Louai Aldayeh1, Mizan Rahman1, and Ross Venook1
1Boston Scientific Neuromodulation, Valencia, CA, United States
Synopsis
Keywords: Safety, Safety
Previous investigations of RF-induced heating of multiple implants
(active as well as passive) have shown that heating of an implant can be
impacted by the presence of a separate neighboring implant.
This work assessed to which extend this phenomenon is
present for the limited scope of two separate MR Conditional ipsilateral DBS implants,
one on one side of the body, and another on the other side. Both, RF-induced heating
and voltage injection level into the IPG were assessed.
A variety of DBS lead configurations, and clinically
relevant distances in-between the two separate implants, were thoroughly
assessed for a data-driven conclusion.
INTRODUCTION
We previously reported on multiple Active Implantable
Medical Devices (AIMDs) [1], showing how RF-induced lead electrode heating of
an IPG/lead-based system can be impacted by the presence of a neighboring
IPG/lead-based system. On the passive implants side, an investigation [2] reported
this for adjacent orthopedic implants, and, depending on the length, angle, and
proximity gap of the second implant compared to the first, RF-induced heating can
potentially increase significantly for specific experimental configurations.
This work investigated a specific scenario of two MR Conditional
AIMDs: one Deep Brain Stimulation (DBS) implant with the lead and Implantable
Pulse Generator (IPG) on the same side of the body (ipsilateral), and another DBS
ipsilateral implant on the other side of the body (with a second IPG).
The question was: Given that each of the two separate DBS ipsilateral
implants is MR Conditional, what is the MR safety status of the combined system
regarding RF-induced distal lead heating and RF-induced
voltage injection level into the IPGs from leads?METHODS
Per the described limited scope, all variants of clinical
routings (trajectories) of a DBS ipsilateral implant - whether for pectoral or abdominal
IPG locations - were evaluated for range finding of distances between the lead (and
IPG) of an Ipsilateral Implant on the one side of the body, to their
counterparts of an Ipsilateral Implant on the other side of the body. These
ranges of distances between the two implants were explored in testing the two
MR-induced hazards likely to be potentially impacted in a multiple-implant
setup: RF-induced distal lead heating and RF- induced injection level into the
IPG from the lead. Both assessments were done in a 1.5T birdcage radiated
testing environment.
We applied specific setups in mounting the two DBS implants in
the test medium so that each system’s routing has clinically relevant exposure
fields. In addition, the mentioned distances between the two separate implants,
were thoroughly explored, and for many variants of DBS leads and their mixed configurations.
This mounting was implemented in a rectangular ASTM phantom [3] utilizing two circular
holders that facilitated the 3D routings and system-to-system distances (See Figures
1 & 2).
Measuring RF-induced distal lead heating is very sensitive
to accurate temperature probe placement on lead electrodes. In addition, during
testing, minor displacements of these electrodes could occur. To minimize the potential
impact of these factors on results, the following order of testing was exercised
on all tests:
- Test setup 1: Mount DBS
system one, and run multiple tests, each with a different incident e-field
signature that is clinically relevant. RF-induced measurements are collected as
the baseline data of single DBS system #1.
·
- Test setup 2: Without
altering the mounted DBS system #1, mount DBS system #2 (per the spatial distances
between the two systems to explore). Run the same multiple tests above done for
Test setup 1. RF-induced measurements are collected from both systems. Data of
DBS system #1 has potential differences (delta) in results when compared to its
baseline data (from Test setup 1). And data of DBS system #2 has a potential
delta in results when compared to its baseline data (from Test setup 3).
-
Test setup 3: Remove
mounted DBS system #1, keeping mounted DBS system #2 intact. Run the same multiple
tests above. RF-induced measurements are collected as the baseline data of single
DBS system #2.
For both RF-induced distal
lead heating and RF-induced injection level into the IPG, dozens of tests were
conducted per the above scheme, reflecting various lead configurations and clinically
relevant spatial distances between the two systems.
Data analysis aimed at the delta percentage difference of DBS
system #1 data from Test setup 2 (i.e., in the presence of the DBS system #1)
as compared to its baseline from Test setup 1, along with the equivalent delta
percentage for DBS system #2.
RESULTS
Our testing showed that RF-induced heating and injection
level into the IPG of one DBS Ipsilateral Implant (on one side of the body), is
not impacted by the presence of another DBS Ipsilateral Implant on other side
of the body.
The change of heating or injection level in one system, detected
due to the presence of the other system, was either negligible, slightly
higher, or slightly lower. The slightly higher results were not consistently
repeatable, and their magnitudes did not exceed the error margin (assessed uncertainty
of ~25%) of the overall measurement protocol.CONCLUSION
Based on these experiments, for a patient implanted with a dual
implant system comprised of two DBS Ipsilateral Implants (with separate IPGs), the
MR Conditional safety performance is not impacted for either system due to the
presence of the other system.
As this investigation utilized Boston Scientific’s Vercise Genus™
DBS systems with Boston Scientific leads or Medtronic® leads (via Vercise™ M8
Adapter), and the two tested systems shared the same MR Conditional label, this
combinatorial system has test performance consistent with the independent MR Conditional
label (safety limits) of either single system.
This finding is encouraging, though it is limited to the
described scope and generalization to other combinatorial AIMD systems are
beyond the scope of this work.Acknowledgements
No acknowledgement found.References
- Aldayeh L, Rahman M, Venook
R. Impact of a neighboring device on in vitro lead heating measurements. ISMRM
2020 proceedings. https://archive.ismrm.org/2020/4180.html
- Guo R, Zheng J, Wang Y,
Zeng Q, Wang Q, Yang R, Kainz W, Chen J. Computational and experimental
investigation of RF-induced heating for multiple orthopedic implants. Magn
Reson Med. 2019 Nov;82(5):1848-1858. doi: 10.1002/mrm.27817. Epub 2019 Jun 10.
PMID: 31183897.
- ASTM F2182-09. Standard
Test Method for Measurement of RF Induced Heating Near Passive Implants During
MRI