav = 200 mm, resnav = 64, Flip
anglenav = 15º, TRnav=4.22ms. The labeling pulse train and readout
were triggered by the pulse signal and played out only during the systolic
period. Hepatic blood flow was selectively
tagged using a tagging plane placed at the portal vein. As a reference, pCASL
scans using a timed breathing protocol were performed, referred to as
“Breathing Holding” (BH). The breath-hold extended over consecutive labeling,
PLD, and 2D SE-EPI phases. Subjects were instructed to take one deep breath
during the following time span until the TR of 7000 ms was reached and go back
into the exhaled breath-hold as soon as possible. Since the repetition time was
set to 7000 ms, 2740 ms were available in each repetition for comfortable
inhalation and exhalation. For comparison, pCASL scans without breath-holding
and navigator-based motion correction were also performed, referred to as “Free
Breathing” (FB).
Before motion
analysis, the interleaved navigator signals during image acquisition were
Fourier transformed and truncated to exclude RF saturation along IS direction
from the SE-EPI readout (Fig. 2). The position for this truncation was
calculated based on peaks fitted from the averaged training navigator and the
peaks from the first 32 interleaved navigators. The diaphragm position was
derived by calculating the phase difference of the interleaved navigator
signals at each acquisition after Fourier transform and truncation. The
unwrapped data from different coils were then combined by using coil clustering
(3) based on the first 32 interleaved navigators.
The motion information was then directly sent back to the sequence and slice
positioning was adjusted in real-time. This motion analysis and real-time
feedback was performed on the scanner, implemented in ICE (Image calculation
environment, Siemens Healthineers AG, Erlangen, Germany).RESULTS
The respiratory motion from navigator signals could be precisely
calculated (Fig. 2) and slice positioning was changed in real-time based on the
motion information. The subtracted perfusion weighted maps show higher
perfusion from NAV-pCASL than from BH and FB-pCASL (Fig. 3). More detailed
structure in the perfusion weighted maps could be observed in the NAV-pCASL
case (Fig. 3, 4).CONCLUSION
This study demonstrates the feasibility of navigator-based slice
tracking technique in liver pCASL using SE-EPI readout.Acknowledgements
This work was supported by
the Deutsche Forschungsgemeinschaft (DFG, German Research
Foundation), 507778062.References
1. Pandharipande
PV, Krinsky GA, Rusinek H, Lee VS. Perfusion imaging of the liver: Current
challenges and future goals. Radiology 2005;234(3):661-673.
2. Kim
SH, Kamaya A, Willmann JK. CT Perfusion of the Liver: Principles and
Applications in Oncology. Radiology 2014;272(2):321-343.
3. Zhang T, Cheng JY, Chen YX, Nishimura
DG, Pauly JM, Vasanawala SS. Robust Self-Navigated Body MRI Using Dense Coil
Arrays. Magn Reson Med 2016;76(1):197-205.