Md Zahidul Islam1, Ananda Kumar2, Jianfeng Zheng1, and Ji Chen1
1Univ of Houston, Houston, TX, United States, 2FDA, Silver Spring, MD, United States
Synopsis
Keywords: Safety, Safety
The effect of ferrite magnetic
beads to suppress the induced currents on a straight wire active implantable
medical devices (AIMDs) lead was investigated. Leads with one, two and no
ferrite beads were studied experimentally and using numerical simulations. The
application of ferrite beads on the leads significantly reduced induced
currents, and thus heating on experiments outside the MRI scanner. In the MRI magnet
no significant reduction in heating was observed due to magnetic saturation of
the ferrite beads. Suitable magnetic material for current suppression needs to
be developed for application in MRI scanners.
Introduction
Radiofrequency (RF)-induced
heating is a safety concern when patients with AIMDs undergo magnetic resonance
imaging (MRI). The leads used with AIMDs mainly consist of metallic wires
covered by exterior insulating materials. During MRI, the interaction between
the metallic lead and the radiated RF field from the MRI scanner may lead to an
antenna effect, which could induce a strong current along the lead body. The induced
energy propagates towards the tip electrodes and dissipates into tissues near
tip electrodes, leading to potential thermal injury1. An approach of
using ferromagnetic materials is studied here to reduce the current along the
lead and mitigate RF-induced heating. Methods
Manganese Zinc ferromagnetic beads were applied in a solid straight
insulated wire with a diameter of 1.58 mm. Due to the high permeability of the
ferrite bead, the induced current along the lead is reduced, and consequently
the reduction of the RF-induced heating near the electrodes. Three different
cases have been studied, as shown in Figure 1. A numerical simulation was
performed to evaluate the transfer functions (TF) with and without ferrite
beads using the method of moments using surface equivalence principles (FEKO,
Altair, USA). Transfer functions were also measured inside the ASTM phantom to understand
the ferrite beads effect; the TF measurement system is shown in Figure 2. To
measure the RF-induced heating inside the ASTM phantom, the solid wire with or
without ferrite beads was placed inside the gel (0.47S/m) filled ASTM phantom
under 1.5T RF birdcage coil manufactured by Zurich Med Tech (ZMT), shown in
Figure 3. Fiber optic probes were used to read the temperate at the tip of the
solid wire.Results
The results of both numerical
modeling and experimental studies are given in the figures below. Discussion
Figure 4 (a) shows the measured transfer functions inside the ASTM
phantom with and without ferrite beads. The amplitude of the current through
the solid wire was reduced with the use of single ferrite bead at 15 cm
distance from the tip. The current amplitude was further reduced by usi6.40ng
another ferrite bead at 30 cm distance from the tip. Electromagnetic model of
the straight solid wire was developed with and without ferrite beads, and
numerical simulation was performed to get the transfer functions. The transfer
functions from the simulation are shown in Figure 4 (b), and similar effects
due to the magnetic beads are observed. The measured RF-induced heating results
show a significant reduction due to ferrite beads, as shown in Figure 5 (b). The
force due to the magnetic beads were measured in a 1.5T MRI scanner, as shown
in Figure 5(c). With one bead and two beads, the deflection angle was 21
degrees and 35 degrees, and the pulling force was 9.31 g and 23.67 g
(equivalent mass), respectively.Conclusion
This study has demonstrated the possibility of using ferromagnetic
materials in the lead to reduce RF-induced heating. The measured and simulated
transfer functions agree and show the amplitude of the current reduction
through the lead by the application of the ferrite beads. Measured RF heating under
the RF body coil was also reduced significantly due to the application of
ferrite beads in the wire. The effect of the ferrite magnetic beads was reduced
once used inside the MR scanner due to saturation of the ferrite material by
the static magnetic field of the scanner. More suitable materials should be
used in the future. Disclaimer
The mention of commercial
products, their sources, or their use in connection with material reported
herein is not to be construed as either an actual or implied endorsement of
such products by the Department of Health and Human Services.Acknowledgements
No acknowledgement found.References
J. Liu, J. Zheng, Q. Wang, W. Kainz, and J. Chen,
"A Transmission Line Model for the Evaluation of MRI RF-induced Fields on AIMDs,"
IEEE Transactions Microwave Theory and Techniques, 2018.