Ke Zhang1, Simon M.F. Triphan1, Oliver Sedlaczek1,2, Christian Ziener2, Hans-Ulrich Kauczor1, Heinz-Peter Schlemmer2, and Felix T. Kurz2
1Department of Diagnostic and Interventional Radiology, Heidelberg University Hospital, Heidelberg, Germany, 2Department of Radiology, German Cancer Research Center, Heidelberg, Germany
Synopsis
Keywords: Pulse Sequence Design, Arterial spin labelling
Velocity-selective pulses include VS saturation pulses (VSS) and
VS Inversion (VSI) pulses. Previous study (1) concluded that both dual-sBIR8-VSS and sinc-VSI
achieved the highest SNR efficiency among the VS labeling schemes. Overall, the
dual-sBIR8-VSS pulse was the most robust against field imperfections, whereas sinc-modulated
VSI pulse showed greater tSNR and was the best among the VSI methods. In this study, VSI
sequence with rectangular small flip-angle RF pulses (rect-VSI), sinc-VSI with
and without VS gradients during the control condition are compared. Bloch
simulation and in vivo experiments for their robustness against B
1,
B
0 variation and eddy current (EC) are investigated.
PURPOSE
To compare three different vessel-selective inversion (VSI) pulses for
noninvasive arterial spin labeling (ASL) measurement of perfusion in human
brain.METHODS
Numerical simulations using the Bloch equations based
on matrix rotation were performed to assess the properties of the Fourier
transform based velocity-selective inversion (FT-VSI) pulse trains with three
different settings using Matlab (MathWorks, Inc., Natick, MA, USA). The Mz
profile in the presence of B1 (0.6 to 1.4 of the nominal value, step
size 0.57) and B0 (-200 to 200 Hz, step size 40 Hz) variations, with
arterial T1 (1932 ms) and T2 (275 ms) relaxation were
simulated. To investigate the sensitivity to the eddy current (EC) effects,
Bloch simulations were performed for these different VS pulses with EC
amplitude of 0.25% and time constants of 0.1 to 1000 ms. Measurements were
performed using a 20-channel head/neck receiver coil on a 3T scanner (Prisma, Siemens
Healthineers AG, Erlangen, Germany). Pulse sequence diagrams for VSI-ASL with
3D acquisition are same as demonstrated in previous study (2). It includes five blocks within each repetition:
slab-selective saturation, label/control modules using FT-VSI pulse trains,
background suppression (BGS), flow-dephasing for suppressing large-vessel
signal, and 3D GRASE readout (2). Three FT-VSI pulse trains are compared: 1)
traditional VSI sequence with rectangular small flip-angle RF pulses
(rect-VSI); 2) sinc-modulated FT-VSI pulse (sinc-VSI) with VS gradients during
control condition, for sinc-VSI the amplitude of the nth pulse is given by
sinc(n/5), where n = -4:1:4). The flip angles of these 9 rectangular pulses add
up to 180°; 3) sinc-modulated FT-VSI pulse (sinc-VSI) without VS gradients
during control condition. The design of pulse trains are compared in Figure 1. Sequence parameters were as
follows: TE =17.2 ms, FOV=220×220 mm3, FA=120°, slice and in-plane
partial Fourier = 6/8, segments=2, matrix size=64×64×24, slice oversample =
16.7%, resolution=3.4×3.4×5 mm3, bandwidth = 2298 Hz, labeling/control
duration = 64 ms, cutoff velocity =2.8 cm/s, delay between the pre-saturation
and the labeling modules = 2 s, post-labeling delay (PLD) =1200 ms, TR = 4000
ms, timing for BGS = 0.52, 0.91 and 1.08 s; control-labeling pairs = 20, TA =
5.2 min. Each pulse scheme was
tested in three healthy subjects.RESULTS
The simulated Mz-velocity responses are shown in Figure 2.
Both rect-VSI and sinc-VSI based pulses were capable of labeling spins in a
narrow velocity band with reasonable robustness against field inhomogeneities.
The sinc-VSI pulse produced much smoother response in the “unperturbed”
velocity bands. The control profile of sinc-VSI without VS gradients during
control condition showed the best performance despite B0/B1+
inhomogeneities. Figure 3 shows
the simulated magnetization of the static tissue (V = 0) under the label
condition showing EC sensitivity of VSI pulses to EC components with different
time constants (ranging from 0.1 ms to 1000 ms) at different distances to the
iso-center of the magnet. The sinc-VSI without VS gradients during control
condition demonstrated good resistance to EC effects. Normalized perfusion
signal equal to the averaged subtraction of label/control pairs (∆S) divided by
S0,tissue, (∆S/S0,tissue), was calculated as an indicator
of perfusion signal level (Figure 4).
In the in vivo experiment, the sinc-VSI without VS gradients gave the highest
ASL signals at the gray matter but lower ASL signals at the region of basilar
arteries.DISCUSSION
The reason for this phenomenon is likely due to
the unbalanced selection of moving spins with different flow direction. In the third
scheme all moving spin with all flow directions were inverted in control
condition. However, only the moving spins along z direction were selected in
label condition. After applying VS gradients along all three directions during
labeling, the signal at the region of basilar arteries was restored, as shown
in Figure 5.CONCLUSION
This study compared three FT-VSI ASL schemes. In simulation, the
sinc-VSI without VS gradients during the control condition demonstrated the
best performance against B0/B1+ inhomogeneity
and EC effects. The in vivo experiment showed that the sinc-VSI without VS
gradients gives the highest ASL signals. However, in the in vivo experiment
this scheme showed ASL signal voids at the level of basilar arteries. To avoid this,
a VS gradient along x and y may need to be added in the labeling section.Acknowledgements
This work was supported by the Deutsche
Forschungsgemeinschaft (DFG, German Research Foundation), 507778062.References
1. Guo J, Das S, Hernandez-Garcia L.
Comparison of velocity-selective arterial spin labeling schemes. Magn Reson Med
2021;85(4):2027-2039.
2. Liu D, Xu F, Li
W, van Zijl PC, Lin DD, Qin Q. Improved velocity-selective-inversion arterial
spin labeling for cerebral blood flow mapping with 3D acquisition. Magn Reson
Med 2020;84(5):2512-2522.