Sukhoon Oh1, Seon-Eui Hong2, Hyung-Do Choi2, Dongwoo Chang3, Kiseong Kim4, and Sangwoo Kim5
1Korea Basic Science Institute, Cheongju, Korea, Republic of, 2Electronics and Telecommunications Research Institute, Daejeon, Korea, Republic of, 3Chungbuk National University, Cheongju, Korea, Republic of, 4BioBrain, Inc., Daejeon, Korea, Republic of, 5Daewon University College, Jechoen, Korea, Republic of
Synopsis
Keywords: Safety, Safety
Numerical investigations on
the safety of patient assistants exposed to a 1.2T open MRI are performed at different
poses of patient assistants (including wearing latex gloves), because current
regulations are concerned with only patient RF safety. Compared with the
patient, up to 29.8% of the patient 10-gram SAR was observed in the patient
assistant. To prevent possible RF hazards of a patient assistant during MRI
scans, certain clauses regarding the patient assistant’s poses or wearing
gloves must be added to the existing MRI screening forms.
Introduction
Patient assistants’ supports
during MR scans can afford considerable psychological comfort to patients,
enabling timely completion of all the required scans at an open MRI system. It
is one of the important advantages of an open MRI system comparing with a
closed magnet. Most regulations or RF safety guidelines for MRI concern only
the RF exposure of a patient1. However, there have been few studies and
guidelines on RF exposure of patient assistants during MRI scan2. Access of
patient assistants to patients is enabled by the shape of the open MRI system,
even during MRI scans. Therefore, it is necessary to evaluate RF energy
exposure of patient assistants. In this study, we numerically investigated RF
energy exposures of patient assistants using planar–shaped RF coil, which is
widely used transmit RF coil type, particularly at 1.2 T in an open MRI system.Methods
The planar Tx coil3,4 is top–down symmetrically (in the
z-direction) composed of RF shielding plates (800 mm in diameter), end-ring (15
mm in width), and planar RF coils (with 10 mm gap between end-ring and coil,
with 400 mm space between the coils), as shown in Fig. 1. Eight constant
current sources were circularly placed at the gap in equispacing order to
create a B1+ field around the z-axis in a CP mode at 51
MHz. The amount of applied RF power of the Tx RF coil was scaled to create the
same magnitude of B1+, for example, 2 μT5, at the
isocenter of the magnet for each pose of the patient assistant. Four different poses
of patient assistants were defined by changing the poses and patient contact
conditions using the posable Duke model (age: 34, sex: male, BMI: 23.1 kg/m3,
tissues: 77)6,7 as shown in Fig. 2. Including sitting pose (Fig. 2(a)), the
patient assistant also touched the patient (age: 11, sex: female, BMI: 16.7 kg/m3,
tissues: 75) with one hand (Fig. 2(b)), two hands stacked up (Fig. 2(c)), and
finally two hands on the patient’s hand and thigh (Fig. 2(d)). EM field
simulations of RF exposure of patient assistants were performed using Sim4Life (Zurich
Med Tech AG, Switzerland) with GPUs (two NVDIA Quadro RTX A6000s), CPU (Intel
Xeon 6126), and 704 GB of system memory. The 10-g averaged SAR was analyzed
from the patient assistant for each pose.Results and Discusstion
The lowest max10-gram SAR was found for sitting pose (Fig. 2(a),
0.1041 W/kg). The highest max 10-gram SAR was measured for the pose of one hand
of the patient’s assistant touched the patient (Fig. 2(b), 5.2606 W/kg). It is
approximately 50.5 times higher than the lowest max 10-gram SAR of the patient
assistant. The other two poses (Fig. 2(c) & Fig. 2(d)) of the patient
assistant showed an approximately 11% lower max 10-gram SAR level than one-hand
touching pose. This is because of the total mass difference within the exposed
electric field of the planar Tx RF coil, particularly around the edge of the
coil. A smaller mass tends to result in greater RF exposure. In addition, the
induced SAR level of the patient assistant would worsen if a patient assistant
was electrically connected to the patient (i.e., touching each other). No
significant differences in the max 10-gram SAR levels in patients were found,
regardless of the pose of the patient assistant.
The max 10-gram SARs were found at the finger tip of the left hand, which
is electrical connected to the patient. This signifies level of RF exposure in
the patient assistant. For this reason, we suggest wearing electrically
insulated gloves, such as latex, to prevent direct contact of the patient body.
While assisting patients by holding their body parts during MRI scans, the
patient assistants should be aware of the possible risk of RF exposure. In
addition, insulating gloves must be worn to minimize the possibility of
unwanted adverse events.Conclusion
We investigated the amount of
RF exposure in four different poses of a patient assistant during MRI scans at
1.2 T in an open MRI system. The planar Tx RF coil induces a higher max 10-gram
SAR level in the patient assistant, up to 50 times higher when the patient
assistant holds the patient’s body comparing with sitting pose. Based on our
study, we suggest adding certain clauses to the existing MRI screening forms to
prevent or minimize any possible RF hazards.Acknowledgements
This work was supported by the IT R&D program of MSIP/IITP [2019-0-00102, A Study on Public Health and Safety in a Complex EMF Environment].References
1. International Electrotechnical Commission
(IEC). IEC 60601-2-33: Medical electrical equipment-Part 2–33: Particular
requirements for the basic safety and essential performance of magnetic
resonance equipment for medical diagnosis, 2015.
2. Hong SE, Oh S, Choi HD. RF exposure assessment
for various poses of patient assistant in open MRI environment. Appl. Sci. 2021;11:4947.
3. Wei S, Yang W, Wang H, Chen L. Analysis of a
planar transmit coil used for vertical field MRI. In Proceedings of the 4th
International Conference on Biomedical Engineering and Informatics. Shanghai,
China, 2011;118–121.
4. Boskamp, EB. Flat RF body coil design for open
MRI. In Proceedings of the 22nd Annual EMBS International Conference. Chicago,
USA, 2000;2387–2389.
5. Collins CM, Smith MB. Signal–to–noise ratio
and absorbed power as functions of main magnetic field strength, and definition
of “90°” RF pulse for the head in the birdcage coil. Magn. Reson. Med. 2001;45:684–691.
6. Christ A, Kainz W, Hahn EG, Honegger K,
Zefferer M, Neufeld E, Rascher W, Janka R, Bautz W, Chen J. The virtual
family-development of surface-based anatomical models of two adults and two
children for dosimetric simulations. Phys. Med. Biol. 2010;55:N23–N38.
7. Gosselin MC, Neufeld E, Moser H, Huber E, Farcito
S, Gerber L, Jedensjö M, Hilber I, Di Gennaro FD, Lloyd B. Development of a new
generation of high-resolution anatomical models for medical device evaluation:
The virtual population 3.0. Phys. Med. Biol. 2014;59:5287–5303.